Optimization of PET/CT image quality using the GE "Sharp IR' point-spread function reconstruction algorithm

被引:16
作者
Vennart, Nicholas J. [1 ,2 ]
Bird, Nicholas [1 ]
Buscombe, John [1 ]
Cheow, Heok K. [1 ]
Nowosinska, Ewa [1 ]
Heard, Sarah [1 ]
机构
[1] Cambridge Univ Hosp Fdn Trust, Dept Nucl Med, Cambridge, England
[2] Queen Elizabeth Hosp, Dept Med Phys, Gateshead NE9 6SX, England
关键词
contrast; Discovery; 690; image quality; PET; CT; point-spread function; signal-to-noise; time-of-flight; TIME-OF-FLIGHT; PERFORMANCE; SCANNER; IMPACT; IMPROVEMENT; NOISE; PSF;
D O I
10.1097/MNM.0000000000000669
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThe objective of this study was to quantify any improvement with the GE Sharp IR' point-spread function (PSF) reconstruction algorithm in addition to ordered subsets expectation maximum (OSEM) and time-of-flight (TOF) reconstruction algorithms and establish the optimum parameters to be used in clinical studies.Materials and methodsWe conducted a range of experiments using the National Electrical Manufacturers Association image quality phantom filled with a 4:1 signal-to-background ratio. We scanned the phantom using the GE Discovery 690 PET/CT scanner. We varied iteration number and Gaussian filtration. Results were compared for OSEM, OSEM+TOF and OSEM+TOF+PSF reconstructions. A sample of 15 whole-body fluorine-18-fluorodeoxyglucose were reconstructed with OSEM+TOF and OSEM+TOF+PSF using a selection of optimum reconstruction parameters determined in phantom studies. Clinicians qualitatively ranked their preferred images to choose optimum parameters.ResultsThe addition of PSF improved signal-to-noise ratios (SNRs), contrast, hot contrast recovery coefficients and noise over OSEM and OSEM+TOF reconstruction algorithms. SNRs were the highest at two iterations and with 0 or 2mm filters with OSEM+TOF+PSF reconstruction in all phantom studies. Clinicians generally favoured OSEM+TOF+PSF reconstruction with three iterations and a 2mm filter.ConclusionPSF reconstruction significantly improved image quality for both clinical and phantom studies. We recommended the optimum reconstruction parameters using three iterations, 24 subsets and a 2mm filter, which improved SNRs by up to 28.8% for small lesions (P<0.05).
引用
收藏
页码:471 / 479
页数:9
相关论文
共 40 条
[1]   Improvement in PET/CT Image Quality with a Combination of Point-Spread Function and Time-of-Flight in Relation to Reconstruction Parameters [J].
Akamatsu, Go ;
Ishikawa, Kaori ;
Mitsumoto, Katsuhiko ;
Taniguchi, Takafumi ;
Ohya, Nobuyoshi ;
Baba, Shingo ;
Abe, Koichiro ;
Sasaki, Masayuki .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (11) :1716-1722
[2]  
Alesina A, 2006, NBER MACROEC ANN-SER, V20, P1
[3]   Modeling and incorporation of system response functions in 3-D whole body PET [J].
Alessio, Adam M. ;
Kinahan, Paul E. ;
Lewellen, Thomas K. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (07) :828-837
[4]   Application and Evaluation of a Measured Spatially Variant System Model for PET Image Reconstruction [J].
Alessio, Adam M. ;
Stearns, Charles W. ;
Tong, Shan ;
Ross, Steven G. ;
Kohlmyer, Steve ;
Ganin, Alex ;
Kinahan, Paul E. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (03) :938-949
[5]   Physical Performance of the new hybrid PET/CT Discovery-690 [J].
Bettinardi, V. ;
Presotto, L. ;
Rapisarda, E. ;
Picchio, M. ;
Gianolli, L. ;
Gilardi, M. C. .
MEDICAL PHYSICS, 2011, 38 (10) :5394-5411
[6]  
BUDINGER TF, 1983, J NUCL MED, V24, P73
[7]  
Casey M., 2007, CISC VIS NETW IND GL
[8]  
Cherry SR, 2006, J NUCL MED, V47, P1735
[9]   Effect of randoms on signal-to-noise ratio in TOF PET [J].
Conti, Maurizio .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) :1188-1193
[10]  
Daube-Witherspoon ME, 2002, J NUCL MED, V43, P1398