Surgical techniques: how I do it? The Reamer/Irrigator/Aspirator (RIA) System

被引:68
作者
Giannoudis, P. V. [1 ]
Tzioupis, C. [1 ]
Green, J. [1 ]
机构
[1] Leeds Teaching Hosp NHS Trust, Leeds Gen Infirmanr, Clarendon Wing, Dept Trauma & Orthopaed,Acad Unit, Leeds LS1 3EX, W Yorkshire, England
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2009年 / 40卷 / 11期
关键词
RIA; REAMER-IRRIGATOR-ASPIRATOR; TIBIAL SHAFT FRACTURES; PROXIMAL FEMORAL FRACTURES; INTRAMEDULLARY PRESSURE; FAT-EMBOLISM; ILIAC CREST; BONE-GRAFT; OSTEOSYNTHESIS; TRAUMA; FEMUR;
D O I
10.1016/j.injury.2009.07.070
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Although reamed intramedullary nailing has long been considered a safe procedure, pulmonary complications have been reported in some groups of patients Concerns over fat embolisation, adult respiratory distress syndrome(ARDS),and sudden intraoperative death have prompted sonic authors to question whether the benefits of reaming are Outweighed by its potential adverse effects to the patient In response to these complications. recent evaluations have focused oil alternative systems that reduce intramedullary, pressure during reaming The Reamer/Irrigator/Aspirator (RIA) system (Synthes, Inc., West Chester, PA) was developed as a simultaneous reaming and aspiration system to reduce the intramedullary pressure. heat generation. operating time. and systemic effects of reaming. It was designed with an aggressive one-pass reamer head which is connected to a hand-held reamer via a drive shaft, simultaneously Irrigating and aspirating femoral canal Contents during reaming. thus preventing thermal necrosis and fat embolisation After its early use, investigators examined the "aspirate" captured by the RIA which appeared to be loaded with osteogenic Substrates This discovery revealed a second potential use for the RIA-a bone graft harvester. The different design characteristics and technicalities of the RIA system necessitate a deep and clear understanding of its technique, a meticulous preoperative planning and a strict adherence to its principles intraoperatively. We present Our current approach to the use of the RIA system, including the preoperative planning, the details of the operative procedure, the pitfalls and the potential complications that could be encountered (C) 2009 Elsevier Ltd All rights reserved
引用
收藏
页码:1231 / 1236
页数:6
相关论文
共 50 条
  • [11] Expandable nailing system for tibial shaft fractures
    Fortis, A. P.
    Dimas, A.
    Lamprakis, A. A.
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2008, 39 (08): : 940 - 946
  • [12] Open intramedullary nailing in neglected femoral diaphyseal fractures
    Gahukamble, Abhay
    Nithyananth, Manasseh
    Venkatesh, K.
    Amritanand, Rohit
    Cherian, V. M.
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2009, 40 (02): : 209 - 212
  • [13] Fat embolism: the reaming controversy
    Giannoudis, Peter V.
    Tzioupis, Christopher
    Pape, Hans-Christoph
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2006, 37 : S50 - S58
  • [14] EARLY OSTEOSYNTHESIS AND PROPHYLACTIC MECHANICAL VENTILATION IN THE MULTI-TRAUMA PATIENT
    GORIS, RJA
    GIMBRERE, JSF
    VANNIEKERK, JLM
    SCHOOTS, FJ
    BOOY, LHD
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1982, 22 (11) : 895 - 903
  • [15] Cerebral emboli and cognitive function after intramedullary fracture fixation
    Gray, Andrew C.
    Torrens, Lorna
    Christie, James
    Graham, Catriona
    Robinson, C. Michael
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2009, 40 (07): : 742 - 745
  • [16] Effect of re-implanted particles from intramedullary reaming on mechanical properties and callus formation
    Hammer, T. O.
    Wieling, R.
    Green, J. M.
    Suedkamp, N. P.
    Schneider, E.
    Mueller, C. A.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (11): : 1534 - 1538
  • [17] Intramedullary osteosynthesis of complex proximal femoral fractures with the Targon PF nail
    Heinert, G.
    Parker, M. J.
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2007, 38 (11): : 1294 - 1299
  • [18] Cortical heat generation using an irrigating/aspirating single-pass reaming vs conventional stepwise reaming
    Higgins, Thomas E.
    Casey, Virginia
    Bachus, Kent
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (03) : 192 - 197
  • [19] Fat extravasation due to unreamed and experimentally reamed intramedullary nailing of the sheep femur
    Hoegel, Florian
    Kamer, Lukas
    Schlegel, Urs
    Rahn, Berton
    Suedkamp, Norbert P.
    Mueller, Christof A.
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2009, 40 (07): : 718 - 721
  • [20] The influence of a one-step reamer-irrigator-aspirator technique on the intramedullary pressure in the pig femur
    Husebye, Elisabeth Ellingsen
    Lyberg, Torstein
    Madsen, Jan Erik
    Eriksen, Morten
    Roise, Olav
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2006, 37 (10): : 935 - 940