Influence of residual stenosis after percutaneous coronary intervention with stent implantation on development of restenosis and stent thrombosis

被引:19
作者
Sick, P
Hüttl, T
Niebauer, J
Thiele, H
Lauer, B
Harnbrecht, R
Hentschel, B
Schuler, G
机构
[1] Univ Leipzig, Ctr Heart, Dept Cardiol, D-04289 Leipzig, Germany
[2] Univ Leipzig, Inst Med Informat Stat & Epidemiol, Dept Cardiol, D-04289 Leipzig, Germany
关键词
PALMAZ-SCHATZ STENTS; BALLOON-EXPANDABLE-STENT; INTRAVASCULAR ULTRASOUND; FRENCH-MULTICENTER; ARTERY DISEASE; PLACEMENT; ANTICOAGULATION; ANGIOPLASTY; PREDICTORS; THERAPY;
D O I
10.1016/S0002-9149(02)03100-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this study was to assess the effects of residual stenosis after single-stent implantation on the rate of stent thrombosis, as well as restenosis within a 6-month follow-up period. Coronary angiograms of 2,157 patients with 2,523 lesions treated with a single stent were analyzed by quantitative coronary angiography before, immediately after stent implantation, and at a planned 6-month follow-up. Lesions were classified into 4 subgroups according to the degree of residual stenosis after stent implantation: group 1, gross oversizing < - 15%; group 2, slight oversizing - 15% to <0%; group 3, mild residual 0% to < 15%; group 4, moderate residual 15% to <30%. Stent thrombosis rates were not significantly different among the 4 subgroups (group 1: 0 of 60 [0%]; group 2: 2 of 388 [0.5%]; group 3: 8 of 1,370 [0.6%]; group 4: 8 of 705 [1.1%]; p = NS for all). An adequate dosage of ticlopidine (250 mg twice daily) and aspirin (100 mg/day) led to a lower rate of stent thrombosis (6 of 2,189 cases) than inadequate dosages or missing therapy (12 of 343 cases). In 1,882 stenoses with angiographic follow-up (77.7%), gross oversizing of stents lead to a significantly higher increase of percent stenosis (p <0.001) associated with a higher restenosis rate (group 1: 34.7% vs groups 2, 3, and 4: 32.5%, 28.2%, and 29.6%, respectively). A multiple regression analysis was performed. Optimal results with regard to stent thrombosis and restenosis were achieved with mild residual stenoses between 0% and 15% after stent implantation. Oversizing of stents is no longer necessary with an adequate dosage of ticlopidine and aspirin. (C) 2003 by Excerpta Medica, Inc.
引用
收藏
页码:148 / 153
页数:6
相关论文
共 29 条
  • [1] Predictors of restenosis after coronary stent implantation
    Bauters, C
    Hubert, E
    Prat, A
    Bougrimi, K
    Van Belle, E
    McFadden, EP
    Amouyel, P
    Lablanche, JM
    Bertrand, M
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (06) : 1291 - 1298
  • [2] Enhanced onset of platelet inhibition with a loading dose of ticlopidine in ASA-treated stable coronary patients
    Berglund, U
    Lindahl, T
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 1998, 64 (02) : 215 - 217
  • [3] INTER-SERIES DIFFERENCES IN THE RESTENOSIS RATE OF PALMAZ-SCHATZ CORONARY STENT PLACEMENT - DIFFERENCES IN DEMOGRAPHICS AND POSTPROCEDURE LUMEN DIAMETER
    CARROZZA, JP
    KUNTZ, RK
    SCHATZ, RA
    LEON, M
    GOLDBERG, S
    SAVAGE, M
    FISCHMAN, D
    SENERCHIA, C
    DIVER, DJ
    BAIM, DS
    [J]. CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1994, 31 (03): : 173 - 178
  • [4] MORPHOLOGIC CHARACTERISTICS OF LESION FORMATION AND TIME-COURSE OF SMOOTH-MUSCLE CELL-PROLIFERATION IN A PORCINE PROLIFERATIVE RESTENOSIS MODEL
    CARTER, AJ
    LAIRD, JR
    FARB, A
    KUFS, W
    WORTHAM, DC
    VIRMANI, R
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (05) : 1398 - 1405
  • [5] Coronary stenting with a novel stainless steel balloon-expandable stent: Determinants of neointimal formation and changes in arterial geometry after placement in an atherosclerotic model
    Carter, AJ
    Laird, JR
    Kufs, WM
    Bailey, L
    Hoopes, TG
    Reeves, T
    Farb, A
    Virmani, R
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (05) : 1270 - 1277
  • [6] INTRACORONARY STENTING WITHOUT ANTICOAGULATION ACCOMPLISHED WITH INTRAVASCULAR ULTRASOUND GUIDANCE
    COLOMBO, A
    HALL, P
    NAKAMURA, S
    ALMAGOR, Y
    MAIELLO, L
    MARTINI, G
    GAGLIONE, A
    GOLDBERG, SL
    TOBIS, JM
    [J]. CIRCULATION, 1995, 91 (06) : 1676 - 1688
  • [7] Management of restenosis after coronary intervention
    Dangas, G
    Fuster, V
    [J]. AMERICAN HEART JOURNAL, 1996, 132 (02) : 428 - 436
  • [8] ANGIOGRAPHIC PREDICTORS OF RECURRENCE OF RESTENOSIS AFTER WIKTOR STENT IMPLANTATION IN NATIVE CORONARY-ARTERIES
    DEJAEGERE, P
    SERRUYS, PW
    BERTRAND, M
    WIEGAND, V
    MARQUIS, JF
    VROLICX, M
    PIESSENS, J
    VALEIX, B
    KOBER, G
    BONNIER, H
    RUTSCH, W
    UEBIS, R
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1993, 72 (02) : 165 - 170
  • [9] RESTENOSIS AFTER PLACEMENT OF PALMAZ-SCHATZ STENTS IN NATIVE CORONARY-ARTERIES - INITIAL RESULTS OF A MULTICENTER EXPERIENCE
    ELLIS, SG
    SAVAGE, M
    FISCHMAN, D
    BAIM, DS
    LEON, M
    GOLDBERG, S
    HIRSHFELD, JW
    CLEMAN, MW
    TEIRSTEIN, PS
    WALKER, C
    BAILEY, S
    BUCHBINDER, M
    TOPOL, EJ
    SCHATZ, RA
    [J]. CIRCULATION, 1992, 86 (06) : 1836 - 1844
  • [10] A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE
    FISCHMAN, DL
    LEON, MB
    BAIM, DS
    SCHATZ, RA
    SAVAGE, MP
    PENN, I
    DETRE, K
    VELTRI, L
    RICCI, D
    NOBUYOSHI, M
    CLEMAN, M
    HEUSER, R
    ALMOND, D
    TEIRSTEIN, PS
    FISH, RD
    COLOMBO, A
    BRINKER, J
    MOSES, J
    SHAKNOVICH, A
    HIRSHFELD, J
    BAILEY, S
    ELLIS, S
    RAKE, R
    GOLDBERG, S
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) : 496 - 501