Numerical investigation on stress intensity around Bone-Implant interface by 3-Dimensional FEA and experimental verification by optical technique

被引:12
|
作者
Dhatrak, Pankaj [1 ]
Shirsat, Uddhav [2 ]
Sumanth, S. [3 ]
Deshmukh, Vijay [4 ]
机构
[1] Dr Vishwanath Karad MIT WPU, MAEERS MIT, Sch Mech Engn, Pune 38, Maharashtra, India
[2] Bhivarabai Swanat Coll Engn & Researach, Dept Mech Engn, Pune 41, Maharashtra, India
[3] MA Rangoonwal Coll Dent Sci & Res Ctr, Dept Periodont, Pune 01, Maharashtra, India
[4] Int Clin Dent Res Org, Pune 38, Maharashtra, India
关键词
Cancellous bone; Dental Implant; FEA; Optical Technique; Photoelastic stress;
D O I
10.1016/j.matpr.2020.06.097
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The aim of the current study is to examine numerically the stress intensity along bone-implant interface for selected implant systems under occlusal load using finite element analysis (FEA) and confirming the result experimentally by optical technique. To see the biomechanical performance at bone-implant interface the five types of implant systems are chosen for the current study. A three-dimensional (3D) computer aided design (CAD) is prepared using CreO Parametric modeling software. A non-linear contact between implant and surrounding bone is defined to analyze the stress intensity in the cancellous bone under the applied load of 100 N axial directions, 40 N laterals (Bucco-lingual) direction and 100 N oblique at 45 degree to the longitudinal axis of implant. An experimental photoelastic technique (Optical technique) is used to predict the stress intensity in a cancellous bone near the bone-implant interface using Tardy's method of compensation. A homogeneous blend of Araldite and hardener material is used to prepare a photoelastic block model. The maximum value of stress intensity is noted in Type-I implant system at apical part of implant whereas minimum value was noted in Type-III implant systemunder lateral and oblique loads. Stress intensity in cancellous bone at the bone interface is found to be more significant to the applied oblique load. (C) 2020 The Authors. Published by Elsevier Ltd.
引用
收藏
页码:35 / 41
页数:7
相关论文
共 6 条
  • [1] Stress and Strain Analysis of the Bone-Implant Interface: A Comparison of Fiber-Reinforced Composite and Titanium Implants Utilizing 3-Dimensional Finite Element Study
    Shinya, Akikazu
    Ballo, Ahmed M.
    Lassila, Lippo V. J.
    Shinya, Akiyoshi
    Narhi, Timo O.
    Vallittu, Pekka K.
    JOURNAL OF ORAL IMPLANTOLOGY, 2011, 37 : 133 - 140
  • [2] Finite Element Analysis and Experimental Investigations on Stress Distribution of Dental Implants around Implant-Bone Interface
    Dhatrak, Pankaj
    Shirsat, Uddhav
    Sumanth, S.
    Deshmukh, Vijay
    MATERIALS TODAY-PROCEEDINGS, 2018, 5 (02) : 5641 - 5648
  • [3] Effect of vertical bone loss on stress distribution at the bone-implant interface around implants of varying diameters-an in silico 3D finite element analysis
    Gupta, Shipra
    Chopra, Priyanka
    Goyal, Parveen
    Jain, Ashish
    MATERIALS TODAY-PROCEEDINGS, 2021, 45 : 4581 - 4586
  • [4] Investigation of Implant bone interface with non-invasive methods: numerical simulation, strain gauges and optical coherence tomography
    Antonie, Sergiu
    Sinescu, Cosmin
    Negrutiu, Meda
    Sticlaru, Carmen
    Negru, Radu
    Laissue, Philippe L.
    Rominu, Mihai
    Podoleanu, Adrian Gh.
    ADVANCES IN STRENGTH OF MATERIALS, 2009, 399 : 193 - 198
  • [5] Stress and Strain Patterns of 1-Piece and 2-Piece Implant Systems in Bone: A 3-Dimensional Finite Element Analysis
    Ormianer, Zeev
    Ben Amar, Arial
    Duda, Mariusz
    Marku-Cohen, Sharon
    Lewinstein, Israel
    IMPLANT DENTISTRY, 2012, 21 (01) : 39 - 45
  • [6] Effects of Impression Coping Design, Impression Technique, and Dental Undercuts on the Accuracy of Implant Impressions Assessed by 3-Dimensional Optical Scanning: An In Vitro Study
    Sabouhi, Mahmoud
    Bajoghli, Farshad
    Dakhilalian, Mansour
    Beygi, Ali
    Abolhasani, Majid
    IMPLANT DENTISTRY, 2016, 25 (02) : 238 - 246