Early First Trimester Fetal Dose Estimation Method in a Multivendor Study of 16-and 64-MDCT Scanners and Low-Dose Imaging Protocols

被引:16
作者
Jaffe, Tracy A. [1 ]
Neville, Amy M. [1 ]
Anderson-Evans, Colin [2 ]
Long, Sheldon [3 ]
Lowry, Carolyn [1 ]
Yoshizumi, Terry T. [2 ]
Toncheva, Greta [2 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[2] Duke Univ, Med Ctr, Dept Occupat & Environm Safety Off, Durham, NC 27710 USA
[3] CSG Solut, Siemens Healthcare, Cary, NC USA
关键词
CT dosimetry; fetal dose; pregnancy; TUBE CURRENT MODULATION; RADIATION-EXPOSURE; PREGNANT PATIENTS; CT EXAMINATIONS; ROW CT; IRRADIATION; PATIENT; PELVIS; FETUS; DEPTH;
D O I
10.2214/AJR.08.2274
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to corroborate the relation between the estimated absorbed fetal dose derived from directly measured uterine doses early in the first trimester and the volume CT dose index (CTDI(vol)) for 16- and 64-MDCT of the maternal chest, abdomen, and pelvis. MATERIALS AND METHODS. Estimated absorbed fetal dose was measured with a metal oxide semiconductor field effect transistor (MOSFET) dosimeter placed in the expected uterine location in an anthropomorphic phantom of a woman and scanned with 16- and 64-MDCT units of one vendor and a 64-MDCT unit of another vendor. A trauma chest, abdomen, and pelvis protocol and an abdomen and pelvis protocol were used. Absorbed uterine dose was measured directly from the MOSFET detector. The CTDI(vol) for each protocol was recorded from the scanner console. Correlation between mean uterine dose and CTDI(vol) was tested with a goodness of fit model. RESULTS. The absorbed uterine dose ranged from 9.25 to 37.7 mGy. Absorbed fetal dose in the early first trimester correlated with CTDI(vol) in a linear regression equation. For the 16-MDCT scanner, at 130 kVp, the fetal dose was 2.091 x CTDI(vol) - 9.489. For the 64-MDCT scanner from the same vendor, at 120 kVp, the fetal dose was 1.113 x CTDI(vol) + 1.773. For the 64-MDCT scanner from the other vendor, at 120 kVp, the fetal dose was 1.378 x CTDI(vol) - 1.014. The goodness of fit results (R(2)) for the equations were 0.97, 0.98, and 0.99. CONCLUSION. Estimated absorbed fetal dose during the first trimester of pregnancy is linearly associated with CTDI(vol) regardless of beam energy, detector configuration, and scanner manufacturer.
引用
收藏
页码:1019 / 1024
页数:6
相关论文
共 37 条
[1]   Estimation of fetal and effective dose for CT examinations [J].
Adams, EJ ;
Brettle, DS ;
Jones, AP ;
Hounsell, AR ;
Mott, DJ .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 :272-278
[2]   Radiation dose to the fetus for pregnant patients undergoing multidetector CT imaging: Monte Carlo Simulations estimating fetal dose for a range of gestational age and patient size1 [J].
Angel, Erin ;
Wellnitz, Clinton V. ;
Goodsitt, Mitchell M. ;
Yaghmai, Nazanin ;
DeMarco, John J. ;
Cagnon, Christopher H. ;
Sayre, James W. ;
Cody, Dianna D. ;
Stevens, Donna M. ;
Primak, Andrew N. ;
McCollough, Cynthia H. ;
McNitt-Gray, Michael F. .
RADIOLOGY, 2008, 249 (01) :220-227
[3]   Radiation exposure and the pregnant patient [J].
Berlin, L .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (06) :1377-1379
[4]   Accidental embryo irradiation during barium enema examinations - An estimation of absorbed dose [J].
Damilakis, J ;
Perisinakis, K ;
Grammatikakis, J ;
Panayiotakis, G ;
Gourtsoyiannis, N .
INVESTIGATIVE RADIOLOGY, 1996, 31 (04) :242-245
[5]   Estimation of fetal radiation dose from computed tomography scanning in late pregnancy - Depth-dose data from routine examinations [J].
Damilakis, J ;
Perisinakis, K ;
Voloudaki, A ;
Gourtsoyiannis, N .
INVESTIGATIVE RADIOLOGY, 2000, 35 (09) :527-533
[6]  
DAMILAKIS J, 2004, EUR RADIOL SYLLABUS, V14, P33
[7]  
Dietrich M F, 2005, Health Phys, V88, pS10, DOI 10.1097/01.HP.0000147810.99736.70
[8]   Risk of childhood cancer from fetal irradiation [J].
Doll, R ;
Wakeford, R .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 :130-139
[9]   MR imaging of the female pelvis [J].
Fielding, JR .
RADIOLOGIC CLINICS OF NORTH AMERICA, 2003, 41 (01) :179-+
[10]   Estimation of dose to the unborn child at diagnostic X-ray examinations based on data registered in RIS/PACS [J].
Helmrot, Ebba ;
Pettersson, Hakan ;
Sandborg, Michael ;
Alten, Jonas Nilsson .
EUROPEAN RADIOLOGY, 2007, 17 (01) :205-209