Anterior chamber volume measurements with Visante optical coherence tomography and Pentacam: repeatability and level of agreement

被引:24
作者
Labiris, Georgios [1 ]
Gkika, Maria [1 ]
Katsanos, Andreas [1 ,2 ]
Fanariotis, Michael [1 ]
Alvanos, Efstratios [1 ]
Kozobolis, Vassilios [1 ]
机构
[1] Democritus Univ Thrace, Sch Med, Dept Ophthalmol, Alexandroupolis, Greece
[2] Univ Hosp Larissa, Dept Ophthalmol, Larisa, Greece
关键词
anterior segment imaging; ophthalmic devices; Pentacam; screening; Visante;
D O I
10.1111/j.1442-9071.2009.02132.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
P>Background: The purpose of this study is to examine the repeatability of Pentacam-derived anterior chamber volume measurements, and assess the level of agreement of anterior chamber volume measurements obtained with the Visante and the Pentacam imaging devices. Methods: Anterior chamber volume was calculated with the Pentacam device and the Visante using the Wang formula. Results were analysed by means of the Wilcoxon signed rank test, Bland-Altman analysis, Cronbach's alpha test and intraclass correlation coefficient. P-values < 0.01 were considered statistically significant. Results: One randomly selected eye from each of 20 persons was analysed. Spherical equivalent ranged between -2.50 and +1.50 D, and their age ranged from 25 to 79 years (mean +/- SD: 54.4 +/- 19.2 years). Mean anterior chamber volume measured with Pentacam was 171.1 +/- 39.6 mu L (minimum: 103.0 mu L, maximum: 236.0 mu L), and mean anterior chamber volume measured by Visante was 171.4 +/- 42.4 mu L (minimum: 96.4 mu L, maximum: 245.5 mu L). There was no statistically significant difference between Pentacam and Visante measurements of the anterior chamber volume (P = 0.691, Wilcoxon signed rank test). Cronbach's alpha and intraclass correlation coefficient for Pentacam were 0.998 and 0.991, respectively. Conclusions: Anterior chamber volume measurements obtained with the Visante using the Wang formula were in good agreement with the corresponding ones obtained with Pentacam. This recently described method is easy to perform and might contribute to both clinical and research settings.
引用
收藏
页码:772 / 774
页数:3
相关论文
共 9 条
[1]  
BEHRING A, 2005, OPHTHAL RES, V37, P13
[2]  
BLAND LM, 1986, LANCET, V8, P307
[3]  
FONTANA ST, 1980, ARCH OPHTHALMOL-CHIC, V98, P1803
[4]  
JOHNSON SB, 1977, INVEST OPHTH VIS SCI, V16, P633
[5]   SIMPLE PHOTOGRAMMETRIC METHOD OF MEASURING ANTERIOR-CHAMBER VOLUME [J].
JOHNSON, SB ;
COAKES, RL ;
BRUBAKER, RF .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1978, 85 (04) :469-474
[6]   Validity and Repeatability of Anterior Chamber Depth Measurements With Pentacam and Orbscan [J].
Lackner, Birgit ;
Schmidinger, Gerald ;
Skorpik, Christian .
OPTOMETRY AND VISION SCIENCE, 2005, 82 (09) :858-861
[7]  
LEE DA, 1984, ARCH OPHTHALMOL-CHIC, V102, P46
[8]   A method of measuring anterior chamber volume using the anterior segment optical coherence tomographer and specialized software [J].
Wang, Ningli ;
Wang, Bingsong ;
Zhai, Gaoshou ;
Lei, Kun ;
Wang, Lan ;
Congdon, Nathan .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (05) :879-881
[9]  
Yi Jeong-Ho, 2008, Korean J Ophthalmol, V22, P242, DOI 10.3341/kjo.2008.22.4.242