Adenosine Triphosphate Quantification Correlates Poorly with Microbial Contamination of Duodenoscopes

被引:25
作者
Olafsdottir, Lovisa B. [1 ,2 ]
Wright, Sharon B. [1 ,2 ]
Smithey, Anne [1 ]
Heroux, Riley [1 ]
Hirsch, Elizabeth B. [3 ,4 ]
Chen, Alice [1 ]
Lane, Benjamin [1 ]
Sawhney, Mandeep S. [5 ]
Snyder, Graham M. [1 ,2 ]
机构
[1] Beth Israel Deaconess Med Ctr, Div Infect Control Hosp Epidemiol, Silverman Inst Hlth Care Qual & Safety, Boston, MA 02215 USA
[2] Beth Israel Deaconess Med Ctr, Dept Med, Div Infect Dis, Boston, MA 02215 USA
[3] Beth Israel Deaconess Med Ctr, Dept Pharm, Boston, MA 02215 USA
[4] Northeastern Univ, Dept Pharm & Hlth Syst Sci, Boston, MA 02115 USA
[5] Beth Israel Deaconess Med Ctr, Div Gastroenterol, Boston, MA 02215 USA
关键词
FLEXIBLE GASTROINTESTINAL ENDOSCOPES; ATP-BIOLUMINESCENCE; PREVENT INFECTIONS; DECONTAMINATION; CLEANLINESS; INDICATORS; GUIDELINE; CHANNELS;
D O I
10.1017/ice.2017.58
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
OBJECTIVE. The aim of this study was to quantify the correlation between adenosine triphosphate (ATP) measurements and bacterial cultures from duodenoscopes for evaluation of contamination following high-level disinfection. DESIGN. Duodenoscopes used for any intended endoscopic retrograde cholangiopancreatography (ERCP) procedure were included. Microbiologic and ATP data were collected concomitantly and in the same manner from ERCP duodenoscopes. SETTING. A high-volume endoscopy unit at a tertiary referral acute-care facility. METHODs. Duodenoscopes were sampled for ATP and bacterial contamination in a contemporaneous and highly standardized fashion using a "flush-brush-flush" method for the working channel (WC) and a dry flocked swab for the elevator mechanism (EM). Specimens were processed for any aerobic bacterial growth (colony-forming units, CFU). Growth of CFU > 0 and ATP relative light unit (RLU) > 0 was considered a contaminated result. Frequency of discord between among WC and EM measurements were calculated using 2 X 2 contingency tables. The Spearman correlation coefficient was used to calculate the relatedness of bacterial contamination and ATP as continuous measurements. RESULTS. The Spearman correlation coefficient did not demonstrate significant relatedness between ATP and CFU for either a WC or EM site. Among 390 duodenoscope sampling events, ATP and CFU assessments of contamination were discordant in 82 of 390 WC measurements (21%) and 331 of 390 of EM measurements (84.9%). The EM was frequently and markedly positive by ATP measurement. CONCLUSION. ATP measurements correlate poorly with a microbiologic standard assessing duodenoscope contamination, particularly for EM sampling. ATP may reflect biological material other than nonviable aerobic bacteria and may not serve as an adequate marker of bacterial contamination.
引用
收藏
页码:678 / 684
页数:7
相关论文
共 28 条
[1]   Evaluation of ATP Bioluminescence Assays for Potential Use in a Hospital Setting [J].
Aiken, Zoie A. ;
Wilson, Michael ;
Pratten, Jonathan .
INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2011, 32 (05) :507-509
[2]   Comparison of clinically relevant benchmarks and channel sampling methods used to assess manual cleaning compliance for flexible gastrointestinal endoscopes [J].
Alfa, Michelle J. ;
Olson, Nancy ;
Murray, Brenda-Lee .
AMERICAN JOURNAL OF INFECTION CONTROL, 2014, 42 (01) :E1-E5
[3]   Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels [J].
Alfa, Michelle J. ;
Fatima, Iram ;
Olson, Nancy .
AMERICAN JOURNAL OF INFECTION CONTROL, 2013, 41 (03) :245-248
[4]   The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels [J].
Alfa, Michelle J. ;
Fatima, Iram ;
Olson, Nancy .
AMERICAN JOURNAL OF INFECTION CONTROL, 2013, 41 (03) :249-253
[5]   Use of ATP bioluminescence for assessing the cleanliness of hospital surfaces: A review of the published literature (1990-2012) [J].
Amodio, Emanuele ;
Dino, Claudia .
JOURNAL OF INFECTION AND PUBLIC HEALTH, 2014, 7 (02) :92-98
[6]  
[Anonymous], EURO SURVEILL
[7]  
[Anonymous], 2015, FREQ ASK QUEST FAQS
[8]   Multidrug-Resistant Klebsiella Pneumoniae Outbreak after Endoscopic Retrograde Cholangiopancreatography [J].
Aumeran, C. ;
Poincloux, L. ;
Souweine, B. ;
Robin, F. ;
Laurichesse, H. ;
Baud, O. ;
Bommelaer, G. ;
Traore, O. .
ENDOSKOPIE HEUTE, 2011, 24 (02) :124-129
[9]   Usefulness of Adenosinetriphosphate Bioluminescence Assay (ATPmetry) for Monitoring the Reprocessing of Endoscopes [J].
Batailler, Pierre ;
Saviuc, Philippe ;
Picot-Gueraud, Romain ;
Bosson, Jean-Luc ;
Mallaret, Marie-Reine .
INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2015, 36 (12) :1437-1443
[10]   ESGE-ESGENA guideline for quality assurance in reprocessing: Microbiological surveillance testing in endoscopy [J].
Beilenhoff, U. ;
Neumann, C. S. ;
Rey, J. F. ;
Biering, H. ;
Blum, R. ;
Schmidt, V. .
ENDOSCOPY, 2007, 39 (02) :175-181