Measurements of the Modulation Transfer Function, Normalized Noise Power Spectrum and Detective Quantum Efficiency for two flat panel detectors: A Fluoroscopic and a Cone Beam Computer Tomography Flat Panel Detectors

被引:6
作者
Benitez, Ricardo Betancourt [1 ,2 ]
Ning, Ruola [1 ]
Conover, David [1 ]
Liu, Shaohua [1 ]
机构
[1] Univ Rochester, Med Ctr, Dept Imaging Sci, Rochester, NY 14642 USA
[2] Univ Rochester, Dept Phys & Astron, Rochester, NY 14627 USA
关键词
Cone Beam Computer Tomography; Modulation Transfer Function; Edge Spread Function; Normalized Noise Power Spectrum; Detective Quantum Efficiency; Fluoroscopy Flat Panel Detector; Cone Beam Flat Panel Detector; X-RAY-DETECTOR; DIGITAL MAMMOGRAPHY; PHYSICAL-CHARACTERISTICS; PERFORMANCE; CT; SYSTEM; ANGIOGRAPHY; IMAGERS; DQE(F);
D O I
10.3233/XST-2009-0234
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The physical performance of two Flat Panel Detectors has been evaluated. The first Flat Panel Detector is for Fluoroscopic applications, Varian PaxScan 2520, and the second is for Cone Beam Computer Tomography applications, Varian PaxScan 4030CB. First, the spectrum of the X-ray source was measured. Second, the linearity of the detectors was investigated by using an ionization chamber and the average ADU values of the detectors. Third, the temporal resolution was characterized by evaluating their image lag. Fourth, their spatial resolution was characterized by the pre-sampling Modulation Transfer Function. Fifth, the Normalized Noise Power Spectrum was calculated for various exposures levels. Finally, the Detective Quantum Efficiency was obtained as a function of spatial frequency and entrance exposure. The results illustrate that the physical performance in Detective Quantum Efficiency and Normalized Noise Power Spectrum of the Cone Beam Computer Tomography detector is superior to that of the fluoroscopic detector whereas the latter detector has a higher spatial resolution as demonstrated by larger values of its Modulation Transfer Function at large spatial frequencies.
引用
收藏
页码:279 / 293
页数:15
相关论文
共 6 条
  • [1] Comparison measurements of DQE for two flat panel detectors:: Fluoroscopic detector vs. cone beam CT detector
    Benitez, Ricardo Betancourt
    Ning, Ruola
    Conover, David
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [2] Detective quantum efficiency of direct, flat panel x-ray imaging detectors for fluoroscopy
    Hunt, DC
    Zhao, W
    Rowlands, JA
    PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 408 - 417
  • [3] Detective quantum efficiency of an x-ray image intensifier chain as a benchmark for amorphous silicon flat panel detectors
    Koch, A
    Macherel, JM
    Wirth, T
    de Groot, P
    Ducourant, T
    Couder, D
    Moy, JP
    Calais, E
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 115 - 120
  • [4] Signal, noise power spectrum, and detective quantum efficiency of indirect-detection flat-panel imagers for diagnostic radiology
    Siewerdsen, JH
    Antonuk, LE
    El-Mohri, Y
    Yorkston, J
    Huang, W
    Cunningham, IA
    MEDICAL PHYSICS, 1998, 25 (05) : 614 - 628
  • [5] Effect of ghosting on the modulation transfer function of amorphous selenium based flat panel detectors - art. no. 61421S
    Chowdhury, Lamia
    Tousignant, Olivier
    DeCrescenzo, Giovanni
    Gauthier, P.
    Leboeuf, J.
    Rowlands, J. A.
    Medical Imaging 2006: Physics of Medical Imaging, Pts 1-3, 2006, 6142 : S1421 - S1421
  • [6] Effect of depth dependent modulation transfer function and K-fluorescence reabsorption on the detective quantum efficiency of indirect conversion flat panel x-ray imaging systems using CsI
    Rowlands, JA
    Ji, WG
    Zhao, W
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 257 - 267