Mobile app Aston contrast sensitivity test

被引:23
作者
Kingsnorth, Alec [1 ]
Drew, Tom [1 ]
Grewal, Bikramjit [1 ]
Wolffsohn, James S. [1 ]
机构
[1] Aston Univ, Ophthalm Res Grp, Life & Hlth Sci, Birmingham, W Midlands, England
关键词
contrast sensitivity; CSV‐ 1000; mobile app; Pelli– Robson chart; repeatability; validity; TEST-RETEST; CSV-1000; AGE; DISCRIMINATION; RELIABILITY; VISTECH; VISION; ACUITY; CHARTS; ADULTS;
D O I
10.1111/cxo.12362
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background Contrast detection is an important aspect of the assessment of visual function; however, clinical tests evaluate limited spatial frequencies and contrasts. This study validates the accuracy and inter-test repeatability of a swept-frequency near and distance mobile app Aston contrast sensitivity test, which overcomes this limitation compared to traditional charts. Method Twenty subjects wearing their full refractive correction underwent contrast sensitivity testing on the new near application (near app), distance app, CSV-1000 and Pelli-Robson charts with full correction and with vision degraded by 0.8 and 0.2 Bangerter degradation foils. In addition repeated measures using the 0.8 occluding foil were taken. Results The mobile apps (near more than distance, p = 0.005) recorded a higher contrast sensitivity than printed tests (p < 0.001); however, all charts showed a reduction in measured contrast sensitivity with degradation (p < 0.001) and a similar decrease with increasing spatial frequency (interaction > 0.05). Although the coefficient of repeatability was lowest for the Pelli-Robson charts (0.14 log units), the mobile app charts measured more spatial frequencies, took less time and were more repeatable (near: 0.26 to 0.37 log units; distance: 0.34 to 0.39 log units) than the CSV-1000 (0.30 to 0.93 log units). The duration to complete the CSV-1000 was 124 +/- 37-seconds, Pelli-Robson 78 +/- 27-seconds, near app 53 +/- 15-seconds and distance app 107 +/- 36-seconds. Conclusions While there were differences between charts in contrast levels measured, the new Aston near and distance apps are valid, repeatable and time-efficient method of assessing contrast sensitivity at multiple spatial frequencies.
引用
收藏
页码:350 / 355
页数:6
相关论文
共 33 条
[1]   Contrast sensitivity, healthy aging and noise [J].
Allard, Remy ;
Renaud, Judith ;
Molinatti, Sandra ;
Faubert, Jocelyn .
VISION RESEARCH, 2013, 92 :47-52
[2]  
APKARIAN P, 1987, INVEST OPHTH VIS SCI, V28, P607
[3]   Differerces in Vision Between Clinic and Horns and the Effect of Lighting in Older Adults With and Without Glaucoma [J].
Bhorade, Anjali M. ;
Perlmutter, Monica S. ;
Wilson, Brad ;
Kambarian, Jamie ;
Chang, Sidney ;
Pekmezci, Melike ;
Gordon, Mae .
JAMA OPHTHALMOLOGY, 2013, 131 (12) :1554-1562
[4]   MEASUREMENT OF SPATIAL CONTRAST SENSITIVITY IN CASES OF BLURRED VISION ASSOCIATED WITH CEREBRAL LESIONS [J].
BODISWOLLNER, I ;
DIAMOND, SP .
BRAIN, 1976, 99 (DEC) :695-710
[5]   APPLICATION OF FOURIER ANALYSIS TO VISIBILITY OF GRATINGS [J].
CAMPBELL, FW ;
ROBSON, JG .
JOURNAL OF PHYSIOLOGY-LONDON, 1968, 197 (03) :551-&
[6]   OPTICAL AND RETINAL FACTORS AFFECTING VISUAL RESOLUTION [J].
CAMPBELL, FW ;
GREEN, DG .
JOURNAL OF PHYSIOLOGY-LONDON, 1965, 181 (03) :576-+
[7]   Differential effects of aging on spatial contrast sensitivity to linear and polar sine-wave gratings [J].
Costa, T. L. ;
Nogueira, R. M. T. B. L. ;
Pereira, A. G. F. ;
Santos, N. A. .
BRAZILIAN JOURNAL OF MEDICAL AND BIOLOGICAL RESEARCH, 2013, 46 (10) :855-860
[8]  
DEREFELDT G, 1979, ACTA OPHTHALMOL, V57, P679
[9]   NONSELECTIVE LOSS OF CONTRAST SENSITIVITY IN VISUAL-SYSTEM TESTING IN EARLY TYPE-I DIABETES [J].
DILEO, MAS ;
CAPUTO, S ;
FALSINI, B ;
PORCIATTI, V ;
MINNELLA, A ;
GRECO, AV ;
GHIRLANDA, G .
DIABETES CARE, 1992, 15 (05) :620-625
[10]   Rapid and Reliable Assessment of the Contrast Sensitivity Function on an iPad [J].
Dorr, Michael ;
Lesmes, Luis A. ;
Lu, Zhong-Lin ;
Bex, Peter J. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (12) :7266-7273