Evaluation of the Abbott m2000rt RealTime™ HIV-1 assay with manual sample preparation compared with the ROCHE COBAS® AmpliPrep™/AMPLICOR™ HIV-1 MONITOR® v1.5 using specimens from East Africa

被引:31
作者
Crump, John A. [1 ,2 ,3 ,4 ]
Scott, Lesley E. [5 ]
Msuya, Emma [3 ]
Morrissey, Anne B. [1 ]
Kimaro, Ekyafyose E. [3 ]
Shao, John F. [3 ,4 ]
Stevens, Wendy S. [5 ,6 ]
机构
[1] Duke Univ, Med Ctr, Dept Med, Div Infect Dis & Int Health, Durham, NC 27710 USA
[2] Duke Univ, Duke Global Hlth Inst, Durham, NC 27710 USA
[3] Kilimanjaro Christian Med Ctr, Moshi, Tanzania
[4] Tumaini Univ, Kilimanjaro Christian Med Coll, Moshi, Tanzania
[5] Univ Witwatersrand, Dept Mol Med & Haematol, Sch Pathol, Fac Hlth Sci, ZA-2050 Wits, South Africa
[6] Natl Hlth Lab Serv, Johannesburg, South Africa
基金
美国国家卫生研究院;
关键词
HIV; Reverse transcriptase polymerase chain reaction; Viral load; Validation studies; Tanzania; GENETICALLY DIVERSE PANEL; DRIED BLOOD SPOTS; VIRUS TYPE-1 RNA; QUANTITATIVE DETECTION; PERFORMANCE; NUCLISENS; PLASMA; TANZANIA; PCR; QUANTIFICATION;
D O I
10.1016/j.jviromet.2009.08.013
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The performance of the Abbott m2000rt RealTime (TM) HIV-1 assay (RealTime HIV-1) with manual sample preparation was compared against the ROCHE COBAS (R) AmpliPrep (TM)/AMPLICOR (TM) HIV-1 MONITOR Test v0.5 (CAP/CA HIV-1) using samples collected from 100 donors infected with HIV and 20 donors not infected with HIV in northern Tanzania where HIV-1 subtypes A, C, D, and their recombinant forms predominate. The RealTime HIV-1 appeared to have more within-run variability at high HIV-1 RNA concentrations, but total assay variability over the dynamic range tested was within the manufacturer's claim of < 0.3 SD copies/mL Accuracy studies showed 100% concordance for positive and negative values. When continuous values were examined, CAP/CA HIV-1 yielded higher values than the RealTime HIV-1 at higher nominal HIV-1 RNA concentrations. The RealTime HIV-1 assay showed excellent linearity between 2.5 and 7.0 log copies/mL Of negative samples, 100% showed negative results, and > 95% of samples with nominal concentrations of 40 copies/mL were detected at >= 40 copies/mL by RealTime HIV-1. Manual sample preparation may contribute to higher total assay variability. This study suggests that the Abbott m2000rt RealTime HIV-1 assay with manual sample preparation is an acceptable and feasible alternative to the conventional ROCHE COBAS AmpliPrep/AMLICOR HIV-1 Monitor v1.5 assay and that the RealTime HIV-1 assay performs well on samples from East Africa. (c) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:218 / 222
页数:5
相关论文
共 27 条
  • [1] Performance of the Amplicor human immunodeficiency virus type 1 PCR and analysis of specimens with false-negative results
    Barlow, KL
    Tosswill, JHC
    Parry, JV
    Clewley, JP
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (11) : 2846 - 2853
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/ml
    Collins, ML
    Irvine, B
    Tyner, D
    Fine, E
    Zayati, C
    Chang, CA
    Horn, T
    Ahle, D
    Detmer, J
    Shen, LP
    Kolberg, J
    Bushnell, S
    Urdea, MS
    Ho, DD
    [J]. NUCLEIC ACIDS RESEARCH, 1997, 25 (15) : 2979 - 2984
  • [4] Multicenter evaluation of the NucliSens EasyQ HIV-1 v1.1 assay for the quantitative detection of HIV-1 RNA in plasma
    de Mendoza, C
    Koppelman, M
    Montès, B
    Ferre, V
    Soriano, V
    Cuypers, H
    Segondy, M
    Oosterlaken, T
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2005, 127 (01) : 54 - 59
  • [5] Comparison of NucliSens and Roche monitor assays for quantitation of levels of human immunodeficiency virus type 1 RNA in plasma
    Dyer, JR
    Pilcher, CD
    Shepard, R
    Schock, J
    Eron, JJ
    Fiscus, SA
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (02) : 447 - 449
  • [6] HIV-1 viral load assays for resource-limited settings: Authors' reply
    Fiscus, Susan A.
    [J]. PLOS MEDICINE, 2006, 3 (12) : 2461 - 2461
  • [7] Comparative evaluation of the performance of the Abbott real-time human immunodeficiency virus type I (HIV-1) assay for measurement of HIV-1 plasma viral load following automated specimen preparation
    Garcia-Diaz, A.
    Clewley, G. S.
    Booth, C. L.
    Labett, W.
    McAllister, N.
    Geretti, A. M.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2006, 44 (05) : 1788 - 1791
  • [8] Correlation between Human Immunodeficiency Virus Type 1 (HIV-1) RNA Measurements Obtained with Dried Blood Spots and Those Obtained with Plasma by Use of Nuclisens EasyQ HIV-1 and Abbott RealTime HIV Load Tests
    Garrido, Carolina
    Zahonero, Natalia
    Corral, Angelica
    Arredondo, Miguel
    Soriano, Vincent
    de Mendoza, Carmen
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2009, 47 (04) : 1031 - 1036
  • [9] Failure to detect a non-B HIV-1 subtype by the HIV-1 Amplicor Monitor test, version 1.5: a case of unexpected vertical transmission
    Geelen, S
    Lange, J
    Borleffs, A
    Wolfs, T
    Weersink, A
    Schuurman, R
    [J]. AIDS, 2003, 17 (05) : 781 - 782
  • [10] A new ultrasensitive assay for quantitation of HIV-1 RNA in plasma
    Johanson, J
    Abravaya, K
    Caminiti, W
    Erickson, D
    Flanders, R
    Leckie, G
    Marshall, E
    Mullen, C
    Ohhashi, Y
    Perry, R
    Ricci, J
    Salituro, J
    Smith, A
    Tang, N
    Vi, M
    Robinson, J
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2001, 95 (1-2) : 81 - 92