Resistance of coated polyetheretherketone lumbar interbody fusion cages against abrasion under simulated impaction into the disc space

被引:8
|
作者
Kienle, Annette [1 ]
Krieger, Alexander [1 ]
Willems, Karel [2 ]
Wilke, Hans-Joachim [3 ]
机构
[1] SpineServ GmbH & Co KG, Soeflinger Str 100, D-89077 Ulm, Germany
[2] AZ Delta Hosp, Roeselare, Belgium
[3] Univ Ulm, Inst Orthoped Res & Biomech, Ulm, Germany
关键词
Nanocoating; calcium phosphate; titanium; vacuum plasma spray coating; fusion cage; abrasion; wear; bonding strength; spine; BOND STRENGTH; TITANIUM; COATINGS; DEBRIS; PEEK; ADHESION;
D O I
10.1177/2280800018782854
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In order to improve osseointegration, polyetheretherketone (PEEK) interbody fusion cages are frequently surface coated. The bonding strength of the coatings is mostly investigated under unrealistic loading conditions. To close this gap, in this study, uncoated and coated cages were loaded in a clinical setup in order to investigate their real resistance against abrasion. Six uncoated, six calcium phosphate (CaP) nanocoated, and six titanium (Ti) nanocoated PEEK cages were tested in this study. The experimental setup was designed to mimic cage impaction into the intervertebral disc space using polyurethane (PU) foam blocks as vertebral body substitutes. The cage surface was inspected before and after impaction, and their weight was measured. Impaction resulted in abrasion at the tip of the ridges on the implant surface. The mean weight loss was 0.39 mg for the uncoated cages, 0.57 mg for the CaP nanocoated cages, and 0.75 mg for the Ti nanocoated cages. These differences were statistically significant. In conclusion, differences between the three cage types were found concerning the amount of abrasion. However, all three cages lost less weight than a comparative Ti plasma spray coated cage, which showed a mean weight loss of 2.02 mg. This may be because the plasma spray coating is significantly thicker than the two nanocoatings. If compared with the permitted amount of weight loss derived from an FDA guidance document, which is approximately 1.7 mg, the wear of the Ti plasma spray coated cage is above this limit, whereas the wear for all other cage types is below.
引用
收藏
页数:6
相关论文
共 25 条
  • [1] Endplate Changes with Polyetheretherketone Cages in Posterior Lumbar Interbody Fusion
    Elfiky, Tarek Anwar
    Patil, Nirmal Dhananjay
    Allam, Yasser
    Ragab, Raafat
    ASIAN SPINE JOURNAL, 2020, 14 (02) : 229 - 237
  • [2] Does impaction of titanium-coated interbody fusion cages into the disc space cause wear debris or delamination?
    Kienle, Annette
    Graf, Nicolas
    Wilke, Hans-Joachim
    SPINE JOURNAL, 2016, 16 (02) : 235 - 242
  • [3] Titanium-coated polyetheretherketone cages vs. polyetheretherketone cages in lumbar interbody fusion: A systematic review and meta-analysis
    Li, Songfeng
    Li, Xiyong
    Bai, Xiaohui
    Wang, Yunlu
    Han, Pengfei
    Li, Hongzhuo
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2023, 25 (06)
  • [4] Comparison of fusion rates following transforaminal lumbar interbody fusion using polyetheretherketone cages or titanium cages with transpedicular instrumentation
    Nemoto, Osamu
    Asazuma, Takashi
    Yato, Yoshiyuki
    Imabayashi, Hideaki
    Yasuoka, Hiroki
    Fujikawa, Akira
    EUROPEAN SPINE JOURNAL, 2014, 23 (10) : 2150 - 2155
  • [5] Comparison of fusion rates following transforaminal lumbar interbody fusion using polyetheretherketone cages or titanium cages with transpedicular instrumentation
    Osamu Nemoto
    Takashi Asazuma
    Yoshiyuki Yato
    Hideaki Imabayashi
    Hiroki Yasuoka
    Akira Fujikawa
    European Spine Journal, 2014, 23 : 2150 - 2155
  • [6] Vertebral Endplate Cyst as a Predictor of Nonunion After Lumbar Interbody Fusion Comparison of Titanium and Polyetheretherketone Cages
    Tanida, Shimei
    Fujibayashi, Shunsuke
    Otsuki, Bungo
    Masamoto, Kazutaka
    Takahashi, Yoshimitsu
    Nakayama, Takeo
    Matsuda, Shuichi
    SPINE, 2016, 41 (20) : E1216 - E1222
  • [7] Randomized Controlled Trial of Posterior Lumbar Interbody Fusion With Ti- and CaP-Nanocoated Polyetheretherketone Cages: Comparative Study of the 1-Year Radiological and Clinical Outcome
    Willems, Karel
    Lauweryns, Philippe
    Verleye, Gino
    Van Goethem, Johan
    INTERNATIONAL JOURNAL OF SPINE SURGERY, 2019, 13 (06) : 575 - 587
  • [8] Polyetheretherketone Versus Titanium Cages for Posterior Lumbar Interbody Fusion: Meta-Analysis and Review of the Literature
    Massaad, Elie
    Fatima, Nida
    Kiapour, Ali
    Hadzipasic, Muhamed
    Shankar, Ganesh M.
    Shin, John H.
    NEUROSPINE, 2020, 17 (01) : 125 - +
  • [9] Titanium interlayer-mediated hydroxyapatite-coated polyetheretherketone cage in transforaminal lumbar interbody fusion surgery
    Zhu, Ce
    He, Miaomiao
    Mao, Lili
    Yang, Huiliang
    Hu, Bowen
    Zhang, Li
    Feng, Ganjun
    Liu, Limin
    Song, Yueming
    BMC MUSCULOSKELETAL DISORDERS, 2021, 22 (01)
  • [10] A Comparison of Polyetheretherketone and Titanium-Coated Polyetheretherketone in Minimally Invasive Transforaminal Lumbar Interbody Fusion: A Randomized Clinical Trial
    Singhatanadgige, Weerasak
    Tangchitcharoen, Nattapat
    Kerr, Stephen J.
    Tanasansomboon, Teerachat
    Yingsakmongkol, Wicharn
    Kotheeranurak, Vit
    Limthongkul, Worawat
    WORLD NEUROSURGERY, 2022, 168 : E471 - E479