Cutoff values for bacteria and leukocytes for urine flow cytometer Sysmex UF-1000i in urinary tract infections

被引:109
作者
Manoni, Fabio [2 ]
Fornasiero, Lucia [2 ]
Ercolin, Mauro [2 ]
Tinello, Agostino [2 ]
Ferrian, Melissa [2 ]
Hoffer, Paolo [2 ]
Valverde, Sara [1 ]
Gessoni, Gianluca [1 ]
机构
[1] Civil Hosp, Serv Lab Med, I-30015 Venice, Italy
[2] Civil Hosp, Serv Lab Med, I-35043 Padua, Italy
关键词
Urinary tract infection; Rapid diagnosis; Bacteriuria; Pyuria; UF-1000i; DIAGNOSIS; CULTURE;
D O I
10.1016/j.diagmicrobio.2009.06.003
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Because urinary tract infections (UTIs) are a quite common disease, the gold standard for diagnosing UTIs is still bacterial culture, although a large percentage of samples are negative: unnecessary Cultures call be reduced by means of ail effective screening test. The analytic performance of a new urine cytometer, the UF-1000i, has been tested oil 1463 urine samples Submitted to our laboratory for culture. Bacteria and leukocyte counts have been compared by means of the UF-1000i with colony-forriling unit (CFU) quantification on citrate lactose electrolytes deficient agar to assess the best cutoff values. By using quantitative cultures and considering as positive a sample with 10 x 10(5) CFU/mL, 546 positive samples (37%) were observed. If compared with 10 x 10(5) CFU/mL, the cutoff values obtained were 125 bacteria/mu L and 40 leukocytes/mu L, respectively. Analytic parameters Such as sensitivity, specificity, positive predictive value, negative predictive value, and correctly classified incidence were satisfactory. Based oil the results obtained in this study, when using the UF-1000i analyzer for a screening test for UTI, a Cutoff value of 40 white blood cells/mu L should be adopted. The cutoff value for bacteria should be 125/mu L for those clinical conditions in which 10 x 10(5) CFU/mL indicates a positivity. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:103 / 107
页数:5
相关论文
共 14 条
[1]  
Bartlet R, 1983, J INFECT DIS, V79, P179
[2]   Testing by Sysmex UF-100 flow cytometer and with bacterial culture in a diagnostic laboratory: a comparison [J].
Evans, R. ;
Davidson, M. M. ;
Sim, L. R. W. ;
Hay, A. J. .
JOURNAL OF CLINICAL PATHOLOGY, 2006, 59 (06) :661-662
[3]  
KASS EH, 1956, T ASSOC AM PHYSICIAN, V69, P56
[4]   Evaluation of the sysmex UF-100 urine cell Analyzer as a screening test to reduce the need for urine cultures for community-acquired urinary tract unfection [J].
Kim, Shine Young ;
Kim, Young Jin ;
Lee, Sun Min ;
Hwang, Sang Hyun ;
Kim, Hyung Hoi ;
Son, Han Chut ;
Lee, Eun Yup .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2007, 128 (06) :922-925
[5]   Determination of cut-off values for leucocytes and bacteria for urine flow cytometer (UF-100) in urinary tract infections [J].
Koken T. ;
Aktepe O.C. ;
Serteser M. ;
Samli M. ;
Kahraman A. ;
Dogan N. .
International Urology and Nephrology, 2002, 34 (2) :175-178
[6]   DIAGNOSIS OF BACTERIURIA IN MEN - SPECIMEN COLLECTION AND CULTURE INTERPRETATION [J].
LIPSKY, BA ;
IRETON, RC ;
FIHN, SD ;
HACKETT, R ;
BERGER, RE .
JOURNAL OF INFECTIOUS DISEASES, 1987, 155 (05) :847-854
[7]   Field evaluation of a second-generation cytometer UF-100 in diagnosis of acute urinary tract infections in adult patients [J].
Manoni, F ;
Valverde, S ;
Antico, F ;
Salvadego, MM ;
Giacomini, A ;
Gessoni, G .
CLINICAL MICROBIOLOGY AND INFECTION, 2002, 8 (10) :662-668
[8]   LABORATORY IN THE DIAGNOSIS AND MANAGEMENT OF URINARY-TRACT INFECTIONS [J].
PAPPAS, PG .
MEDICAL CLINICS OF NORTH AMERICA, 1991, 75 (02) :313-325
[9]  
PEZZLO M, 1988, Clinical Microbiology Reviews, V1, P268
[10]   DETECTION OF BACTERIURIA AND PYURIA BY URISCREEN, A RAPID ENZYMATIC SCREENING-TEST [J].
PEZZLO, MT ;
AMSTERDAM, D ;
ANHALT, JP ;
LAWRENCE, T ;
STRATTON, NJ ;
VETTER, EA ;
PETERSON, EM ;
DELAMAZA, LM .
JOURNAL OF CLINICAL MICROBIOLOGY, 1992, 30 (03) :680-684