An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose

被引:15
作者
Cao, Bingyan [1 ]
Wang, Rui [1 ]
Gong, Chunxiu [1 ]
Wu, Di [1 ]
Su, Chang [1 ]
Chen, Jiajia [1 ]
Yi, Yajun [1 ]
Liu, Min [1 ]
Liang, Xuejun [1 ]
Li, Wenjing [1 ]
机构
[1] Capital Med Univ, Beijing Childrens Hosp, Dept Endocrinol Genet & Metab, Natl Ctr Childrens Hlth, Beijing 100045, Peoples R China
关键词
TISSUE;
D O I
10.1155/2019/4845729
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims. To evaluate the performance of a factory-calibrated flash glucose monitoring system in children with diabetes compared to venous blood glucose (BG). Methods. A total of 13 hospitalized participants newly diagnosed with type 1 diabetes, aged 114 years old, were involved in the study. Sensor glucose measurements on days 2, 3, 6, 7, 12, and 13 of wear were compared with venous BG. During these days, the venous BG results were obtained either 4 or 7 times per day. Results. The accuracy was evaluated against venous BG, with 469 of 469 (100.0%) sensor and venous BG pairs within consensus error grid zones A and B, including 94.7% in zone A. The overall mean absolute relative difference (MARD) was 11.67%. The MARD of blood glucose lower than 4.0 mmol/L (MARD=16.89%) was higher than blood glucose between 4 and 10 mmol/L (MARD=11.58%) and blood glucose higher than 10 mmol/L (MARD=7.79%). Compared to venous BG, the MARDs of wear days 2, 3, 6, 7, 12, and 13 were 11.53%, 9.66%, 11.79%, 10.89%, 13.18%, and 13.92%, respectively, with no statistically significant difference (P=0.25). The median ARD was highest when the glucose decreased >0.11 mmol/L/min (20.27%) and lower than 10.00% when the glucose changed between 0.06 and 0.11 mmol/L/min, changed <0.06 mmol/L/min, and increased >0.11 mmol/L/min. Conclusions. The accuracy of the system is good and remains stable over 14 days of wear; however, the accuracy depends on the glucose level and rates of glucose concentration changes.
引用
收藏
页数:7
相关论文
共 13 条
  • [1] Introduction
    不详
    [J]. DIABETES CARE, 2017, 40 : S1 - S130
  • [2] The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
    Bailey, Timothy
    Bode, Bruce W.
    Christiansen, Mark P.
    Klaff, Leslie J.
    Alva, Shridhara
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) : 787 - 794
  • [3] Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor
    Boyne, MS
    Silver, DM
    Kaplan, J
    Saudek, CD
    [J]. DIABETES, 2003, 52 (11) : 2790 - 2794
  • [4] A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring
    Cengiz, Eda
    Tamborlane, William V.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 : S11 - S16
  • [5] Deja Grazyna, 2018, Pediatr Endocrinol Diabetes Metab, V24, P11, DOI 10.18544/PEDM-24.01.0098
  • [6] A New Horizon for Glucose Monitoring
    Dovc, Klemen
    Bratina, Natasa
    Battelino, Tadej
    [J]. HORMONE RESEARCH IN PAEDIATRICS, 2015, 83 (03): : 149 - 156
  • [7] An alternative sensor-based method for glucose monitoring in children and young people with diabetes
    Edge, Julie
    Acerini, Carlo
    Campbell, Fiona
    Hamilton-Shield, Julian
    Moudiotis, Chris
    Rahman, Shakeel
    Randell, Tabitha
    Smith, Anne
    Trevelyan, Nicola
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 2017, 102 (06) : 543 - 549
  • [8] CHARACTERIZATION BY MICRODIALYSIS OF INTERCELLULAR GLUCOSE LEVEL IN SUBCUTANEOUS TISSUE IN HUMANS
    JANSSON, PA
    FOWELIN, J
    SMITH, U
    LONNROTH, P
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1988, 255 (02): : E218 - E220
  • [9] Ji Linong, 2017, J Diabetes Sci Technol, V11, P290, DOI 10.1177/1932296816662884
  • [10] Lodwig Volker, 2003, Diabetes Technol Ther, V5, P572, DOI 10.1089/152091503322250596