Optimization of PET/CT image quality using the GE "Sharp IR' point-spread function reconstruction algorithm

被引:17
|
作者
Vennart, Nicholas J. [1 ,2 ]
Bird, Nicholas [1 ]
Buscombe, John [1 ]
Cheow, Heok K. [1 ]
Nowosinska, Ewa [1 ]
Heard, Sarah [1 ]
机构
[1] Cambridge Univ Hosp Fdn Trust, Dept Nucl Med, Cambridge, England
[2] Queen Elizabeth Hosp, Dept Med Phys, Gateshead NE9 6SX, England
关键词
contrast; Discovery; 690; image quality; PET; CT; point-spread function; signal-to-noise; time-of-flight; TIME-OF-FLIGHT; PERFORMANCE; SCANNER; IMPACT; IMPROVEMENT; NOISE; PSF;
D O I
10.1097/MNM.0000000000000669
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThe objective of this study was to quantify any improvement with the GE Sharp IR' point-spread function (PSF) reconstruction algorithm in addition to ordered subsets expectation maximum (OSEM) and time-of-flight (TOF) reconstruction algorithms and establish the optimum parameters to be used in clinical studies.Materials and methodsWe conducted a range of experiments using the National Electrical Manufacturers Association image quality phantom filled with a 4:1 signal-to-background ratio. We scanned the phantom using the GE Discovery 690 PET/CT scanner. We varied iteration number and Gaussian filtration. Results were compared for OSEM, OSEM+TOF and OSEM+TOF+PSF reconstructions. A sample of 15 whole-body fluorine-18-fluorodeoxyglucose were reconstructed with OSEM+TOF and OSEM+TOF+PSF using a selection of optimum reconstruction parameters determined in phantom studies. Clinicians qualitatively ranked their preferred images to choose optimum parameters.ResultsThe addition of PSF improved signal-to-noise ratios (SNRs), contrast, hot contrast recovery coefficients and noise over OSEM and OSEM+TOF reconstruction algorithms. SNRs were the highest at two iterations and with 0 or 2mm filters with OSEM+TOF+PSF reconstruction in all phantom studies. Clinicians generally favoured OSEM+TOF+PSF reconstruction with three iterations and a 2mm filter.ConclusionPSF reconstruction significantly improved image quality for both clinical and phantom studies. We recommended the optimum reconstruction parameters using three iterations, 24 subsets and a 2mm filter, which improved SNRs by up to 28.8% for small lesions (P<0.05).
引用
收藏
页码:471 / 479
页数:9
相关论文
共 50 条
  • [31] Low-dose High-resolution 18F-FDG-PET/CT Using Time-of-flight and Point-spread Function Reconstructions: A Role in the Detection of Breast Carcinoma Axillary Lymph Node Metastases
    Ferdova, Eva
    Baxa, Jan
    Narsanska, Andrea
    Hes, Ondrej
    Finek, Jindrich
    Topolcan, Onrej
    Ferda, Jiri
    ANTICANCER RESEARCH, 2018, 38 (07) : 4145 - 4148
  • [32] The association of tumor-to-background ratios and SUVmax deviations related to point spread function and time-of-flight F18-FDG-PET/CT reconstruction in colorectal liver metastases
    Rogasch, Julian M. M.
    Steffen, Ingo G.
    Hofheinz, Frank
    Grosser, Oliver S.
    Furth, Christian
    Mohnike, Konrad
    Hass, Peter
    Walke, Mathias
    Apostolova, Ivayla
    Amthauer, Holger
    EJNMMI RESEARCH, 2015, 5 : 1 - 9
  • [33] Analysis of image quality by regulating beta function of BSREM reconstruction algorithm and comparison with conventional reconstructions in carcinoma breast studies of PET CT with BGO detector
    Dwivedi, Pooja
    Sawant, Viraj
    Vajarkar, Vishal
    Vatsa, Rakhee
    Choudhury, Sayak
    Jha, Ashish Kumar
    Rangarajan, Venkatesh
    NUCLEAR MEDICINE COMMUNICATIONS, 2023, 44 (01) : 56 - 64
  • [34] Dose reduction and image quality in CT examinations using an iterative reconstruction algorithm: a phantom study
    Guariglia, S.
    Meliado, G.
    Zivelonghi, E.
    Pinali, L.
    Montemezzi, S.
    Cavedon, C.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2015, 1 (04):
  • [35] Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A phantom study
    Dabli, Djamel
    Frandon, Julien
    Belaouni, Asmaa
    Akessoul, Philippe
    Addala, Takieddine
    Berny, Laure
    Beregi, Jean-Paul
    Greffer, Joel
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2022, 103 (01) : 31 - 40
  • [36] Evaluation of spatial dependence of point spread function-based PET reconstruction using a traceable point-like 22Na source
    Taisuke Murata
    Kenta Miwa
    Noriaki Miyaji
    Kei Wagatsuma
    Tomoyuki Hasegawa
    Keiichi Oda
    Takuro Umeda
    Takashi Iimori
    Yoshitada Masuda
    Takashi Terauchi
    Mitsuru Koizumi
    EJNMMI Physics, 3
  • [37] Iterative EM Reconstruction of Cardiac Small Animal PET Images Using System Point Spread Function Modelling and MAP with Anatomical Priors
    Spinelli, Antonello E.
    Fiacchi, Giacomo
    D'Ambrosio, Danicla
    Boschi, Stefano
    Franchi, Roberto
    Marengo, Mario
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4413 - +
  • [38] Characterization and optimization of image quality as a function of reconstruction algorithms and parameter settings in a Siemens Inveon small-animal PET scanner using the NEMA NU 4-2008 standards
    Visser, Eric P.
    Disselhorst, Jonathan A.
    van Lier, Monique G. J. T. B.
    Laverman, Peter
    de Jong, Gabie M.
    Oyen, Wim J. G.
    Boerman, Otto C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 629 (01) : 357 - 367
  • [39] The value of Bayesian penalized likelihood reconstruction for improving lesion conspicuity of malignant lung tumors on 18F-FDG PET/CT: comparison with ordered subset expectation maximization reconstruction incorporating time-of-flight model and point spread function correction
    Kurita, Yoshie
    Ichikawa, Yasutaka
    Nakanishi, Toshihiro
    Tomita, Yoya
    Hasegawa, Daisuke
    Murashima, Shuichi
    Hirano, Tadanori
    Sakuma, Hajime
    ANNALS OF NUCLEAR MEDICINE, 2020, 34 (04) : 272 - 279
  • [40] Image Quality and Activity Optimization in Oncologic 18F-FDG PET Using the Digital Biograph Vision PET/CT System
    van Sluis, Joyce
    Boellaard, Ronald
    Dierckx, Rudi A. J. O.
    Stormezand, Gilles N.
    Glaudemans, Andor W. J. M.
    Noordzij, Walter