Dose measurements nearby low energy electronic brachytherapy sources using radiochromic film

被引:9
作者
Devic, Slobodan [1 ,2 ]
Liang, LiHeng [1 ,2 ]
Tomic, Nada [1 ,2 ]
Bekerat, Hamed [1 ,2 ]
Morcos, Marc [1 ,2 ,3 ]
Popovic, Marija [1 ,4 ]
Watson, Peter [1 ,4 ]
Aldelaijan, Saad [1 ,5 ,6 ]
Seuntjens, Jan [1 ,4 ]
机构
[1] McGill Univ, Med Phys Unit, Montreal, PQ, Canada
[2] Jewish Gen Hosp, Dept Radiat Oncol, Montreal, PQ, Canada
[3] Johns Hopkins Univ, Dept Radiat Oncol & Mol Radiat Sci, Baltimore, MD USA
[4] Montreal Gen Hosp, Dept Radiat Oncol, Montreal, PQ, Canada
[5] Montreal Neurol Inst, Biol & Biomed Engn Dept, Montreal, PQ, Canada
[6] King Faisal Specialist Hosp & Res Ctr, Biomed Phys Dept, Riyadh, Saudi Arabia
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2019年 / 64卷
基金
加拿大自然科学与工程研究理事会;
关键词
Electronic brachytherapy; Radiochromic film dosimetry; DOSIMETRY;
D O I
10.1016/j.ejmp.2019.05.017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: We investigate the effect of the GafChromic (TM) film EBT3 model absorbed dose energy response when used for dose measurements around low-energy photon sources. Monte Carlo based correction procedure in synergy with appropriate calibration curves was shown to provide more accurate absorbed dose (either relative or absolute). An assessment was made of possible dose errors that might be encountered if such energy dependent response is ignored. Methods: We measured PDDs in water from a Xoft 50 kVp source using EBT3 film, and compared to PDD measurements acquired with a PTW-TN34013 parallel-plate ionization chamber. For the x-ray source, we simulated spectra using the EGSnrc (BEAMnrc) Monte Carlo code, and calculated Half Value Layer (HVL) at different distances from the source in water. Measurement strips of EBT3 film were positioned at distances of 2-6 cm from the Xoft source in a water phantom using a custom-made holder and irradiated simultaneously. Results: Our results show that film calibration curves obtained at beam qualities near the effective energy of the Xoft 50 kVp source in water lead to variation in absorbed dose energy dependence of the response of around 5%. However, if the calibration curve was established in an MV beam quality, the error in absorbed dose could be as large as 20%. Conclusion: Accurate dose measurements using radiochromic films at low photon energies require that the radiochromic film dosimetry system be calibrated at appropriate corresponding low energies, as large absorbed dose errors are expected when film calibration is performed in MV beam qualities.
引用
收藏
页码:40 / 44
页数:5
相关论文
共 50 条
  • [1] Prospects for quantitative two-dimensional radiochromic film dosimetry for low dose-rate brachytherapy sources
    Le, Yi
    Ali, Imad
    Dempsey, James F.
    Williamson, Jeffrey F.
    MEDICAL PHYSICS, 2006, 33 (12) : 4622 - 4634
  • [2] Measurement of dose enhancement factor for Xoft Axxent electronic brachytherapy device using nanoparticle-embedded alginate film and radiochromic film
    Kakade, Nitin R.
    Das, Abhishek
    Kumar, Rajesh
    Sharma, S. D.
    Maiti, Nandita
    Chadha, Ridhima
    Sapra, B. K.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 (SUPPL 1) : S177 - S182
  • [3] Dosimetric characterization of low energy brachytherapy sources: Measurements
    Meigooni, AS
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 25 - 28
  • [4] A modified dose calculation formalism for electronic brachytherapy sources
    DeWerd, Larry A.
    Culberson, Wesley. S.
    Micka, John A.
    Simiele, Samantha J.
    BRACHYTHERAPY, 2015, 14 (03) : 405 - 408
  • [5] Validation of an electron Monte Carlo dose calculation algorithm in the presence of heterogeneities using EGSnrc and radiochromic film measurements
    Aubry, Jean-Francois
    Bouchard, Hugo
    Bessieres, Igor
    Lacroix, Frederic
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (04): : 2 - 14
  • [6] Dosimetry and calibration of low-energy brachytherapy sources: Implications for dose prescription
    Williamson, JF
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 21 - 24
  • [7] Radiochromic film dosimetry of a low energy proton beam
    Piermattei, A
    Miceli, R
    Azario, L
    Fidanzio, A
    delle Canne, S
    De Angelis, C
    Onori, S
    Pacilio, M
    Petetti, E
    Raffaele, L
    Sabini, MG
    MEDICAL PHYSICS, 2000, 27 (07) : 1655 - 1660
  • [8] Radiochromic film-based quality assurance for CT-based high-dose-rate brachytherapy
    Asgharizadeh, Saeid
    Bekerat, Hamed
    Syme, Alasdair
    Aldelaijan, Saad
    DeBlois, Francois
    Te Vuong
    Evans, Michael
    Seuntjens, Jan
    Devic, Slobodan
    BRACHYTHERAPY, 2015, 14 (04) : 578 - 585
  • [9] Measurements and Monte Carlo calculations to determine the absolute detector response of radiochromic film for brachytherapy dosimetry
    Bohm, TD
    Pearson, DW
    Das, RK
    MEDICAL PHYSICS, 2001, 28 (02) : 142 - 146
  • [10] MCNP modelling of vaginal and uterine applicators used in intracavitary brachytherapy and comparison with radiochromic film measurements
    Ceccolini, E.
    Gerardy, I.
    Rodenas, J.
    Van Dycke, M.
    Gallardo, S.
    Mostacci, D.
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2009, 164 (5-6): : 319 - 324