Characterization of uniform ultrathin layer for z-response measurements in three-dimensional section fluorescence microscopy

被引:7
作者
Vicidomini, G.
Schneider, M.
Bianchini, P.
Krol, S.
Szellas, T.
Diaspro, A. [1 ]
机构
[1] FIRC Inst Mol Oncol Fdn, IFOM, I-20139 Milan, Italy
[2] Univ Genoa, Dept Phys, LAMBS, MicroSCoBIO Res Ctr, I-16146 Genoa, Italy
[3] Univ Genoa, Dept Comp Sci, I-16146 Genoa, Italy
[4] Univ Genoa, Dept Phys, I-16146 Genoa, Italy
[5] Univ Saarland, Dept Biopharmaceut & Pharmaceut Technol, D-66041 Saarbrucken, Germany
[6] Leica Microsyst, D-68165 Mannheim, Germany
关键词
4Pi microscopy; confocal microscopy; optical sectioning; shift-variant system; two-photon excitation microscopy; ultrathin uniform fluorescent PE layer; z-response;
D O I
10.1111/j.1365-2818.2007.01718.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
Layer-by-layer technique is used to adsorb a uniform ultrathin layer of fluorescently labelled polyelectrolytes on a glass cover slip. Due to their thickness, uniformity and fluorescence properties, these ultrathin layers may serve as a simple and applicable standard to directly measure the z-response of different scanning optical microscopes. In this work we use ultrathin layers to measure the z-response of confocal, two-photon excitation and 4Pi laser scanning microscopes. Moreover, due to their uniformity over a wide region, i.e. cover slip surface, it is possible to quantify the z-response of the system over a full field of view area. This property, coupled with a bright fluorescence signal, enables the use of polyelectrolyte layers for representation on sectioned imaging property charts: a very powerful method to characterize image formation properties and capabilities (z-response, off-axis aberration, spherical aberration, etc.) of a three-dimensional scanning system. The sectioned imaging property charts method needs a through-focus dataset taken from such ultrathin layers. Using a comparatively low illumination no significant bleaching occurs during the excitation process, so it is possible to achieve long-term monitoring of the z-response of the system. All the above mentioned properties make such ultrathin layers a suitable candidate for calibration and a powerful tool for real-time evaluation of the optical sectioning capabilities of different three-dimensional scanning systems especially when coupled to sectioned imaging property charts.
引用
收藏
页码:88 / 95
页数:8
相关论文
共 22 条
[11]  
HANNINEN PE, 1995, APPL PHYS LETT, V66, P1698, DOI 10.1063/1.113897
[12]   PROPERTIES OF A 4PI CONFOCAL FLUORESCENCE MICROSCOPE [J].
HELL, S ;
STELZER, EHK .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1992, 9 (12) :2159-2166
[13]   FUNDAMENTAL IMPROVEMENT OF RESOLUTION WITH A 4PI-CONFOCAL FLUORESCENCE MICROSCOPE USING 2-PHOTON EXCITATION [J].
HELL, S ;
STELZER, EHK .
OPTICS COMMUNICATIONS, 1992, 93 (5-6) :277-282
[14]   Toward fluorescence nanoscopy [J].
Hell, SW .
NATURE BIOTECHNOLOGY, 2003, 21 (11) :1347-1355
[15]  
Hooke R., 1665, MICROGRAPHIA
[16]   Nanoscale morphology of polyelectrolyte self-assembled films probed by scanning force and near-field scanning optical microscopy [J].
Lowman, GM ;
Buratto, SK .
THIN SOLID FILMS, 2002, 405 (1-2) :135-140
[17]   Polyelectrolytes I: Polyanion/polycation multilayers at the air/monolayer/water interface as elements for quantitative polymer adsorption studies and preparation of hetero-superlattices on solid surfaces [J].
Ruths, J ;
Essler, F ;
Decher, G ;
Riegler, H .
LANGMUIR, 2000, 16 (23) :8871-8878
[18]  
Schrader M, 1998, J MICROSC-OXFORD, V191, P135, DOI 10.1046/j.1365-2818.1998.00361.x
[19]   Three-dimensional super-resolution with a 4Pi-confocal microscope using image restoration [J].
Schrader, M ;
Hell, SW ;
van der Voort, HTM .
JOURNAL OF APPLIED PHYSICS, 1998, 84 (08) :4033-4042
[20]  
SHEPPARD CJR, 1980, OPTIK, V55, P331