An effective correction algorithm for off-axis portal dosimetry errors

被引:16
作者
Bailey, Daniel W. [1 ]
Kumaraswamy, Lalith [2 ]
Podgorsak, Matthew B. [2 ]
机构
[1] SUNY Buffalo, Dept Phys, Buffalo, NY 14260 USA
[2] Roswell Pk Canc Inst, Dept Radiat Med, Buffalo, NY 14263 USA
关键词
portal dosimetry; IMRT QA; EPID; off-axis dose; PRETREATMENT VERIFICATION; IMAGING DEVICE; EPID DOSIMETRY; IMRT; BEAMS;
D O I
10.1118/1.3187785
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Portal dosimetric images acquired for IMRT pretreatment verification show dose errors of up to 15% near the detector edges as compared to dose predictions calculated by a treatment planning system for these off-axis regions. A method is proposed to account for these off-axis effects by precisely correcting the off-axis output factors, which calibrate the imager for absolute dose. Using this method, agreement between the predicted and the measured doses improves by tip to 15% for fields near the detector edges, resulting in passing rate improvements of as much as 60% for gamma evaluation of 3 rum, 3% within the collimator jaws. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3187785]
引用
收藏
页码:4089 / 4094
页数:6
相关论文
共 13 条
[1]   Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry [J].
Greer, PB .
MEDICAL PHYSICS, 2005, 32 (12) :3558-3568
[2]   Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device [J].
Greer, Peter B. .
MEDICAL PHYSICS, 2007, 34 (10) :3815-3824
[3]   Establishing action levels for EPID-based QA for IMRT [J].
Howell, Rebecca M. ;
Smith, Iris P. N. ;
Jarrio, Christie S. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (03) :16-25
[4]   Investigation of the optimal backscatter for an aSi electronic portal imaging device [J].
Ko, L ;
Kim, JO ;
Siebers, JV .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (09) :1723-1738
[5]   Comparison of the Efficacy of MapCHECK and Portal Dosimetry in the Pretreatment QA of IMRT Treatment Plans [J].
Liu, J. ;
Wang, Z. ;
Kumaraswamy, L. ;
Nazareth, D. ;
Podgorsak, M. ;
Malhotra, H. .
MEDICAL PHYSICS, 2008, 35 (06)
[6]   Monte Carlo modelling of a-Si EPID response: The effect of spectral variations with field size and position [J].
Parent, Laure ;
Seco, Joao ;
Evans, Phil M. ;
Fielding, Andrew ;
Dance, David R. .
MEDICAL PHYSICS, 2006, 33 (12) :4527-4540
[7]   Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device [J].
Pasma, KL ;
Dirkx, MLP ;
Kroonwijk, M ;
Visser, AG ;
Heijmen, BJM .
MEDICAL PHYSICS, 1999, 26 (11) :2373-2378
[8]   Monte Carlo computation of dosimetric amorphous silicon electronic portal images [J].
Siebers, JV ;
Kim, JO ;
Ko, L ;
Keall, PJ ;
Mohan, R .
MEDICAL PHYSICS, 2004, 31 (07) :2135-2146
[9]   Pretreatment verification of IMRT absolute dose distributions using a commercial a-Si EPID [J].
Talamonti, C. ;
Casati, M. ;
Bucciolini, M. .
MEDICAL PHYSICS, 2006, 33 (11) :4367-4378
[10]   The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields [J].
Van Esch, A ;
Depuydt, T ;
Huyskens, DP .
RADIOTHERAPY AND ONCOLOGY, 2004, 71 (02) :223-234