National survey of doses from CT in the UK: 2003

被引:437
作者
Shrimpton, P. C. [1 ]
Hillier, M. C.
Lewis, M. A.
Dunn, M.
机构
[1] Hlth Protect Agcy, Radiat Protect Div, Didcot OX11 0RQ, Oxon, England
[2] Univ London St Georges Hosp, Dept Med Phys, ImPACT, London SW17 0QT, England
[3] Queens Med Ctr, Dept Med Phys, Nottingham NG7 2UH, England
关键词
D O I
10.1259/bjr/93277434
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A review of patient doses from CT examinations in the UK for 2003 has been conducted on the basis of data received from over a quarter of all UK scanners, of which 37% had multislice capability. Questionnaires were employed to collect scan details both for the standard protocols established at each scanner for 12 common types of CT examination on adults and children, and for samples of individual patients. This information was combined with published scanner-specific CT dose index (CTDI) coefficients to estimate values of the standard dose indices CTDIw and CTDIvol for each scan sequence. Knowledge of each scan length allowed assessment of the dose-length product (DLP) for each examiniation, from which effective doses were then estimated. When compared with a previous UK survey for 1991, wide variations were still apparent between CT centres in the doses for standard protocols. The mean UK doses for adult patients were in general lower by up to 50% than those for 1991, although doses were slightly higher for multislice (4+) (MSCT) relative to single slice (SSCT) scanners. Values of CTDIvol for MSCT were broadly similar to European survey data for 2001. The third quartile values of these dose distributions have been used to derive UK national reference doses for examinations on adults (separately for SSCT and MSCT) and children as initial tools for promoting patient protection. The survey has established the PREDICT (Patient Radiation Exposure and Dose in CT) database as a sustainable national resource for monitoring dose trends in CT through the ongoing collation of further survey data.
引用
收藏
页码:968 / 980
页数:13
相关论文
共 45 条
[1]  
[Anonymous], 2000, COMPUTED TOMOGRAPHY
[2]  
[Anonymous], CT PAT DOS EXC SPREA
[3]  
[Anonymous], 88 IPEM
[4]   Dose reduction in pediatric CT: A rational approach [J].
Boone, JM ;
Geraghty, EM ;
Seibert, JA ;
Wootton-Gorges, SL .
RADIOLOGY, 2003, 228 (02) :352-360
[5]   Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey [J].
Brix, G ;
Nagel, HD ;
Stamm, G ;
Veit, R ;
Lechel, U ;
Griebel, J ;
Galanski, M .
EUROPEAN RADIOLOGY, 2003, 13 (08) :1979-1991
[6]  
*BUND STRAHL, 2003, BEK DIAGN REF RAD NU
[7]  
*DEP HLTH, 2003, FORM KH12 NUMB IM RA
[8]  
European Commission, 1999, 16262 EUR EN EUR COM
[9]   Improved pediatric multidetector body CT using a size-based color-coded format [J].
Frush, DP ;
Soden, B ;
Frush, KS ;
Lowry, C .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 178 (03) :721-726
[10]   Radiation dose in CT: are we meeting the challenge? [J].
Golding, SJ ;
Shrimpton, PC .
BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (889) :1-4