Dose measurements nearby low energy electronic brachytherapy sources using radiochromic film

被引:9
作者
Devic, Slobodan [1 ,2 ]
Liang, LiHeng [1 ,2 ]
Tomic, Nada [1 ,2 ]
Bekerat, Hamed [1 ,2 ]
Morcos, Marc [1 ,2 ,3 ]
Popovic, Marija [1 ,4 ]
Watson, Peter [1 ,4 ]
Aldelaijan, Saad [1 ,5 ,6 ]
Seuntjens, Jan [1 ,4 ]
机构
[1] McGill Univ, Med Phys Unit, Montreal, PQ, Canada
[2] Jewish Gen Hosp, Dept Radiat Oncol, Montreal, PQ, Canada
[3] Johns Hopkins Univ, Dept Radiat Oncol & Mol Radiat Sci, Baltimore, MD USA
[4] Montreal Gen Hosp, Dept Radiat Oncol, Montreal, PQ, Canada
[5] Montreal Neurol Inst, Biol & Biomed Engn Dept, Montreal, PQ, Canada
[6] King Faisal Specialist Hosp & Res Ctr, Biomed Phys Dept, Riyadh, Saudi Arabia
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2019年 / 64卷
基金
加拿大自然科学与工程研究理事会;
关键词
Electronic brachytherapy; Radiochromic film dosimetry; DOSIMETRY;
D O I
10.1016/j.ejmp.2019.05.017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: We investigate the effect of the GafChromic (TM) film EBT3 model absorbed dose energy response when used for dose measurements around low-energy photon sources. Monte Carlo based correction procedure in synergy with appropriate calibration curves was shown to provide more accurate absorbed dose (either relative or absolute). An assessment was made of possible dose errors that might be encountered if such energy dependent response is ignored. Methods: We measured PDDs in water from a Xoft 50 kVp source using EBT3 film, and compared to PDD measurements acquired with a PTW-TN34013 parallel-plate ionization chamber. For the x-ray source, we simulated spectra using the EGSnrc (BEAMnrc) Monte Carlo code, and calculated Half Value Layer (HVL) at different distances from the source in water. Measurement strips of EBT3 film were positioned at distances of 2-6 cm from the Xoft source in a water phantom using a custom-made holder and irradiated simultaneously. Results: Our results show that film calibration curves obtained at beam qualities near the effective energy of the Xoft 50 kVp source in water lead to variation in absorbed dose energy dependence of the response of around 5%. However, if the calibration curve was established in an MV beam quality, the error in absorbed dose could be as large as 20%. Conclusion: Accurate dose measurements using radiochromic films at low photon energies require that the radiochromic film dosimetry system be calibrated at appropriate corresponding low energies, as large absorbed dose errors are expected when film calibration is performed in MV beam qualities.
引用
收藏
页码:40 / 44
页数:5
相关论文
共 50 条
  • [31] Proposed linear energy transfer areal detector for protons using radiochromic film
    Mayer, Rulon
    Lin, Liyong
    Fager, Marcus
    Douglas, Dan
    McDonough, James
    Carabe, Alejandro
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2015, 86 (04)
  • [32] Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom
    Garcia-Garduno, Olivia A.
    Larraga-Gutierrez, Jose M.
    Rodriguez-Villafuerte, Mercedes
    Martinez-Davalos, Arnulfo
    Celis, Miguel A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 (02) : 250 - 253
  • [33] On the registration of low energy alpha particle with modified GafChromic EBT2 radiochromic film
    El-Naggar, Hala I.
    Ghanim, E. H.
    El Ghazaly, M.
    Salama, T. T.
    RADIATION PHYSICS AND CHEMISTRY, 2022, 191
  • [34] Beam quality correction factors for dose measurements around 192Ir brachytherapy sources
    Thrapsanioti, Zoi
    Peppa, Vasiliki
    Hourdakis, Costas J.
    Karaiskos, Pantelis
    Lekatou, Aristea
    Pantelis, Evaggelos
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (01):
  • [35] Commissioning of applicator-guided stereotactic body radiation therapy boost with high-dose-rate brachytherapy for advanced cervical cancer using radiochromic film dosimetry
    Aldelaijan, Saad
    Wadi-Ramahi, Shada
    Nobah, Ahmad
    Moftah, Belal
    Devic, Slobodan
    Jastaniyah, Noha
    BRACHYTHERAPY, 2017, 16 (04) : 893 - 902
  • [36] THE CONCEPT OF X-RAY CT DOSE EVALUATION METHOD USING RADIOCHROMIC FILM AND FILM-FOLDING PHANTOM
    Tanki, Nobuyoshi
    Katsuda, Toshizo
    Gotanda, Rumi
    Gotanda, Tatsuhiro
    Imai, Shinya
    Kawaji, Yasuyuki
    Noguchi, Atsushi
    Kuwano, Tadao
    Fujita, Hideki
    Takeda, Yoshihiro
    RADIATION PROTECTION DOSIMETRY, 2021, 193 (02) : 96 - 104
  • [37] Exploring the potential of mixed-source brachytherapy for the treatment of cervical cancer using high-dose rate 192Ir and/or 50 kV electronic sources
    Cohen, David M.
    Hiatt, Jessica R.
    O'Connor, Brigid
    Curran, Bruce
    Sternick, Edward S.
    Wazer, David E.
    BRACHYTHERAPY, 2011, 10 (02) : 141 - 146
  • [38] SKIN DOSE MEASUREMENTS USING RADIOCHROMIC FILMS, TLDS AND IONISATION CHAMBER AND COMPARISON WITH MONTE CARLO SIMULATION
    Alashrah, Saleh
    Kandaiya, Sivamany
    Maalej, Nabil
    El-Taher, A.
    RADIATION PROTECTION DOSIMETRY, 2014, 162 (03) : 338 - 344
  • [39] An investigation into factors affecting the precision of CT radiation dose profile width measurements using radiochromic films
    Li, Baojun
    Behrman, Richard H.
    MEDICAL PHYSICS, 2015, 42 (04) : 1765 - 1772
  • [40] Single-fraction low-energy electronic brachytherapy for conjunctival lymphoma
    Sarria, Gustavo R.
    Cabrera, Carla M.
    Sarria, Gustavo J.
    Buitrago, Mario
    Fuentes, Paola
    Serpa, Solon
    Giordano, Frank A.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2020, 12 (03) : 267 - 272