A vibrating plate fabricated by the methods of microelectromechanical systems (MEMS) for the simultaneous measurement of density and viscosity: Results for argon at temperatures between 323 and 423K at pressures up to 68 MPa

被引:25
作者
Goodwin, A. R. H.
Fitt, A. D.
Ronaldson, K. A.
Wakeham, W. A.
机构
[1] Schlumberger Doll Res Ctr, Ridgefield, CT 06877 USA
[2] Univ Southampton, Sch Math, Southampton SO17 1BJ, Hants, England
[3] Univ Southampton, Sch Engn Sci, Southampton SO17 1BJ, Hants, England
关键词
argon; density; MEMS; microelectromechanical systems technology; viscosity;
D O I
10.1007/s10765-006-0114-6
中图分类号
O414.1 [热力学];
学科分类号
摘要
In the petroleum industry, measurements of the density and viscosity of petroleum reservoir fluids are required to determine the value of the produced fluid and the production strategy. Measurements of the density and viscosity of petroleum fluids require a transducer that can operate at reservoir conditions, and results with an uncertainty of about +/- 1% in density and +/- 10% in viscosity are needed to guide value and exploitation calculations with sufficient rigor. Necessarily, these specifications place robustness as a superior priority to accuracy for the design. A vibrating plate, with dimensions of the order of 1 mm and a mass of about 0.12 mg, clamped along one edge, has been fabricated, with the methods of Microelectromechanical (MEMS) technology, to provide measurements of both density and viscosity of fluids in which it is immersed. The resonance frequency (at pressure p = 0 is about 12 kHz) and quality factor (at p = 0 is about 2800) of the first order bending (flexural) mode of the plate are combined with semi-empirical working equations, coefficients obtained by calibration, and the mechanical properties of the plate to provide the density and viscosity of the fluid into which it is immersed. When the device was surrounded by argon at temperatures between 348 and 423 K and at pressures between 20 and 68 MPa, the density and viscosity were determined with an expanded (k = 2) uncertainty, including the calibration, of about +/- 0.35% and +/- 3%, respectively. These results, when compared with accepted correlations for argon reported in the literature, were found to lie within +/- 0.8% for density and less than +/- 5% for viscosity of literature values, which are within a reasonable multiple of the relative combined expanded (k = 2) uncertainty.
引用
收藏
页码:1650 / 1676
页数:27
相关论文
共 58 条
[21]  
Johnson J. F., 1977, PHYS METHODS CHEM 6, P63
[22]  
KAWATA M, 1991, EXPT THERMODYNAMICS, V3, P51
[23]   The influence of a viscous fluid on the vibration dynamics of scanning near-field optical microscopy fiber probes and atomic force microscopy cantilevers [J].
Kirstein, S ;
Mertesdorf, M ;
Schonhoff, M .
JOURNAL OF APPLIED PHYSICS, 1998, 84 (04) :1782-1790
[24]   Bulk micromachining of silicon [J].
Kovacs, GTA ;
Maluf, NI ;
Petersen, KE .
PROCEEDINGS OF THE IEEE, 1998, 86 (08) :1536-1551
[25]  
KUNZEL W, 1987, RECOMMENDED REFERENC, P45
[26]   Accurate density measurements of reference liquids by a magnetic suspension balance [J].
Kuramoto, N ;
Fujii, K ;
Waseda, A .
METROLOGIA, 2004, 41 (02) :S84-S94
[27]   Viscosity and thermal conductivity equations for nitrogen, oxygen, argon, and air [J].
Lemmon, EW ;
Jacobsen, RT .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2004, 25 (01) :21-69
[28]  
LEMMON EW, 2005, REFERENCE FLUID PROP
[29]   Measurement of the viscosity and density of two reference fluids, with nominal viscosities at T=298 K and p=0.1 MPa of (16 and 29) mPa•s, at temperatures between (298 and 393) K and Pressures below 55 MPa [J].
Lundstrum, R ;
Goodwin, ARH ;
Hsu, K ;
Frels, M ;
Caudwell, DR ;
Trusler, JPM ;
Marsh, KN .
JOURNAL OF CHEMICAL AND ENGINEERING DATA, 2005, 50 (04) :1377-1388
[30]   Hydrodynamics of oscillating atomic force microscopy cantilevers in viscous fluids -: art. no. 074907 [J].
Maali, A ;
Hurth, C ;
Boisgard, R ;
Jai, C ;
Cohen-Bouhacina, T ;
Aimé, JP .
JOURNAL OF APPLIED PHYSICS, 2005, 97 (07)