Usability and Teachability of Continuous Glucose Monitoring Devices in Older Adults and Diabetes Educators: Task Analysis and Ease-of-Use Survey

被引:6
作者
Psavko, Simon [1 ]
Katz, Noam [1 ]
Mirchi, Tina [1 ]
Green, Courtney R. [1 ,2 ]
机构
[1] Dexcom Inc, San Diego, CA USA
[2] Dexcom Inc, 6340 Sequence Dr, San Diego, CA 92121 USA
来源
JMIR HUMAN FACTORS | 2022年 / 9卷 / 04期
关键词
medical devices; wearable devices; older adults; task analysis; usability testing; continuous glucose monitoring; glucose monitoring; glucose levels; diabetes; usability; HYPOGLYCEMIA; TECHNOLOGY; FRAMEWORK; ADOPTION; CARE;
D O I
10.2196/42057
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Continuous glucose monitoring (CGM) devices continuously sense and relay glucose concentration data from the interstitial fluid to a mobile phone or receiver. Older adults benefit from this continuous monitoring of glucose levels. Proper deployment of the sensing wire is facilitated by a specialized applicator. Objective: Our aim was to assess a new seventh-generation (G7) CGM device (Dexcom, Inc) for use by adults 65 years of age or older and certified diabetes care and education specialists (CDCESs). Ease of use related to intradermal insertion and mobile app setup will be assessed and compared to the fifth-and sixth-generation systems. Methods: Formal task analysis was conducted to enumerate the number and complexity of tasks associated with CGM deployment. We recruited 10 older adults with no prior CGM experience and 10 CDCESs to assess ease of use through hands-on insertion and initiation of a G7 system followed by a survey and, for older adults, a system usability scale survey. Results: About half as many tasks are needed to deploy G7 compared to G6. Older adults and CDCESs reported overall high usability of the G7 CGM device. CDCESs noted G7's easier setup compared to previous generations. The system usability scale score for the CGM system was 92.8, which reflects excellent usability. Conclusions: For CDCESs and for older adults using the G7 CGM system, cognitive burden is relatively low and reduced compared to previous CGM systems. Easing of this burden and simplification of the glucose monitoring aspect of proper diabetes management will likely contribute to improved outcomes in this population.
引用
收藏
页数:8
相关论文
共 35 条
[1]  
Allen NA, 2020, Diabetes Digit Health, P131, DOI 10.1016/B978-0-12-817485-2.00010-9
[2]   7. Diabetes Technology: Standards of Medical Care in Diabetes-2022 [J].
不详 .
DIABETES CARE, 2022, 45 :S97-S112
[3]  
[Anonymous], 2016, Applying human factors and usability engineering to medical devices
[4]  
[Anonymous], MED ACC CGMS JUST GO
[5]  
[Anonymous], 2021, MOR CGMS NOW COV MED
[6]  
[Anonymous], 2022, DL33822 GLUC MON
[7]  
[Anonymous], 2015, 6236612015EN IEC ISO
[8]  
Association for the Advancement of Medical Instrumentation, 2013, HUM FACT ENG DES MED
[9]  
Bangor A, 2009, J USABILITY STUD, V4, P114
[10]   Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range [J].
Battelino, Tadej ;
Danne, Thomas ;
Bergenstal, Richard M. ;
Amiel, Stephanie A. ;
Beck, Roy ;
Biester, Torben ;
Bosi, Emanuele ;
Buckingham, Bruce A. ;
Cefalu, William T. ;
Close, Kelly L. ;
Cobelli, Claudio ;
Dassau, Eyal ;
DeVries, J. Hans ;
Donaghue, Kim C. ;
Dovc, Klemen ;
Doyle, Francis J. ;
Garg, Satish ;
Grunberger, George ;
Heller, Simon ;
Heinemann, Lutz ;
Hirsch, Irl B. ;
Hovorka, Roman ;
Jia, Weiping ;
Kordonouri, Olga ;
Kovatchev, Boris ;
Kowalski, Aaron ;
Laffel, Lori ;
Levine, Brian ;
Mayorov, Alexander ;
Mathieu, Chantal ;
Murphy, Helen R. ;
Nimri, Revital ;
Norgaard, Kirsten ;
Parkin, Christopher G. ;
Renard, Eric ;
Rodbard, David ;
Saboo, Banshi ;
Schatz, Desmond ;
Stoner, Keaton ;
Urakami, Tatsuiko ;
Weinzimer, Stuart A. ;
Phillip, Moshe .
DIABETES CARE, 2019, 42 (08) :1593-1603