ROC-curve approach for determining the detection limit of a field chemical sensor

被引:13
作者
Fraga, Carlos G. [1 ]
Melville, Angela M. [1 ]
Wright, Bob W. [1 ]
机构
[1] Pacific NW Natl Lab, Richland, WA 99354 USA
关键词
D O I
10.1039/b607843e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The detection limit of a field chemical sensor under realistic operating conditions is determined by receiver operator characteristic (ROC) curves. The chemical sensor is an ion mobility spectrometry (IMS) device used to detect a chemical marker in diesel fuel. The detection limit is the lowest concentration of the marker in diesel fuel that obtains the desired true-positive probability (TPP) and false-positive probability (FPP). A TPP of 0.90 and a FPP of 0.10 were selected as acceptable levels for the field sensor in this study. The detection limit under realistic operating conditions is found to be between 2 to 4 ppm (w/w). The upper value is the detection limit under challenging conditions. The ROC-based detection limit is very reliable because it is determined from multiple and repetitive sensor analyses under realistic circumstances. ROC curves also clearly illustrate and gauge the effects data preprocessing and sampling environments have on the sensor's detection limit.
引用
收藏
页码:230 / 236
页数:7
相关论文
共 21 条
[1]   Hepascore: An accurate validated predictor of liver fibrosis in chronic hepatitis C infection [J].
Adams, LA ;
Bulsara, M ;
Rossi, E ;
Deboer, B ;
Speers, D ;
George, J ;
Kench, J ;
Farrell, G ;
McCaughan, GW ;
Jeffrey, GP .
CLINICAL CHEMISTRY, 2005, 51 (10) :1867-1873
[2]  
[Anonymous], 2005, ION MOBILITY SPECTRO, DOI [DOI 10.1201/9781420038972, DOI 10.1016/j.ijms.2010.09.009]
[3]   Receiver operating characteristics curves and related decision measures: A tutorial [J].
Brown, CD ;
Davis, HT .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2006, 80 (01) :24-38
[4]  
CARRANO J, 2004, CHEM BIOL SENSOR STU
[5]  
CHRISTESEN S, 2004, FEDERATION ANAL CHEM
[6]   Analysis of gaseous toxic industrial compounds and chemical warfare agent simulants by atmospheric pressure ionization mass spectrometry [J].
Cotte-Rodríguez, I ;
Justes, DR ;
Nanita, SC ;
Noll, RJ ;
Mulligan, CC ;
Sanders, NL ;
Cooks, RG .
ANALYST, 2006, 131 (04) :579-589
[7]   Detection: International update, and some emerging di-lemmas involving calibration, the blank, and multiple detection decisions [J].
Currie, LA .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1997, 37 (01) :151-181
[8]   DETECTION LIMITS AND RECEIVER OPERATING CHARACTERISTIC CURVE ANALYSIS IN THE EVALUATION OF SPECIFIC IGE ASSAYS [J].
DOLEN, WK .
ALLERGY PROCEEDINGS, 1994, 15 (06) :319-322
[9]   RECEIVER OPERATOR CHARACTERISTIC (ROC) CURVES AND NONNORMAL DATA - AN EMPIRICAL-STUDY [J].
GODDARD, MJ ;
HINBERG, I .
STATISTICS IN MEDICINE, 1990, 9 (03) :325-337
[10]  
Heeschen C, 1999, CLIN CHEM, V45, P1789