Evaluation of the Abbott m2000rt RealTime™ HIV-1 assay with manual sample preparation compared with the ROCHE COBAS® AmpliPrep™/AMPLICOR™ HIV-1 MONITOR® v1.5 using specimens from East Africa

被引:31
作者
Crump, John A. [1 ,2 ,3 ,4 ]
Scott, Lesley E. [5 ]
Msuya, Emma [3 ]
Morrissey, Anne B. [1 ]
Kimaro, Ekyafyose E. [3 ]
Shao, John F. [3 ,4 ]
Stevens, Wendy S. [5 ,6 ]
机构
[1] Duke Univ, Med Ctr, Dept Med, Div Infect Dis & Int Health, Durham, NC 27710 USA
[2] Duke Univ, Duke Global Hlth Inst, Durham, NC 27710 USA
[3] Kilimanjaro Christian Med Ctr, Moshi, Tanzania
[4] Tumaini Univ, Kilimanjaro Christian Med Coll, Moshi, Tanzania
[5] Univ Witwatersrand, Dept Mol Med & Haematol, Sch Pathol, Fac Hlth Sci, ZA-2050 Wits, South Africa
[6] Natl Hlth Lab Serv, Johannesburg, South Africa
基金
美国国家卫生研究院;
关键词
HIV; Reverse transcriptase polymerase chain reaction; Viral load; Validation studies; Tanzania; GENETICALLY DIVERSE PANEL; DRIED BLOOD SPOTS; VIRUS TYPE-1 RNA; QUANTITATIVE DETECTION; PERFORMANCE; NUCLISENS; PLASMA; TANZANIA; PCR; QUANTIFICATION;
D O I
10.1016/j.jviromet.2009.08.013
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The performance of the Abbott m2000rt RealTime (TM) HIV-1 assay (RealTime HIV-1) with manual sample preparation was compared against the ROCHE COBAS (R) AmpliPrep (TM)/AMPLICOR (TM) HIV-1 MONITOR Test v0.5 (CAP/CA HIV-1) using samples collected from 100 donors infected with HIV and 20 donors not infected with HIV in northern Tanzania where HIV-1 subtypes A, C, D, and their recombinant forms predominate. The RealTime HIV-1 appeared to have more within-run variability at high HIV-1 RNA concentrations, but total assay variability over the dynamic range tested was within the manufacturer's claim of < 0.3 SD copies/mL Accuracy studies showed 100% concordance for positive and negative values. When continuous values were examined, CAP/CA HIV-1 yielded higher values than the RealTime HIV-1 at higher nominal HIV-1 RNA concentrations. The RealTime HIV-1 assay showed excellent linearity between 2.5 and 7.0 log copies/mL Of negative samples, 100% showed negative results, and > 95% of samples with nominal concentrations of 40 copies/mL were detected at >= 40 copies/mL by RealTime HIV-1. Manual sample preparation may contribute to higher total assay variability. This study suggests that the Abbott m2000rt RealTime HIV-1 assay with manual sample preparation is an acceptable and feasible alternative to the conventional ROCHE COBAS AmpliPrep/AMLICOR HIV-1 Monitor v1.5 assay and that the RealTime HIV-1 assay performs well on samples from East Africa. (c) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:218 / 222
页数:5
相关论文
共 27 条
[1]   Performance of the Amplicor human immunodeficiency virus type 1 PCR and analysis of specimens with false-negative results [J].
Barlow, KL ;
Tosswill, JHC ;
Parry, JV ;
Clewley, JP .
JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (11) :2846-2853
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/ml [J].
Collins, ML ;
Irvine, B ;
Tyner, D ;
Fine, E ;
Zayati, C ;
Chang, CA ;
Horn, T ;
Ahle, D ;
Detmer, J ;
Shen, LP ;
Kolberg, J ;
Bushnell, S ;
Urdea, MS ;
Ho, DD .
NUCLEIC ACIDS RESEARCH, 1997, 25 (15) :2979-2984
[4]   Multicenter evaluation of the NucliSens EasyQ HIV-1 v1.1 assay for the quantitative detection of HIV-1 RNA in plasma [J].
de Mendoza, C ;
Koppelman, M ;
Montès, B ;
Ferre, V ;
Soriano, V ;
Cuypers, H ;
Segondy, M ;
Oosterlaken, T .
JOURNAL OF VIROLOGICAL METHODS, 2005, 127 (01) :54-59
[5]   Comparison of NucliSens and Roche monitor assays for quantitation of levels of human immunodeficiency virus type 1 RNA in plasma [J].
Dyer, JR ;
Pilcher, CD ;
Shepard, R ;
Schock, J ;
Eron, JJ ;
Fiscus, SA .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (02) :447-449
[6]   HIV-1 viral load assays for resource-limited settings: Authors' reply [J].
Fiscus, Susan A. .
PLOS MEDICINE, 2006, 3 (12) :2461-2461
[7]   Comparative evaluation of the performance of the Abbott real-time human immunodeficiency virus type I (HIV-1) assay for measurement of HIV-1 plasma viral load following automated specimen preparation [J].
Garcia-Diaz, A. ;
Clewley, G. S. ;
Booth, C. L. ;
Labett, W. ;
McAllister, N. ;
Geretti, A. M. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2006, 44 (05) :1788-1791
[8]   Correlation between Human Immunodeficiency Virus Type 1 (HIV-1) RNA Measurements Obtained with Dried Blood Spots and Those Obtained with Plasma by Use of Nuclisens EasyQ HIV-1 and Abbott RealTime HIV Load Tests [J].
Garrido, Carolina ;
Zahonero, Natalia ;
Corral, Angelica ;
Arredondo, Miguel ;
Soriano, Vincent ;
de Mendoza, Carmen .
JOURNAL OF CLINICAL MICROBIOLOGY, 2009, 47 (04) :1031-1036
[9]   Failure to detect a non-B HIV-1 subtype by the HIV-1 Amplicor Monitor test, version 1.5: a case of unexpected vertical transmission [J].
Geelen, S ;
Lange, J ;
Borleffs, A ;
Wolfs, T ;
Weersink, A ;
Schuurman, R .
AIDS, 2003, 17 (05) :781-782
[10]   A new ultrasensitive assay for quantitation of HIV-1 RNA in plasma [J].
Johanson, J ;
Abravaya, K ;
Caminiti, W ;
Erickson, D ;
Flanders, R ;
Leckie, G ;
Marshall, E ;
Mullen, C ;
Ohhashi, Y ;
Perry, R ;
Ricci, J ;
Salituro, J ;
Smith, A ;
Tang, N ;
Vi, M ;
Robinson, J .
JOURNAL OF VIROLOGICAL METHODS, 2001, 95 (1-2) :81-92