Intermediate screws in short segment pedicular fixation for thoracic and lumbar fractures - A biomechanical study

被引:67
作者
Anekstein, Yoram [1 ]
Brosh, Tamar
Mirovsky, Yigal
机构
[1] Assaf Harofeh Med Ctr, Dept Orthopaed Surg, IL-70300 Zerifin, Israel
[2] Tel Aviv Univ, Biomech Lab, Sch Dent Med, IL-69978 Tel Aviv, Israel
来源
JOURNAL OF SPINAL DISORDERS & TECHNIQUES | 2007年 / 20卷 / 01期
关键词
fracture; spine; pedicular screw; intermediate; biomechanical;
D O I
10.1097/01.bsd.0000211240.98963.f6
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
To determine the effect of adding pedicle screws at the level of a burst fracture (intermediate screws) on the stiffness of a short segment pedicle fixation, an in vitro biomechanical study was carried out. Six fresh-frozen pig lumbar spine specimens were used. The flexibility of the intact specimens was examined in flexion, extension, lateral bending, and torsion. An unstable burst fracture model was created by the dropped-mass technique. The unstable spine specimens were instrumented with pedicle screws. The flexibility was tested again with and without intermediate screws. The addition of intermediate screws provided a smaller range of motion in flexion-extension (P < 0.001), torsion (P < 0.001), and lateral bending (P = 0.014). The slopes of the load displacement curves increased in flexion (P < 0.001), extension (P = 0.003), lateral bending (P = 0.003), and torsion (P = 0.006), signifying a decrease in flexibility. The addition of intermediate screws at the level of a burst fracture significantly increases the stiffness of a short segment pedicular fixation.
引用
收藏
页码:72 / 77
页数:6
相关论文
共 23 条
  • [1] Short-segment pedicle instrumentation of thoracolumbar burst fractures - Does transpedicular intracorporeal grafting prevent early failure?
    Alanay, A
    Acaroglu, E
    Yazici, M
    Oznur, A
    Surat, A
    [J]. SPINE, 2001, 26 (02) : 213 - 217
  • [2] ALVINE GF, 1997, J BONE JOINT SURG S, V79, P306
  • [3] STANDARD TEST FOR LABORATORY ANIMAL BONE
    BURSTEIN, AH
    FRANKEL, VH
    [J]. JOURNAL OF BIOMECHANICS, 1971, 4 (02) : 155 - &
  • [4] Short-segment pedicle instrumentation - Biomechanical analysis of supplemental hook fixation
    Chiba, M
    McLain, RF
    Yerby, SA
    Moseley, TA
    Smith, TS
    Benson, DR
    [J]. SPINE, 1996, 21 (03) : 288 - 294
  • [5] INFLUENCE OF BONE-MINERAL DENSITY ON THE FIXATION OF THORACOLUMBAR IMPLANTS - A COMPARATIVE-STUDY OF TRANSPEDICULAR SCREWS, LAMINAR HOOKS, AND SPINOUS PROCESS WIRES
    COE, JD
    WARDEN, KE
    HERZIG, MA
    MCAFEE, PC
    [J]. SPINE, 1990, 15 (09) : 902 - 907
  • [6] Mechanical evaluation of cross-link designs in rigid pedicle screw systems
    Dick, JC
    Zdeblick, TA
    Bartel, BD
    Kunz, DN
    [J]. SPINE, 1997, 22 (04) : 370 - 375
  • [7] DICK JC, 1994, J SPINAL DISORD, V7, P402
  • [8] A NEW DEVICE FOR INTERNAL-FIXATION OF THORACOLUMBAR AND LUMBAR SPINE FRACTURES - THE FIXATEUR-INTERNE
    DICK, W
    KLUGER, P
    MAGERL, F
    WOERSDORFER, O
    ZACH, G
    [J]. PARAPLEGIA, 1985, 23 (04): : 225 - 232
  • [9] 1992 VOLVO AWARD IN EXPERIMENTAL STUDIES - VERTEBRAL BURST FRACTURES - AN EXPERIMENTAL, MORPHOLOGIC, AND RADIOGRAPHIC STUDY
    FREDRICKSON, BE
    EDWARDS, WT
    RAUSCHNING, W
    BAYLEY, JC
    YUAN, HA
    [J]. SPINE, 1992, 17 (09) : 1012 - 1201
  • [10] GARFIN SR, 1997, ORTHOPAEDIC KNOWLEDG, P19