Suppressed electrostatic ion chromatography with tetraborate as eluent and its application to the determination of inorganic anions in snow and rainwater

被引:10
作者
Hu, WZ
Tanaka, K
Haddad, PR
Hasebe, K
机构
[1] Univ Tasmania, Sch Chem, Hobart, Tas 7001, Australia
[2] Hokkaido Univ, Grad Sch Sci, Div Chem, Sapporo, Hokkaido 0600810, Japan
[3] Natl Ind Res Inst Nagoya, Kita Ku, Nagoya, Aichi 4628510, Japan
[4] Hokkaido Univ, Grad Sch Environm Earth Sci, Div Mat Sci, Sapporo, Hokkaido 0600810, Japan
关键词
electrostatic ion chromatography; water analysis; stationary phases; LC; Zwitterionic surfactant stationary phase; tetraborate; inorganic anions;
D O I
10.1016/S0021-9673(00)00340-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Tetraborate is investigated as the eluent ion for suppressed electrostatic ion chromatography (EIC) using a zwitterionic stationary phase. Good separation of a range of inorganic anions (SO42-, Cl-, NO2- Br-, NO3- ClO3-, and I-) was obtained, with detection limits for highly conducting ions (SO42-, Cl-, NO2-, Br- and NO3-) being less than 8.10(-8) M, and for weakly conducting anions (ClO3- and I-) being 2.7.10(-7) and 5.8.10(-7) M, respectively. Calibration curves were linear up to 1.8 mM of each analyte, Retention times were found to increase with increasing eluent concentration and a linear relationship was observed between log k' and log[Na2B4O7] for all analytes. This behaviour is attributed to the progressive formation of a binary electrical double layer at the surface of the zwitterionic stationary phase. Retention times of analytes could be manipulated by varying the concentration of the eluent. This new suppressed-EIC system was applied to the (SO42-, Cl-, NO3- NO2- and Br-) in snow and rainwater samples. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:161 / 165
页数:5
相关论文
共 8 条
[1]   DIRECT DETERMINATION OF INORGANIC-IONS AT SUB-PPB LEVELS BY ION CHROMATOGRAPHY USING WATER AS A MOBILE-PHASE [J].
HU, WZ ;
MIYAZAKI, A ;
TAO, H ;
ITOH, A ;
UMEMURA, T ;
HARAGUCHI, H .
ANALYTICAL CHEMISTRY, 1995, 67 (20) :3713-3716
[2]   Electrostatic ion chromatography using dilute electrolytes as eluents: a new method for separation of anions [J].
Hu, WZ ;
Haddad, PR .
ANALYTICAL COMMUNICATIONS, 1998, 35 (10) :317-320
[3]   Electrostatic ion chromatography using hydroxide solutions as mobile phase with suppressed conductivity detection [J].
Hu, WZ ;
Haddad, PR ;
Hasebe, K ;
Tanaka, K .
ANALYTICAL COMMUNICATIONS, 1999, 36 (08) :309-312
[4]   Direct determination of bromide, nitrate, and iodide in saline matrixes using electrostatic ion chromatography with an electrolyte as eluent [J].
Hu, WZ ;
Haddad, PR ;
Hasebe, K ;
Tanaka, K ;
Tong, P ;
Khoo, C .
ANALYTICAL CHEMISTRY, 1999, 71 (08) :1617-1620
[5]   ELECTROSTATIC ION CHROMATOGRAPHY [J].
HU, WZ ;
TAKEUCHI, T ;
HARAGUCHI, H .
ANALYTICAL CHEMISTRY, 1993, 65 (17) :2204-2208
[6]   New approach for controlling the partitioning of analyte ions in ion chromatography with a water mobile phase [J].
Hu, WZ ;
Haraguchi, H .
JOURNAL OF CHROMATOGRAPHY A, 1996, 723 (02) :251-258
[7]   Cyanuric acid-based eluent for suppressed anion chromatography [J].
Maurino, V ;
Minero, C .
ANALYTICAL CHEMISTRY, 1997, 69 (16) :3333-3338
[8]  
WEISS J, 1995, ION CHROMATOGRAPHY, P67