Comparison of simultaneous invasive and noninvasive measurements of pressure gradients in congenital aortic valve stenosis

被引:21
作者
Barker, PCA [1 ]
Ensing, G [1 ]
Ludomirsky, A [1 ]
Bradley, DJ [1 ]
Lloyd, TR [1 ]
Rocchini, AP [1 ]
机构
[1] Univ Michigan, Congenital Heart Ctr, Ann Arbor, MI 48109 USA
关键词
D O I
10.1067/mje.2002.126415
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose. Congenital aortic valve stenosis is a common problem in pediatric cardiology. The catheter peak to peak systolic gradient is the accepted standard used for prognosis and intervention, but noninvasive correlation in pediatric patients is frequently associated with underestimation or overestimation of this gradient. The purpose of this study was to compare different noninvasive measurements with simultaneous catheter gradients to identify which best predicts the catheter peak to peak gradient. Methods. Twenty-five simultaneous Doppler and catheter measurements of aortic stenosis gradient were performed in 14 children (all 14 before valvuloplasty and 11 after valvuloplasty). Noninvasive estimates of pressure gradient were compared with catheter measurements with linear regression and Bland-Altman analysis. Results: The Doppler peak instantaneous pressure gradient overestimated the catheter peak to peak gradient but correlated well with the catheter peak instantaneous gradient. The Doppler mean systolic gradient correlated well with the catheter peak to peak gradient at low gradients and underestimated higher catheter gradients but agreed well at all levels with the catheter mean gradient. The modification of a catheter-derived correlation equation produced good correlation with the catheter peak to peak gradient (slope, 1.14; intercept, -1.8; R, 0.92), as did the use of estimated pressure recovery (slope, 1.04; intercept, 5.0; R, 0.94), calculated from a defined fluid mechanic equation. Conclusion: The catheter peak to peak gradient can be accurately estimated noninvasively using estimated pressure recovery or correlation equations incorporating Doppler measurements.
引用
收藏
页码:1496 / 1502
页数:7
相关论文
共 23 条
[1]   DISCREPANCIES BETWEEN DOPPLER AND CATHETER GRADIENTS IN AORTIC PROSTHETIC VALVES INVITRO - A MANIFESTATION OF LOCALIZED GRADIENTS AND PRESSURE RECOVERY [J].
BAUMGARTNER, H ;
KHAN, S ;
DEROBERTIS, M ;
CZER, L ;
MAURER, G .
CIRCULATION, 1990, 82 (04) :1467-1475
[2]   Overestimation of catheter gradients by Doppler ultrasound in patients with aortic stenosis: A predictable manifestation of pressure recovery [J].
Baumgartner, H ;
Stefenelli, T ;
Niederberger, J ;
Schima, H ;
Maurer, G .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 33 (06) :1655-1661
[3]   EFFECT OF STENOSIS GEOMETRY ON THE DOPPLER-CATHETER GRADIENT RELATION INVITRO - A MANIFESTATION OF PRESSURE RECOVERY [J].
BAUMGARTNER, H ;
SCHIMA, H ;
TULZER, G ;
KUHN, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (04) :1018-1025
[4]  
BAUMGARTNER H, 1991, CIRCULATION S2, V84, P165
[5]  
BAUMGARTNER H, 1991, CIRCULATION S2, V84, P164
[6]   HEMODYNAMIC DETERMINANTS OF THE PEAK SYSTOLIC LEFT VENTRICULAR-AORTIC PRESSURE-GRADIENT IN CHILDREN WITH VALVAR AORTIC-STENOSIS [J].
BEEKMAN, RH ;
ROCCHINI, AP ;
GILLON, JH ;
MANCINI, GBJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (08) :813-815
[7]   DOPPLER EVALUATION OF AORTIC-VALVE AREA IN CHILDREN WITH AORTIC-STENOSIS [J].
BENGUR, AR ;
SNIDER, AR ;
MELIONES, JN ;
VERMILION, RP .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (06) :1499-1505
[8]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[9]   Turbulent/viscous interactions control Doppler/catheter pressure discrepancies in aortic stenosis - The role of the Reynolds number [J].
Cape, EG ;
Jones, M ;
Yamada, I ;
VanAuker, MD ;
ValdesCruz, LM .
CIRCULATION, 1996, 94 (11) :2975-2981
[10]   FLUID-MECHANICS OF AORTIC-STENOSIS .1. THEORY AND STEADY FLOW EXPERIMENTS [J].
CLARK, C .
JOURNAL OF BIOMECHANICS, 1976, 9 (08) :521-528