Comparison of Fluid Dynamics Variations Between Chimney and Fenestrated Endografts for Pararenal Aneurysms Repair: A Patient Specific Computational Study as Motivation for Clinical Decision-Making

被引:8
作者
Moulakakis, Konstantinos G. [1 ]
Kakisis, John [1 ]
Gonidaki, Eleni [2 ]
Lazaris, Andreas M. [1 ]
Tsangaris, Sokrates [2 ]
Geroulakos, George [1 ]
Manopoulos, Christos [2 ]
机构
[1] Natl & Kapodistrian Univ Athens, Attikon Univ Hosp, Med Sch, Dept Vasc Surg, Athens, Greece
[2] Natl Tech Univ Athens, Sch Mech Engn, Fluids Sect, Biofluid Mech & Biomed Engn Lab, Athens, Greece
关键词
computational fluid dynamics; aneurysm; pararenal; endovascular repair; chimney; fenestrated; ABDOMINAL AORTIC-ANEURYSM; BLOOD-FLOW; INTRALUMINAL THROMBUS; ENDOVASCULAR REPAIR; BOUNDARY-CONDITIONS; STENT-GRAFTS; HEMODYNAMICS; ULTRASOUND; MECHANICS; VELOCITY;
D O I
10.1177/1538574419867531
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background-Aim: Limited data exist concerning the fluid dynamic changes induced by endovascular aortic repair with fenestrated and chimney graft modalities in pararenal aneurysms. We aimed to investigate and compare the wall shear stress (WSS) and flow dynamics for the branch vessels before and after endovascular aortic repair with fenestrated and chimney techniques. Methods: Modeling was done for patient specific pararenal aortic aneurysms employing fenestrated and chimney grafts (Materialise Mimics 10.0) before and after the endovascular procedure, using computed tomography scans of patients. Surface and spatial grids were created using the ANSYS CFD meshing software 2019 R2. Assessment of blood flow, streamlines, and WSS before and after aneurysm repair was performed. Results: The endovascular repair with chimney grafts leaded to a 43% to 53% reduction in perfusion in renal arteries. In fenestrated reconstruction, we observed a 15% reduced perfusion in both renal arteries. In both cases, we observed a decrease in the recirculation phenomena of the aorta after endovascular repair. Concerning the grafts of the renal arteries, we observed in both the transverse and longitudinal axes low WSS regions with simultaneous recirculation of the flow 1 cm distal to the ostium sites in both aortic graft models. High WSS regions appeared in the sites of ostium. Conclusions: We observed reduced renal perfusion in chimney grafts compared to fenestrated grafts, probably caused by the long and kinked characteristics of these devices.
引用
收藏
页码:572 / 582
页数:11
相关论文
共 47 条
[1]   A longitudinal comparison of hemodynamics and intraluminal thrombus deposition in abdominal aortic aneurysms [J].
Arzani, Amirhossein ;
Suh, Ga-Young ;
Dalman, Ronald L. ;
Shadden, Shawn C. .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2014, 307 (12) :H1786-H1795
[2]   Comparison of fenestrated endovascular aneurysm repair and chimney graft techniques for pararenal aortic aneurysm [J].
Banno, Hiroshi ;
Cochennec, Frederic ;
Marzelle, Jean ;
Becquemin, Jean-Pierre .
JOURNAL OF VASCULAR SURGERY, 2014, 60 (01) :31-39
[3]   Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms [J].
Casciaro, Mariano E. ;
Dottori, Javier ;
El-Batti, Salma ;
Alsac, Jean-Marc ;
Mousseaux, Elie ;
Larrabide, Ignacio ;
Craiem, Damian .
VASCULAR AND ENDOVASCULAR SURGERY, 2018, 52 (08) :621-628
[4]   Collected World Experience About the Performance of the Snorkel/Chimney Endovascular Technique in the Treatment of Complex Aortic Pathologies The PERICLES Registry [J].
Donas, Konstantinos P. ;
Lee, Jason T. ;
Lachat, Mario ;
Torsello, Giovanni ;
Veith, Frank J. .
ANNALS OF SURGERY, 2015, 262 (03) :546-553
[5]   Effect of Curvature on Displacement Forces Acting on Aortic Endografts: A 3-Dimensional Computational Analysis [J].
Figueroa, C. Alberto ;
Taylor, Charles A. ;
Yeh, Victoria ;
Chiou, Allen J. ;
Zarins, Christopher K. .
JOURNAL OF ENDOVASCULAR THERAPY, 2009, 16 (03) :284-294
[6]   Fluid Mechanics of Blood Clot Formation [J].
Fogelson, Aaron L. ;
Neeves, Keith B. .
ANNUAL REVIEW OF FLUID MECHANICS, VOL 47, 2015, 47 :377-403
[7]   Characterization of an abdominal aortic velocity waveform in patients with abdominal aortic aneurysm [J].
Fraser, Katharine H. ;
Meagher, Siobhan ;
Blake, James R. ;
Easson, William J. ;
Hoskins, Peter R. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2008, 34 (01) :73-80
[8]   Probabilistic 4D blood flow tracking and uncertainty estimation [J].
Friman, Ola ;
Hennemuth, Anja ;
Harloff, Andreas ;
Bock, Jelena ;
Markl, Michael ;
Peitgen, Heinz-Otto .
MEDICAL IMAGE ANALYSIS, 2011, 15 (05) :720-728
[9]  
Fuster V, 2005, CORONARY THROMBOSIS, P1636
[10]   Occlusion of Parallel/Bridging Stent-Grafts in the Treatment of Complex Aneurysms: A Purely Technical Issue . . . or Not? [J].
Georgakarakos, Efstratios ;
Manopoulos, Christos ;
Lazarides, Miltos K. ;
Tsangaris, Sokrates .
JOURNAL OF ENDOVASCULAR THERAPY, 2015, 22 (03) :401-405