Prospective evaluation of hepatitis C virus antibody detection in whole blood collected on dried blood spots with the INNOTEST? HCV Ab IV enzyme immunoassay

被引:4
作者
Vetter, Beatrice N. [1 ]
Reipold, Elena Ivanova [1 ]
Ongarello, Stefano [1 ]
Fajardo, Emmanuel [1 ]
Tyshkovskiy, Alexander [2 ,3 ]
Ben, Iryna [4 ]
Vasylyev, Marta [4 ,5 ]
机构
[1] Fdn Innovat New Diagnost FIND, 9 Chemin Mines, CH-1202 Geneva, Switzerland
[2] Harvard Med Sch, Brigham & Womens Hosp, Dept Med, Div Genet, Boston, MA 02115 USA
[3] Moscow MV Lomonosov State Univ, Belozersky Inst Physicochem Biol, Moscow 119234, Russia
[4] Lviv Reg Publ Hlth Ctr, UA-29027 Lvov, Ukraine
[5] Sci Med Diagnost Ctr Astar, UA-29000 Lvov, Ukraine
关键词
Hepatitis C virus; Dried Blood spot; Sensitivity; Specificity; In vitro diagnostics; Enzyme immunoassay;
D O I
10.1016/j.jcv.2021.104783
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Introduction: Dried blood spots (DBS) have potential to improve access to screening for antibodies to hepatitis C virus (HCV). However, although several studies on off-label use of DBS have been performed, to date no HCV antibody serology test is formally approved for use with DBS. This study evaluated the performance of the INNOTEST? HCV Ab IV enzyme immunoassay in paired DBS and plasma samples, to determine whether DBS may be added to the intended use. Methods: Adults with no history of HCV treatment were prospectively enrolled from two sites in Ukraine. DBS were prepared from fingerstick whole blood (fDBS) and venous whole blood (vDBS) samples. Undiluted and serially diluted DBS and plasma samples were tested. Results: Samples from 149 HCV positive and 151 HCV negative participants were included. Sensitivity and specificity of the INNOTEST? HCV Ab IV assay were both 100 % (95 % confidence intervals 95.7?100) for samples collected on fDBS or vDBS compared with plasma as the reference standard. In all undiluted samples, negative and positive percentage agreement and overall rate of agreement were 100 % between all sample types (Cohen?s kappa coefficient of 1). In serially diluted samples, agreement was high (>95 %) between fDBS and vDBS, and as expected, positive percentage agreement between both DBS sample types and plasma was lower (>66 %). Conclusions: Performance of the INNOTEST? HCV Ab IV assay in DBS was acceptable, thus whole blood collected on DBS may represent an alternative sample type for this assay in settings where venous blood collection is not possible.
引用
收藏
页数:6
相关论文
共 19 条
[1]  
[Anonymous], 2017, WHO |
[2]   Simultaneous detection of hepatitis c virus antigen and antibodies in dried blood spots [J].
Brandao, C. P. U. ;
Marques, B. L. C. ;
Marques, V. A. ;
Villela-Nogueira, C. A. ;
Do O, K. M. R. ;
de Paula, M. T. ;
Lewis-Ximenez, L. L. ;
Lampe, E. ;
Sa Ferreira, J. A. ;
Villar, L. M. .
JOURNAL OF CLINICAL VIROLOGY, 2013, 57 (02) :98-102
[3]   Evaluation of dried blood spot samples for hepatitis C virus detection and quantification [J].
Crespo Marques, Brunna Lemos ;
do Espirito-Santo, Marcia Paschoal ;
Marques, Vanessa Alves ;
Miguel, Juliana Custodio ;
da Silva, Elisangela Ferreira ;
Villela-Nogueira, Cristiane Alves ;
Lewis-Ximenez, Lia Laura ;
Lampe, Elisabeth ;
Villar, Livia Melo .
JOURNAL OF CLINICAL VIROLOGY, 2016, 82 :139-144
[4]   Dried blood spots: Effects of less than optimal collection, shipping time, heat, and humidity [J].
