Innovative monochromatic x-ray source for high-quality and low-dose medical imaging

被引:13
作者
Silver, Eric H. [1 ]
Shulman, Seth D. [1 ]
Rehani, Madan M. [2 ]
机构
[1] Imagine Sci Inc, 90 Kerry Pl, Norwood, MA 02062 USA
[2] Massachusetts Gen Hosp, Boston, MA 02114 USA
关键词
monochromatic x‐ rays; image quality; mammography; phantoms; radiation dose; MAMMOGRAPHY; RADIATION; DETECTORS;
D O I
10.1002/mp.14677
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose An estimated 377 million diagnostic and interventional radiological exams are performed annually in the United States and approximately 4 to 5 billion globally. All use x-ray tubes that emit x-rays over a broad energy band, a technology that is more than a century old. Only a small fraction of the radiation is useful for imaging while the remaining fraction either increases the radiation dose received by the patient or degrades the image. Monochromatic x-rays can provide lower dose images in many of these radiological applications while maintaining or improving image quality. We report the development of the first monochromatic x-ray source suitable for low-dose, high-quality imaging in the clinic and demonstrate its first application and performance with mammography phantoms. Methods X-ray fluorescence was used to generate monochromatic x-rays with selectable energies from 18 to 60 keV. This patented technology was incorporated into a laboratory prototype of a monochromatic x-ray mammography system. Image quality was evaluated as a function of radiation dose in standard breast phantoms using the signal-to-noise ratio (SNR) measured for high and low contrast masses and microcalcifications. Spatial imaging properties were assessed from these images as well as from modulation transfer function (MTF) analysis. Measurements using an iodine contrast agent were also performed. The results were compared to those obtained using a commercially available, conventional x-ray mammography system. Results Our prototype system reduced radiation dose by factors of five to ten times for the same SNRs as obtained from the conventional system. This performance was demonstrated in phantoms simulating a wide range of lesion sizes and microcalcifications in a variety of breast thicknesses. The high SNRs for very thick breast phantoms provide evidence that screening with less breast compression is possible while maintaining image quality. Contrast-enhanced digital mammography (CEDM) with monochromatic x-rays was shown to provide a simpler and more effective technique at substantially lower radiation dose. The MTF value at 20% was 9 lp/mm. Conclusions The monochromatic x-ray system is more sensitive for imaging a wide range of breast sizes and compositions than conventional broadband mammography. High image quality and lower dose are its hallmarks. It also makes CEDM much more effective than current methods developed for use with conventional broadband mammography systems.
引用
收藏
页码:1064 / 1078
页数:15
相关论文
共 50 条
[31]   The impact of X-ray scatter correction software on abdomen radiography in terms of image quality and radiation dose [J].
Sayed, M. ;
Knapp, K. M. ;
Fulford, J. ;
Heales, C. ;
Alqahtani, S. J. .
RADIOGRAPHY, 2024, 30 (04) :1125-1135
[32]   High-resolution, low-dose phase contrast X-ray tomography for 3D diagnosis of human breast cancers [J].
Zhao, Yunzhe ;
Brun, Emmanuel ;
Coan, Paola ;
Huang, Zhifeng ;
Sztrokay, Aniko ;
Diemoz, Paul Claude ;
Liebhardt, Susanne ;
Mittone, Alberto ;
Gasilov, Sergei ;
Miao, Jianwei ;
Bravin, Alberto .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2012, 109 (45) :18290-18294
[33]   Monochromatic X-rays for Low-Dose Digital Mammography Preliminary Results [J].
Yoon, Kwon-Ha ;
Kwon, Young Man ;
Choi, Byoung-Jung ;
Son, Hyun Hwa ;
Ryu, Cheol Woo ;
Chon, Kwon Su ;
Park, Seong Hoon ;
Juhng, Sun Kwan .
INVESTIGATIVE RADIOLOGY, 2012, 47 (12) :683-687
[34]   Distributed source x-ray tube technology for tomosynthesis imaging [J].
Sprenger, F. ;
Calderon-Colon, X. ;
Cheng, Y. ;
Englestad, K. ;
Lu, J. ;
Maltz, J. ;
Paidi, A. ;
Qian, X. ;
Spronk, D. ;
Sultana, S. ;
Yang, G. ;
Zhou, O. .
MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
[35]   Low-dose X-ray irradiation induces morphological changes and cytoskeleton reorganization in osteoblasts [J].
Huang, Qun ;
Zhou, Zhiping ;
Yan, Fei ;
Dong, Qirong ;
Wang, Liming ;
Sha, Weiping ;
Xu, Qin ;
Zhu, Xianwei ;
Zhao, Lei .
EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2020, 20 (06)
[36]   Radiation hygiene in medical X-ray imaging [J].
Nekolla, E. A. ;
Griebel, J. ;
Brix, G. .
RADIOLOGE, 2010, 50 (11) :1039-1052
[37]   Breast Cancer Induced by X-Ray Mammography Screening? A Review Based on Recent Understanding of Low-Dose Radiobiology [J].
Pauwels, Ernest K. J. ;
Foray, Nicolas ;
Bourguignon, Michel H. .
MEDICAL PRINCIPLES AND PRACTICE, 2016, 25 (02) :101-109
[38]   Developing a novel radiosensitive metal ion-loaded PVA/NBT film as medical low-dose X-ray dosimeter [J].
Sutanto, Heri ;
Lo, Nixon Kenny ;
Alkian, Ilham ;
Triadyaksa, Pandji ;
Marhaendrajaya, Indras ;
Jonathan, Felix ;
Irviandi, Risnu ;
Arifin, Zaenal .
RADIATION PHYSICS AND CHEMISTRY, 2024, 224
[39]   Analysis and interpretation of the first monochromatic X-ray tomography data collected at the Australian Synchrotron Imaging and Medical beamline [J].
Stevenson, Andrew W. ;
Hall, Christopher J. ;
Mayo, Sheridan C. ;
Haeusermann, Daniel ;
Maksimenko, Anton ;
Gureyev, Timur E. ;
Nesterets, Yakov I. ;
Wilkins, Stephen W. ;
Lewis, Robert A. .
JOURNAL OF SYNCHROTRON RADIATION, 2012, 19 :728-750
[40]   The ultrafast high-peak power lasers in future biomedical and medical X-ray imaging [J].
Kieffer, J. C. ;
Fourmaux, S. ;
Krol, A. .
19TH INTERNATIONAL CONFERENCE AND SCHOOL ON QUANTUM ELECTRONICS: LASER PHYSICS AND APPLICATIONS, 2017, 10226