Correction of inner-filter effect in fluorescence excitation-emission matrix spectrometry using Raman scatter

被引:138
作者
Larsson, Tobias [1 ]
Wedborg, Margareta [1 ]
Turner, David [1 ]
机构
[1] Gothenburg Univ, Dept Chem, SE-41296 Gothenburg, Sweden
关键词
fluorescence; excitation-emission matrix; inner-filter effect; absorbance correction;
D O I
10.1016/j.aca.2006.09.067
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Fluorescence excitation-emission matrix (EEM) spectroscopy is a useful tool for interpretation of fluorescence information from natural. water samples. One of the major problems with this technique is the inner-filter effect (IFE), i.e. absorption of light at both the excitation and emission wavelengths. The common solutions are to either dilute the sample or apply some form of mathematical correction, most often based on the measured absorbance of the sample. Since dilution is not always possible, e.g. in on-line or in situ EEM recordings, and corrections based on absorbance are hampered primarily by the use of a separate absorbance instrument, neither of these solutions is optimal. In this work, we propose a mathematical correction procedure based on the intensity of Raman scatter from water. This procedure was found to reduce the error after correction by up to 50% in comparison with two absorbance correction procedures. Furthermore, it does not require the use of a separate absorbance measurement, and it is applicable to on-line and in situ EEM recordings, where the IFE would otherwise cause problems. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:357 / 363
页数:7