Crimmins, Eileen M. ;
Zhang, Yuan S. ;
Kim, Jung Ki ;
Frochen, Stephen ;
Kang, Hyewon ;
Shim, Hyunju ;
Ailshire, Jennifer ;
Potter, Alan ;
Cofferen, Jake ;
Faul, Jessica .
AMERICAN JOURNAL OF HUMAN BIOLOGY, 2020, 32 (05)
[5]   Comparison of Hepatitis C Virus RNA and antibody detection in dried blood spots and plasma specimens [J].
Dokubo, E. Kainne ;
Evans, Jennifer ;
Winkelman, Valerie ;
Cyrus, Sherri ;
Tobler, Leslie H. ;
Asher, Alice ;
Briceno, Alya ;
Page, Kimberly .
JOURNAL OF CLINICAL VIROLOGY, 2014, 59 (04) :223-227
[6]   Performance of ANTI-HCV testing in dried blood spots and saliva according to HIV status [J].
Flores, Geane Lopes ;
Cruz, Helena Medina ;
Marques, Vanessa Alves ;
Villela-Nogueira, Cristiane Alves ;
Potsch, Denise Vigo ;
May, Silvia Beatriz ;
Brandao-Mello, Carlos Eduardo ;
Amendola Pires, Marcia Maria ;
Pilotto, Jose Henrique ;
Pollo-Flores, Priscila ;
Cathala Esberard, Eliane Bordalo ;
Ivantes, Claudia ;
Lewis-Ximenez, Lia Laura ;
Lampe, Elisabeth ;
Villar, Livia Melo .
JOURNAL OF MEDICAL VIROLOGY, 2017, 89 (08) :1435-1441
[7]   Diagnostic accuracy of serological diagnosis of hepatitis C and B using dried blood spot samples (DBS): two systematic reviews and meta-analyses [J].
Lange, Berit ;
Cohn, Jennifer ;
Roberts, Teri ;
Camp, Johannes ;
Chauffour, Jeanne ;
Gummadi, Nina ;
Ishizaki, Azumi ;
Nagarathnam, Anupriya ;
Tuaillon, Edouard ;
van de Perre, Philippe ;
Pichler, Christine ;
Easterbrook, Philippa ;
Denkinger, Claudia M. .
BMC INFECTIOUS DISEASES, 2017, 17
[8]   Diagnostic accuracy of detection and quantification of HBV-DNA and HCV-RNA using dried blood spot (DBS) samples - a systematic review and meta-analysis [J].
Lange, Berit ;
Roberts, Teri ;
Cohn, Jennifer ;
Greenman, Jamie ;
Camp, Johannes ;
Ishizaki, Azumi ;
Messac, Luke ;
Tuaillon, Edouard ;
van de Perre, Philippe ;
Pichler, Christine ;
Denkinger, Claudia M. ;
Easterbrook, Philippa .
BMC INFECTIOUS DISEASES, 2017, 17 :71-85
[9]   Performance of an antigen-antibody combined assay for hepatitis C virus testing without venipuncture [J].
Larrat, Sylvie ;
Bourdon, Cecile ;
Baccard, Monique ;
Garnaud, Cecile ;
Mathieu, Sophie ;
Quesada, Jean-Louis ;
Signori-Schmuck, Anne ;
Germi, Raphaele ;
Blanc, Myriam ;
Leclercq, Pascale ;
Hilleret, Marie-Noelle ;
Leroy, Vincent ;
Zarski, Jean-Pierre ;
Morand, Patrice .
JOURNAL OF CLINICAL VIROLOGY, 2012, 55 (03) :220-225
[10]   Dried blood spot samples: Optimization of commercial EIAs for hepatitis C antibody detection and stability under different storage conditions [J].
Marques, B. L. C. ;
Brandao, C. U. ;
Silva, E. F. ;
Marques, V. A. ;
Villela-Nogueira, C. A. ;
Do O, K. M. R. ;
de Paula, M. T. ;
Lewis-Ximenez, L. L. ;
Lampe, E. ;
Villar, L. M. .
JOURNAL OF MEDICAL VIROLOGY, 2012, 84 (10) :1600-1607