Heads-up intraoperative endoscopic imaging: A prospective evaluation of techniques and limitations

被引:17
作者
Levy, ML
Day, JD
Albuquerque, F
Schumaker, G
Giannotta, SL
McComb, JG
机构
[1] UNIV SO CALIF,CHILDRENS HOSP LOS ANGELES,DIV NEUROL SURG,LOS ANGELES,CA
[2] DARTMOUTH HITCHCOCK MED CTR,DIV NEUROL SURG,LEBANON,NH 03766
关键词
heads-up imaging; neuroendoscopy;
D O I
10.1097/00006123-199703000-00020
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: Endoscopes have been used adjunctively for spinal and cranial microsurgical procedures and directly for ventricular exploration, fenestration, and catheterization. Technological advances now allow for multi-imaging technologies, including a so-called ''heads-up display,'' allowing the surgeon to view the operative field and the endoscopic image simultaneously. INSTRUMENTATION: A high-resolution, active matrix liquid crystal display is built into the frame of the eyewear, with a display density of 182,000 pixels (280 x 650 pixels) and a resolution of 200 lines. The display occupies approximately 20% of the visual field. The headgear weighs 4 oz, with dimensions of 9 x 18 x 17 cm. The modular interface weighs 7 oz. Two different visualization systems can be used. The first uses a single integrated interface (IM 300), whereas the second (remote 900) is a 900-MHz frequency modulation wireless system that provides for a line-of-sight link between the NTSC source and the modular interface (range = 100 ft). RESULTS: Heads-up adjunctive endoscopy was used in 60 patients during 18 months. The procedures included 16 craniotomies, 1 intradural lumbar biopsy, and 42 ventriculoperitoneal shunt placements. Follow-up was obtained at a minimum of 1 year. Patient age ranged from 1 month to 58 years. The only complication was a transient loss of auditory evoked potentials after contact of the eighth nerve by the endoscope. CONCLUSION: We describe a portable, light-weight heads-up display imaging system, which we have used in 60 operative procedures. Benefits of the heads-up system include portability and a high-resolution digital monocular image, which reduces eye strain and vertigo. The ideal headgear will likely be an extremely high-resolution liquid crystal display-based or cathode ray tube-based semi-immersive system, with all of the benefits currently described for two-dimensional heads-up systems (i.e., light weight, portability, image quality, and the avoidance of complications associated with immersive systems).
引用
收藏
页码:526 / 530
页数:5
相关论文
共 31 条
[1]   NEUROSURGICAL ENDOSCOPY USING SIDE-VIEWING TELESCOPE - TECHNICAL NOTE [J].
APUZZO, MLJ ;
HEIFETZ, MD ;
WEISS, MH ;
KURZE, T .
JOURNAL OF NEUROSURGERY, 1977, 46 (03) :398-400
[2]   3-D IMAGING IN GASTROINTESTINAL LAPAROSCOPY [J].
BIRKETT, DH .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1993, 7 (06) :556-557
[3]   RETROPERITONEAL LAPAROSCOPIC NEPHRECTOMY UTILIZING 3-DIMENSIONAL CAMERA CASE-REPORT [J].
CHIU, AW ;
BABAYAN, RK .
JOURNAL OF ENDOUROLOGY, 1994, 8 (02) :139-141
[4]   ENDOSCOPE-ASSISTED CRANIOTOMY - APPROACH TO THE ROSTRAL BRAIN-STEM [J].
COHEN, AR ;
PERNECZKY, A ;
RODZIEWICZ, GS ;
GINGOLD, SI .
NEUROSURGERY, 1995, 36 (06) :1128-1129
[5]   VIRTUAL-REALITY AND LAPAROSCOPIC SURGERY [J].
COLEMAN, J ;
NDUKA, CC ;
DARZI, A .
BRITISH JOURNAL OF SURGERY, 1994, 81 (12) :1709-1711
[6]  
DAY JD, 1996, ASP NEUR SURG ANN M
[7]  
DAY JD, 1996, AM ASS NEUR SURG ANN
[8]  
*DISPL TECH, 1993, NEW FRONT VID IM INS
[9]  
ENISLIDIS C, 1995, J MED VIRT REALITY, P22
[10]   VENTRICULOFIBERSCOPE - A NEW TECHNIQUE FOR ENDOSCOPIC DIAGNOSIS AND OPERATION [J].
FUKUSHIMA, T ;
ISHIJIMA, B ;
HIRAKAWA, K ;
NAKAMURA, N ;
SANO, K .
JOURNAL OF NEUROSURGERY, 1973, 38 (02) :251-256