Measurement of serum 17-hydroxyprogesterone using isotope dilution liquid chromatography-tandem mass spectrometry candidate reference method and evaluation of the performance for three routine methods

被引:4
|
作者
Long, Qichen [1 ]
Zhang, Tianjiao [2 ]
Yan, Ying [2 ]
Zhao, Haijian [2 ]
Zhou, Weiyan [2 ]
Zeng, Jie [2 ]
Li, Shuijun [3 ]
Zhang, Jiangtao [2 ]
Zeng, Qingzhang [1 ]
Zhao, Beibei [4 ]
Zhang, Chuanbao [1 ]
Chen, Wenxiang [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Chinese Acad Med Sci, Beijing Hosp,Beijing Engn Res Ctr Lab Med, Natl Ctr Gerontol,Inst Geriatr Med,Natl Ctr Clin, 1 Dahua Rd, Beijing 100730, Peoples R China
[2] Chinese Acad Med Sci, Beijing Hosp, Natl Ctr Clin Labs, Natl Ctr Gerontol,Inst Geriatr Med,Beijing Engn R, Beijing, Peoples R China
[3] Shanghai Xuhui Cent Hosp, Cent Lab, Shanghai, Peoples R China
[4] Guangzhou KingMed Diagnost Grp Co Ltd, Guangzhou, Peoples R China
关键词
congenital adrenal hyperplasia; liquid chromatography/tandem mass spectrometry; method evaluation; reference measurement procedure; serum; 17-hydroxyprogesterone; LC-MS/MS; TESTOSTERONE; 17-ALPHA-HYDROXYPROGESTERONE; ANDROSTENEDIONE; QUANTIFICATION; SPECIFICATIONS; DEFICIENCY; CHILDREN;
D O I
10.1515/cclm-2020-0410
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Accurate measurements of serum 17-hydroxyprogesterone (17OHP) are essential for diagnosis and treatment monitoring for congenital adrenal hyperplasia patients. The performance of serum 17OHP routine methods remains highly variable that calls for a candidate reference measurement procedure (cRMP) to improve the standardization of serum 17OHP measurements. Methods: Serum samples spiked with internal standards were extractedwith a combination of solid-phase extraction and liquid-liquid extraction. The 17OHP was quantified by the isotope dilution coupled with liquid chromatography/ tandem mass spectrometry (ID-LC/MS/MS) with electrospray ionization in positive ion mode. Nine structural analogs of 17OHP were evaluated for interferences. The precision and analytical recovery were assessed. Twenty native and 40 spiked serum for performance evaluation were measured by the cRMP and two clinical LC/MS routine methods. Results: No apparent interferences were found with the 17OHP measurement. The within-run, between-run, and total precision for our method were 0.4-0.8%, 0.6-2.0%, and 1.0-2.1% for four pooled serum (2.46-102.72 nmol/L), respectively. The recoveries of added 17OHP were 100.0-100.2%. For the performance of two LC/MS routine methods, they showed relative deviation ranges of -22.1 to 1.1% and -6.7 to 12.8%, respectively. Conclusions: We developed and validated a reliable serum 17OHP method using ID-LC/MS/MS. The desirable accuracy and precision of this method enable it to serve as a promising cRMP to improve the standardization for serum17OHP routine measurements.
引用
收藏
页码:523 / 532
页数:10
相关论文
共 50 条
  • [11] An isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for aldosterone measurement in human plasma
    Zhang, Qiaoxuan
    Han, Liqiao
    Zheng, Songbai
    Ouyang, Fen
    Wu, Xiaobin
    Yan, Jun
    Zhan, Min
    Ke, Peifeng
    Zhuang, Junhua
    Huang, Xianzhang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2021, 413 (17) : 4471 - 4481
  • [12] Development and evaluation of a candidate reference method for the determination of total cortisol in human serum using isotope dilution liquid chromatography/mass spectrometry and liquid chromatography/tandem mass spectrometry
    Tai, SSC
    Welch, MJ
    ANALYTICAL CHEMISTRY, 2004, 76 (04) : 1008 - 1014
  • [13] Quantification of mycophenolic acid in plasma by isotope dilution liquid chromatography-tandem mass spectrometry candidate reference method
    Lin, Jing
    Shen, Min
    Yu, Ting
    Wang, Huimin
    Pan, Jingjue
    Huang, Gaipeng
    Li, Quanle
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2025, 417 (08) : 1607 - 1617
  • [14] A candidate reference measurement procedure for quantification of glycocholic acid in human serum based on isotope dilution liquid chromatography-tandem mass spectrometry
    Zhang, Pingping
    Wang, Huimin
    Liang, Man
    Wang, Zhifang
    Liu, Chunlong
    Han, Yanlin
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (23) : 5121 - 5131
  • [15] Development and evaluation of a reference measurement procedure for the determination of estradiol-17β in human serum using isotope-dilution liquid chromatography-tandem mass spectrometry
    Tai, SSC
    Welsh, MJ
    ANALYTICAL CHEMISTRY, 2005, 77 (19) : 6359 - 6363
  • [16] Endogenous isobaric interference on serum 17 hydroxyprogesterone by liquid chromatography-tandem mass spectrometry methods
    Lo, Clara Wai Shan
    Hoad, Kirsten
    Loh, Tze Ping
    van den Berg, Sjoerd
    Cooke, Brian R.
    Greaves, Ronda F.
    Hartmann, Michaela F.
    Wudy, Stefan A.
    Ho, Chung Shun
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (03) : E64 - E66
  • [17] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Susan S.-C. Tai
    Bei Xu
    Michael J. Welch
    Karen W. Phinney
    Analytical and Bioanalytical Chemistry, 2007, 388 : 1087 - 1094
  • [18] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    Phinney, Karen W.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 388 (5-6) : 1087 - 1094
  • [19] Determination of 17α-hydroxyprogesterone in serum by liquid chromatography-tandem mass spectrometry and immunoassay
    Turpeinen, U
    Itkonen, O
    Ahola, L
    Stenman, UH
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2005, 65 (01): : 3 - 12
  • [20] Determination of serum cortisol using isotope dilution–liquid chromatography–mass spectrometry as a candidate reference method
    Pyong Gil Jung
    Byungjoo Kim
    Sang-Ryoul Park
    Hun-Young So
    Lian Hua Shi
    Yongseong Kim
    Analytical and Bioanalytical Chemistry, 2004, 380 : 782 - 788