An effective correction algorithm for off-axis portal dosimetry errors
被引:16
作者:
Bailey, Daniel W.
[1
]
Kumaraswamy, Lalith
[2
]
Podgorsak, Matthew B.
[2
]
机构:
[1] SUNY Buffalo, Dept Phys, Buffalo, NY 14260 USA
[2] Roswell Pk Canc Inst, Dept Radiat Med, Buffalo, NY 14263 USA
关键词:
portal dosimetry;
IMRT QA;
EPID;
off-axis dose;
PRETREATMENT VERIFICATION;
IMAGING DEVICE;
EPID DOSIMETRY;
IMRT;
BEAMS;
D O I:
10.1118/1.3187785
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Portal dosimetric images acquired for IMRT pretreatment verification show dose errors of up to 15% near the detector edges as compared to dose predictions calculated by a treatment planning system for these off-axis regions. A method is proposed to account for these off-axis effects by precisely correcting the off-axis output factors, which calibrate the imager for absolute dose. Using this method, agreement between the predicted and the measured doses improves by tip to 15% for fields near the detector edges, resulting in passing rate improvements of as much as 60% for gamma evaluation of 3 rum, 3% within the collimator jaws. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3187785]