Use of the Off-the-Shelf t-Branch Device to Treat an Acute Type Ia Endoleak in a Symptomatic Juxtarenal Abdominal Aortic Aneurysm

被引:9
作者
Khanafer, Adib [1 ]
Khashram, Manar [2 ]
Ruiz, Carmen-Maria [1 ]
Mann, Dana [3 ]
Laing, Andrew [3 ]
机构
[1] Christchurch Hosp, Vasc Endovasc & Renal Transplant Unit, Ricarrton Ave, Christchurch 4071, Canterbury, New Zealand
[2] Univ Otago, Dept Vasc Surg, Christchurch, New Zealand
[3] Christchurch Hosp, Intervent Radiol, Ricarrton Ave, Christchurch 4071, Canterbury, New Zealand
关键词
abdominal aortic aneurysm; bifurcated stent-graft; endovascular aneurysm repair; juxtarenal aneurysm; multibranched stent-graft; sac enlargement; t-branch device; type la endoleak; FENESTRATED ENDOVASCULAR GRAFT; SUITABILITY; ENDOGRAFTS; REPAIR;
D O I
10.1177/1526602815618493
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To report the use of the off-the-shelf Zenith t-Branch Thoracoabdominal Endovascular Graft for the treatment of a symptomatic acute type Ia endoleak subsequent to previous infrarenal bifurcated endovascular aneurysm repair. Case Report: A 72-year-old man presented 4 years after implantation of an Anaconda stent-graft with sac expansion and type II endoleak, which was treated over the next 18 months with repeated coil and glue embolization of the leak and sac. Six months after the last attempt, imaging disclosed a new type Ia endoleak. Sac enlargement became symptomatic, and emergent treatment was performed using the off-the-shelf Zenith b-Branch device. The 4 visceral arteries were bridged with Fluency stent-grafts. The Anaconda stent-graft precluded the use of a standard Zenith universal distal body bifurcated device below the t-Branch graft, so a reversed tapered 16x20x82-mm iliac limb was placed to bridge the t-Branch to the larger Anaconda limb; the smaller 12-mm contralateral Anaconda limb was occluded. The procedure was concluded with a femorofemoral bypass. At 10 months after the procedure, the patient is well; the sac diameter has decreased, and there is no endoleak on imaging. Conclusion: The multibranched Zenith t-branch device may be useful in treating proximal endoleaks in bifurcated stent-grafts.
引用
收藏
页码:212 / 215
页数:4
相关论文
共 15 条
[1]   Rupture After Endovascular Abdominal Aortic Aneurysm Repair: A Multicenter Study [J].
Antonopoulos, Constantine N. ;
Kakisis, John D. ;
Giannakopoulos, Triantafillos G. ;
Andrikopoulos, Vasilios ;
Antoniadis, Pavlos ;
Bessias, Nikolaos ;
Dervisis, Konstantinos ;
Georgopoulos, Sotirios ;
Giannoukas, Athanasios ;
Kaperonis, Elias ;
Kiskinis, Dimitrios ;
Klonaris, Christos ;
Machairas, Anastasios ;
Papavassiliou, Vasilios ;
Saleptsis, Vasilios ;
Saratzis, Nikolaos ;
Seretis, Konstantinos ;
Tampakis, Charalambos ;
Liapis, Christos D. .
VASCULAR AND ENDOVASCULAR SURGERY, 2014, 48 (7-8) :476-481
[2]   Anatomical Suitability of the T-Branch Stent-Graft in Patients With Thoracoabdominal Aortic Aneurysms Treated Using Custom-Made Multibranched Endografts [J].
Bisdas, Theodosios ;
Donas, Konstantinos P. ;
Bosiers, Michel ;
Torsello, Giovanni ;
Austermann, Martin .
JOURNAL OF ENDOVASCULAR THERAPY, 2013, 20 (05) :672-677
[3]  
Buth Jacob, 2003, Semin Vasc Surg, V16, P95, DOI 10.1016/S0895-7967(03)00007-3
[4]   Reporting standards for endovascular aortic aneurysm repair [J].
Chaikof, EL ;
Blankensteijn, JD ;
Harris, PL ;
White, GH ;
Zarins, CK ;
Bernhard, VM ;
Matsumura, JS ;
May, J ;
Veith, FJ ;
Fillinger, MF ;
Rutherford, RB ;
Kent, KC .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (05) :1048-1060
[5]   Periscopes, Snorkels and Chimneys: No Smoke Without Fire? [J].
Chaudhuri, A. .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2014, 47 (03) :218-220
[6]   The role of open and endovascular treatment with fenestrated and chimney endografts for patients with juxtarenal aortic aneurysms [J].
Donas, Konstantinos P. ;
Eisenack, Markus ;
Panuccio, Giuseppe ;
Austermann, Martin ;
Osada, Nani ;
Torsello, Giovanni .
JOURNAL OF VASCULAR SURGERY, 2012, 56 (02) :285-290
[7]   Late conversion of aortic stent grafts [J].
Kelso, Rebecca L. ;
Lyden, Scan P. ;
Butler, Brett ;
Greenberg, Roy K. ;
Eagleto, Matthew J. ;
Clair, Daniel G. .
JOURNAL OF VASCULAR SURGERY, 2009, 49 (03) :589-595
[8]   Zenith p-branch standard fenestrated endovascular graft for juxtarenal abdominal aortic aneurysms [J].
Kitagawa, Atsushi ;
Greenberg, Roy K. ;
Eagleton, Matthew J. ;
Mastracci, Tara M. .
JOURNAL OF VASCULAR SURGERY, 2013, 58 (02) :291-300
[9]   Suitability of the Zenith p-Branch Standard Fenestrated Endovascular Graft for Treatment of Ruptured Abdominal Aortic Aneurysms [J].
Kristmundsson, Thorarinn ;
Sveinsson, Magnus ;
Bjorses, Katarina ;
Tornqvist, Per ;
Dias, Nuno .
JOURNAL OF ENDOVASCULAR THERAPY, 2015, 22 (05) :760-764
[10]   The chimney graft technique for preserving visceral vessels during endovascular treatment of aortic pathologies [J].
Moulakakis, Konstantinos G. ;
Mylonas, Spyridon N. ;
Avgerinos, Efthimios ;
Papapetrou, Anastasios ;
Kakisis, John D. ;
Brountzos, Elias N. ;
Liapis, Christos D. .
JOURNAL OF VASCULAR SURGERY, 2012, 55 (05) :1497-1503