Introduction of a New Parameter for Evaluation of Digital Radiography System Performance

被引:1
|
作者
Choopani, Mohammad Reza [1 ]
Chaparian, Ali [2 ]
机构
[1] Isfahan Univ Med Sci, Fac Paramed, Student Res Comm, Esfahan, Iran
[2] Isfahan Univ Med Sci, Technol Radiol Dept, Esfahan, Iran
来源
JOURNAL OF MEDICAL SIGNALS & SENSORS | 2020年 / 10卷 / 03期
关键词
Contrast detail; contrast-detail radiography phantom; digital radiography; image quality; radiation dose; IMAGE QUALITY; COMPUTED RADIOGRAPHY; DETECTOR;
D O I
10.4103/jmss.JMSS_53_19
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: The aim of this study was to compare the image quality and radiation doses in various digital radiography systems using contrast-detail radiography (CDRAD) phantom. Methods: The image quality and radiation dose for seven different digital radiography systems were compared using the CDRAD phantom. Incident air kerma (IAK) values were measured for certain exposure settings in all digital radiography systems. The images from the CDRAD phantom were evaluated by three observers. The results were displayed in the form of a contrast-detail (CD) curve. In addition, the inverse image quality figure (IQFinv)-to-IAK ratios were used for quantitative comparison of different digital radiography system performance. Results: Results of this study showed that the CD curves cannot be suitable criterion for determining the performance of digital radiography systems. For this reason, IQFinv-to-radiation dose (IAK) ratios in a fixed radiation condition were used. The highest performance in terms of producing high-quality images and low radiation dose was related to X-ray unit 1 and the lowest performance was for X-ray unit 5. Conclusion: The ratio of IQFinv to IAK for performance evaluation of digital radiography systems is an innovation of this study. A digital radiography system with a higher IQFinv-to-IAK ratio is associated with lower patient dose and better image quality. Therefore, it is recommended to equip the new imaging centers with the systems that have higher IQFinv-to-IAK ratios.
引用
收藏
页码:196 / 200
页数:5
相关论文
共 50 条
  • [1] Introduction of Digital Radiography
    Jingu M.
    Yosetsu Gakkai Shi/Journal of the Japan Welding Society, 2023, 94 (02): : 213 - 217
  • [2] PERFORMANCE ASSESSMENT OF A COMMERCIAL DIGITAL RADIOGRAPHY SYSTEM
    NICKOLOFF, EL
    MEDICAL PHYSICS, 1984, 11 (05) : 744 - 744
  • [3] Physical and psychophysical evaluation of a new digital specimen radiography system for use in mammography
    Krupinski, EA
    Roehrig, H
    Schempp, W
    MEDICAL IMAGING 2000: IMAGE PERCEPTION AND PERFORMANCE, 2000, 3981 : 208 - 215
  • [4] New underwater digital radiography inspection system
    不详
    INSIGHT, 2010, 52 (11) : 591 - 591
  • [5] High Precision Performance Evaluation for Digital Radiography Detectors
    Kim, Dong Sik
    Lee, Eunae
    INTERNATIONAL WORKSHOP ON ADVANCED IMAGE TECHNOLOGY (IWAIT) 2019, 2019, 11049
  • [6] CONFERENCE ON DIGITAL RADIOGRAPHY - INTRODUCTION
    BRODY, WR
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1981, 314 : R9 - R10
  • [7] DIGITAL CHEST RADIOGRAPHY - PERFORMANCE EVALUATION OF A PROTOTYPE UNIT
    BARNES, GT
    SONES, RA
    TESIC, MM
    RADIOLOGY, 1985, 154 (03) : 801 - 806
  • [8] Update on digital radiography Introduction
    Schaefer-Prokop, Cornelia
    Uffmann, Martin
    EUROPEAN JOURNAL OF RADIOLOGY, 2009, 72 (02) : 193 - 193
  • [9] A STUDY OF LOW CONTRAST PERFORMANCE OF A DIGITAL RADIOGRAPHY SYSTEM
    SHAW, CG
    BASSANO, DA
    MEDICAL PHYSICS, 1982, 9 (04) : 609 - 609
  • [10] PRELIMINARY EVALUATION OF A DIGITAL SYSTEM FOR ROTATIONAL PANORAMIC RADIOGRAPHY
    DOVE, SB
    MCDAVID, WD
    WELANDER, U
    TRONJE, G
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1992, 73 (05): : 623 - 632