Patient skin dose in cardiac interventional procedures: Conventional fluoroscopy versus pulsed fluoroscopy

被引:25
作者
Chida, Koichi
Fuda, Kenji
Saito, Haruo
Takai, Yoshihiro
Takahashi, Shoki
Yamada, Shogo
Kohzuki, Masahiro
Zuguchi, Masayuki
机构
[1] Tohoku Univ, Fac Med, Sch Hlth Sci, Dept Radiol Technol,Aoba Ku, Sendai, Miyagi 9808575, Japan
[2] NTT Tohoku Hosp, Dept Radiol, Sendai, Miyagi, Japan
[3] Tohoku Univ, Grad Sch Med, Dept Radiol, Sendai, Miyagi 980, Japan
[4] Tohoku Univ, Grad Sch Med, Dept Internal Med & Rehabil Sci, Sendai, Miyagi 980, Japan
关键词
catheterization; coronary intervention; angiography; radiation exposure; coronary diseases;
D O I
10.1002/ccd.20961
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To investigate whether pulsed fluoroscopy reduces a patient's exposure compared with the exposure owing to conventional (continuous) fluoroscopy, we simulated the skin radiation doses of patients at cardiac catheterization facilities with various X-ray systems used in fluoroscopically guided intervention procedures. Background: Although many reports have noted that pulsed fluoroscopy provides important further reductions in radiation exposure, it has been determined that when comparing dose rates between different vendor systems, "pulsed fluoroscopy" does not reduce patients' exposure as compared with "conventional fluoroscopy". Methods: We examined 13 X-ray systems; 10 used pulsed fluoroscopy and three used conventional fluoroscopy. The entrance surface doses with fluoroscopy were compared for the 13 X-ray systems by using acrylic plates (20-cm thick) and a skin dose monitor. The X-ray conditions used in the measurements were those normally used in the facilities performing percutaneous coronary intervention. Results: The average surface dose for systems from three different vendors producing conventional fluoroscopy systems was 23.93 +/- 2.77 mGy/min vs. an average surface dose of 22.52 +/- 4.50 mGy/min from five vendors of pulsed fluoroscopy systems (25, 30, and 50 pulses/sec) (P=0.646). The average entrance surface dose was significantly (P<0.0001) higher with conventional fluoroscopy and pulsed fluoroscopy at 25, 30, and 50 pulses/sec(23.05 +/- 3.78 mGy/min)than with pulsed fluoroscopy at 15 pulses/sec (13.86 +/- 3.22 mGy/min). Conclusions: Pulsed fluoroscopy did not in itself reduce radiation exposure. In general, the-use of pulsed fluoroscopy at a pulse rate lower than 25 pulses/sec should reduce the skin dose in fluoroscopically guided intervention procedures. Nevertheless, some X-ray systems are not designed to reduce the dose rate as the number of pulses per second is decreased. Physicians should be aware of the entrance surface dose of the X-ray system that they use for cardiac IVR. (C) 2006 Wiley-Liss, Inc.
引用
收藏
页码:115 / 121
页数:7
相关论文
共 22 条
  • [1] PERCEPTUAL COMPARISON OF PULSED AND CONTINUOUS FLUOROSCOPY
    AUFRICHTIG, R
    XUE, P
    THOMAS, CW
    GILMORE, GC
    WILSON, DL
    [J]. MEDICAL PHYSICS, 1994, 21 (02) : 245 - 256
  • [2] BAIM DS, 2000, GROSSMANS CARDIAC CA, P15
  • [3] Radiation-induced skin injuries and fluoroscopy
    Berlin, L
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (01) : 21 - 25
  • [4] Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures
    Chida, K
    Saito, H
    Otani, H
    Kohzuki, M
    Takahashi, S
    Yamada, S
    Shirato, K
    Zuguchi, M
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (03) : 774 - 778
  • [5] Does digital acquisition reduce patients' skin dose in cardiac interventional procedures? An experimental study
    Chida, K
    Saito, H
    Zuguchi, M
    Shirotori, K
    Kumagai, S
    Nakayama, H
    Matsubara, K
    Kohzuki, M
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 183 (04) : 1111 - 1114
  • [6] Indicators of the maximum radiation dose to the skin during percutaneous coronary intervention in different target vessels
    Chida, Koichi
    Saito, Haruo
    Kagaya, Yutaka
    Kohzuki, Masahiro
    Takai, Yoshihiro
    Takahashi, Shoki
    Yamada, Shogo
    Zuguchi, Masayuki
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 68 (02) : 236 - 241
  • [7] REDUCTION OF RADIATION EXPOSURE WHILE MAINTAINING HIGH-QUALITY FLUOROSCOPIC IMAGES DURING INTERVENTIONAL CARDIOLOGY USING NOVEL X-RAY TUBE TECHNOLOGY WITH EXTRA BEAM FILTERING
    DENBOER, A
    DEFEYTER, PJ
    HUMMEL, WA
    KEANE, D
    ROELANDT, JRTC
    [J]. CIRCULATION, 1994, 89 (06) : 2710 - 2714
  • [8] *FAO UN, 1996, SAF SER IAEA, V1
  • [9] ACCF/AHA/HRS/SCAI clinical competence statement on physician knowledge to optimize patient safety and image quality in fluoroscopically guided invasive cardiovascular procedures - A report of the American College of Cardiology Foundation/American Heart Association/American College of Physicians Task Force on Clinical Competence and Training
    Hirshfeld, JW
    Balter, SD
    Brinker, JA
    Kern, MJ
    Klein, LW
    Lindsay, BD
    Tommaso, CL
    Tracy, CM
    Wagner, LK
    [J]. CIRCULATION, 2005, 111 (04) : 511 - 532
  • [10] EFFECT OF PULSED PROGRESSIVE FLUOROSCOPY ON REDUCTION OF RADIATION-DOSE IN THE CARDIAC-CATHETERIZATION LABORATORY
    HOLMES, DR
    WONDROW, MA
    GRAY, JE
    VETTER, RJ
    FELLOWS, JL
    JULSRUD, PR
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (01) : 159 - 162