Self-calibration methods in measurement channels for linear and non-linear conversion functions case

被引:1
作者
Gawedzki, Waclaw [1 ]
机构
[1] AGH Univ Sci & Technol, Dept Measurement & Instrumentat, PL-30059 Krakow, Poland
来源
2006 IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE PROCEEDINGS, VOLS 1-5 | 2006年
关键词
self-calibration; non-electrical monitoring; sensors;
D O I
10.1109/IMTC.2006.328648
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Self-calibration methods for long-term monitoring of non-electrical quantities are presented in the paper. There have been discussed the self-calibration methods of measurement channels containing integrated ratiometric output sensors and ratiometric output sensors with strong attenuation (bridge circuits). It has been shown that under certain system solutions this type sensors make measurement results independent from possible changes of the parameters of elements of electronic measuring systems caused by the temporal and thermal instability, and, also, the measurement accuracy improves. In the paper an algorithm is proposed, which enables set(calibration of measurement channels with non-linear conversion functions.
引用
收藏
页码:1506 / 1510
页数:5
相关论文
共 7 条
  • [1] *CAMPB SCI, 2001, VOLT MEAS ACC SELF C
  • [2] GAWEDZKI W, 1997, P 9 INT S EL INSTR I, P189
  • [3] GAWEDZKI W, 1999, IEEE C IMTC 99 VEN I
  • [4] GAWEDZKI W, 1996, IMEKO TC4 8 INT S NE
  • [5] LIU JG, 1999, MEASUREMENT, V26, P129, DOI DOI 10.1016/S0263-2241(99)00025-1
  • [6] *MATH WORKS INC, MATL SIM WIND US GUI
  • [7] Taylor J.R., 1982, An Introduction to Error Analysis