Resistance of coated polyetheretherketone lumbar interbody fusion cages against abrasion under simulated impaction into the disc space

被引:8
作者
Kienle, Annette [1 ]
Krieger, Alexander [1 ]
Willems, Karel [2 ]
Wilke, Hans-Joachim [3 ]
机构
[1] SpineServ GmbH & Co KG, Soeflinger Str 100, D-89077 Ulm, Germany
[2] AZ Delta Hosp, Roeselare, Belgium
[3] Univ Ulm, Inst Orthoped Res & Biomech, Ulm, Germany
关键词
Nanocoating; calcium phosphate; titanium; vacuum plasma spray coating; fusion cage; abrasion; wear; bonding strength; spine; BOND STRENGTH; TITANIUM; COATINGS; DEBRIS; PEEK; ADHESION;
D O I
10.1177/2280800018782854
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In order to improve osseointegration, polyetheretherketone (PEEK) interbody fusion cages are frequently surface coated. The bonding strength of the coatings is mostly investigated under unrealistic loading conditions. To close this gap, in this study, uncoated and coated cages were loaded in a clinical setup in order to investigate their real resistance against abrasion. Six uncoated, six calcium phosphate (CaP) nanocoated, and six titanium (Ti) nanocoated PEEK cages were tested in this study. The experimental setup was designed to mimic cage impaction into the intervertebral disc space using polyurethane (PU) foam blocks as vertebral body substitutes. The cage surface was inspected before and after impaction, and their weight was measured. Impaction resulted in abrasion at the tip of the ridges on the implant surface. The mean weight loss was 0.39 mg for the uncoated cages, 0.57 mg for the CaP nanocoated cages, and 0.75 mg for the Ti nanocoated cages. These differences were statistically significant. In conclusion, differences between the three cage types were found concerning the amount of abrasion. However, all three cages lost less weight than a comparative Ti plasma spray coated cage, which showed a mean weight loss of 2.02 mg. This may be because the plasma spray coating is significantly thicker than the two nanocoatings. If compared with the permitted amount of weight loss derived from an FDA guidance document, which is approximately 1.7 mg, the wear of the Ti plasma spray coated cage is above this limit, whereas the wear for all other cage types is below.
引用
收藏
页数:6
相关论文
共 25 条
  • [21] Titanium-coated PEEK Versus Uncoated PEEK Cages in Lumbar Interbody Fusion A Systematic Review and Meta-analysis of Randomized Controlled Trial
    Lv, Zheng-tao
    Xu, Yong
    Cao, Bin
    Dai, Jun
    Zhang, Si-yuan
    Huang, Jun-ming
    Liang, Shuang
    Jiang, Feng-xian
    CLINICAL SPINE SURGERY, 2023, 36 (05): : 198 - 209
  • [22] Promotion of higher rates of early fusion using activated titanium versus polyetheretherketone cages in adults undergoing 1-and 2-level transforaminal lumbar interbody fusion procedures: a randomized controlled trial
    Toop, Nathaniel
    Dhaliwal, Joravar
    Gifford, Connor S.
    Gibbs, David
    Keister, Alex
    Miracle, Shelby
    Forghani, Reza
    Grossbach, Andrew J.
    Farhadi, H. Francis
    JOURNAL OF NEUROSURGERY-SPINE, 2023, 39 (05) : 709 - 718
  • [23] FINITE ELEMENT ANALYSIS OF 3D PRINTED STAND-ALONE TRANSFORAMINAL LUMBAR INTERBODY FUSION CAGES UNDER VARIOUS LOADINGS
    Zhang, Yufei
    Kang, Minjae
    Su, Siheng
    PROCEEDINGS OF ASME 2023 INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, IMECE2023, VOL 13, 2023,
  • [24] Injection of Bupivacaine into Disc Space to Detect Painful Nonunion after Anterior Lumbar Interbody Fusion (ALIF) Surgery in Patients with Discogenic Low Back Pain
    Kimura, Seiji
    Ohtori, Seiji
    Orita, Sumihisa
    Inoue, Gen
    Eguchi, Yawara
    Takaso, Masashi
    Ochiai, Nobuyasu
    Kuniyoshi, Kazuki
    Aoki, Yasuchika
    Ishikawa, Tetsuhiro
    Miyagi, Masayuki
    Kamoda, Hiroto
    Suzuki, Miyako
    Sakuma, Yoshihiro
    Kubota, Gou
    Oikawa, Yasuhiro
    Inage, Kazuhide
    Sainoh, Takeshi
    Yamauchi, Kazuyo
    Toyone, Tomoaki
    Nakamura, Junichi
    Kishida, Shunji
    Sato, Jun
    Takahashi, Kazuhisa
    YONSEI MEDICAL JOURNAL, 2014, 55 (02) : 487 - 492
  • [25] Effect of Soft Tissue Thickness Over the Posterior Border of the Vertebral Body and Disc Space on Cage Placement During Posterior Lumbar Interbody Fusion: A Cadaveric Study
    Yang, Jae Hyuk
    Kasat, Niraj Sharad
    Whang, Jin Ho
    Kim, Min Keun
    Min, Kyueng-Whan
    Hong, Jae Young
    Modi, Hitesh N.
    Suh, Seung Woo