Optimization of PET/CT image quality using the GE "Sharp IR' point-spread function reconstruction algorithm

被引:17
|
作者
Vennart, Nicholas J. [1 ,2 ]
Bird, Nicholas [1 ]
Buscombe, John [1 ]
Cheow, Heok K. [1 ]
Nowosinska, Ewa [1 ]
Heard, Sarah [1 ]
机构
[1] Cambridge Univ Hosp Fdn Trust, Dept Nucl Med, Cambridge, England
[2] Queen Elizabeth Hosp, Dept Med Phys, Gateshead NE9 6SX, England
关键词
contrast; Discovery; 690; image quality; PET; CT; point-spread function; signal-to-noise; time-of-flight; TIME-OF-FLIGHT; PERFORMANCE; SCANNER; IMPACT; IMPROVEMENT; NOISE; PSF;
D O I
10.1097/MNM.0000000000000669
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThe objective of this study was to quantify any improvement with the GE Sharp IR' point-spread function (PSF) reconstruction algorithm in addition to ordered subsets expectation maximum (OSEM) and time-of-flight (TOF) reconstruction algorithms and establish the optimum parameters to be used in clinical studies.Materials and methodsWe conducted a range of experiments using the National Electrical Manufacturers Association image quality phantom filled with a 4:1 signal-to-background ratio. We scanned the phantom using the GE Discovery 690 PET/CT scanner. We varied iteration number and Gaussian filtration. Results were compared for OSEM, OSEM+TOF and OSEM+TOF+PSF reconstructions. A sample of 15 whole-body fluorine-18-fluorodeoxyglucose were reconstructed with OSEM+TOF and OSEM+TOF+PSF using a selection of optimum reconstruction parameters determined in phantom studies. Clinicians qualitatively ranked their preferred images to choose optimum parameters.ResultsThe addition of PSF improved signal-to-noise ratios (SNRs), contrast, hot contrast recovery coefficients and noise over OSEM and OSEM+TOF reconstruction algorithms. SNRs were the highest at two iterations and with 0 or 2mm filters with OSEM+TOF+PSF reconstruction in all phantom studies. Clinicians generally favoured OSEM+TOF+PSF reconstruction with three iterations and a 2mm filter.ConclusionPSF reconstruction significantly improved image quality for both clinical and phantom studies. We recommended the optimum reconstruction parameters using three iterations, 24 subsets and a 2mm filter, which improved SNRs by up to 28.8% for small lesions (P<0.05).
引用
收藏
页码:471 / 479
页数:9
相关论文
共 50 条
  • [21] Impact of point spread function reconstruction on quantitative 18F-FDG-PET/CT imaging parameters and inter-reader reproducibility in solid tumors
    Sheikhbahaei, Sara
    Marcus, Charles
    Wray, Rick
    Rahmim, Arman
    Lodge, Martin A.
    Subramaniam, Rathan M.
    NUCLEAR MEDICINE COMMUNICATIONS, 2016, 37 (03) : 288 - 296
  • [22] Optimizing time-of-flight of PET/CT image quality via penalty (3 value in Bayesian penalized likelihood reconstruction algorithm
    Murat, H.
    Zulki, M. A. A.
    Said, M. A.
    Kechik, M. Awang
    Tahir, D.
    Karim, M. K. Abdul
    RADIOGRAPHY, 2025, 31 (01) : 343 - 349
  • [23] Evaluation of image reconstruction algorithms encompassing Time-Of-Flight and Point Spread Function modelling for quantitative cardiac PET: Phantom studies
    Presotto, L.
    Gianolli, L.
    Gilardi, M. C.
    Bettinardi, V.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2015, 22 (02) : 351 - 363
  • [24] Impact of Point Spread Function Reconstruction on Thoracic Lymph Node Staging With 18F-FDG PET/CT in Non-Small Cell Lung Cancer
    Lasnon, Charline
    Hicks, Rodney J.
    Beauregard, Jean-Mathieu
    Milner, Alvin
    Paciencia, Maria
    Guizard, Anne-Valerie
    Bardet, Stephane
    Gervais, Radj
    Lemoel, Gabriel
    Zalcman, Gerard
    Aide, Nicolas
    CLINICAL NUCLEAR MEDICINE, 2012, 37 (10) : 971 - 976
  • [25] PET image reconstruction with a system matrix containing point spread function derived from single photon incidence response
    Fan Xin
    Wang Hai-Peng
    Yun Ming-Kai
    Sun Xiao-Li
    Cao Xue-Xiang
    Liu Shuang-Quan
    Chai Pei
    Li Dao-Wu
    Liu Bao-Dong
    Wang Lu
    Wei Long
    CHINESE PHYSICS B, 2015, 24 (01)
  • [26] PET image reconstruction with a system matrix containing point spread function derived from single photon incidence response
    樊馨
    王海鹏
    贠明凯
    孙校丽
    曹学香
    刘双全
    柴培
    李道武
    刘宝东
    王璐
    魏龙
    Chinese Physics B, 2015, (01) : 546 - 554
  • [27] Advancement in PET quantification using 3D-OP-OSEM point spread function reconstruction with the HRRT
    Varrone, Andrea
    Sjoholm, Nils
    Eriksson, Lars
    Gulyas, Balazs
    Halldin, Christer
    Farde, Lars
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 (10) : 1639 - 1650
  • [28] RECONSTRUCTION OF DYNAMIC PET IMAGES USING ACCURATE SYSTEM POINT SPREAD FUNCTION MODELING: EFFECTS ON PARAMETRIC IMAGES
    D'Ambrosio, D.
    Fiacchi, G.
    Marengo, M.
    Boschi, S.
    Fanti, S.
    Spinelli, A. E.
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2010, 10 (01) : 73 - 94
  • [29] A novel figure of merit to investigate 68Ga PET/CT image quality based on patient weight and lesion size using Q.Clear reconstruction algorithm: A phantom study
    Santoro, Miriam
    Della Gala, Giuseppe
    Paolani, Giulia
    Zagni, Federico
    Civollani, Simona
    Strolin, Silvia
    Strigari, Lidia
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 106
  • [30] Impact of a Bayesian penalized likelihood reconstruction algorithm on image quality in novel digital PET/CT: clinical implications for the assessment of lung tumors
    Messerli, Michael
    Stolzmann, Paul
    Egger-Sigg, Michele
    Trinckauf, Josephine
    D'Aguanno, Stefano
    Burger, Irene A.
    von Schulthess, Gustav K.
    Kaufmann, Philipp A.
    Huellner, Martin W.
    EJNMMI PHYSICS, 2018, 5