Assessment of Neodymium:Yttrium-Aluminum-Garnet (Nd:YAG) Laser Fiber Wear and Tear Grade During Visual Laser Ablation of Prostatic Adenoma

被引:0
作者
Rozanski, Waldemar [1 ]
Klimek, Leszek [2 ]
Jablonowski, Zbigniew [1 ]
Sosnowski, Marek [1 ]
机构
[1] Med Univ Lodz, Dept Urol 1, PL-90549 Lodz, Poland
[2] Tech Univ Lodz, Inst Mat Engn, PL-90924 Lodz, Poland
关键词
PROSTATECTOMY; VAPORIZATION; RESECTION;
D O I
10.1089/pho.2008.2402
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: The extent of Nd-YAG laser fiber damage due to inappropriate handling during a common surgical procedure was assessed. Materials and Methods: Nd:YAG laser and Urolase noncontact fibers with 7.5 F gold reflective surface tips have been used to treat benign prostatic hyperplasia since 1995. Nine gold Urolase fiber tips damaged during the procedure and one tip not used for the surgical procedure were analyzed. The surfaces and cross-sections of gold fiber tips used with an Nd:YAG laser were observed under a scanning electron microscope. In order to improve imaging quality, mirror cross-sections embedded in resin were analyzed. Results: The observations, both on the macro and micro scales, revealed a high degree of damage to the gold reflective tip surfaces. In order to observe changes taking place in the reflective surface structure, atomic maps of the studied fiber tips were obtained. Mixing of the atoms of the various elements present in the reflective surface structure was observed. The extent of this phenomenon was dependent on the energy applied during the surgery and on the duration of the procedure. Conclusion: The gold reflective surface tip of a Urolase fiber becomes damaged with inappropriate handling during visual laser ablation procedures. The amount of energy accumulated in the fiber tip leads to a considerable rise in temperature, which can be dangerous for living tissue and may lead to accidental damage of the urinary bladder wall or the urethra.
引用
收藏
页码:799 / 805
页数:7
相关论文
共 13 条
[1]   Vaporization of the prostate with the 80 W potassium titanyl phosphate (KTP) laser. Technique and 6 months follow-up after 70 procedures [J].
Bachmann, A ;
Reich, O ;
Wyler, S ;
Ruszat, R ;
Casella, R ;
Gasser, T ;
Hofstetter, A ;
Sulser, T .
UROLOGE A, 2004, 43 (10) :1262-+
[2]   New technologies in endourology -: High-power thulium fiber laser ablation of urinary tissues at 1.94 μm [J].
Fried, NM ;
Murray, KE .
JOURNAL OF ENDOUROLOGY, 2005, 19 (01) :25-31
[3]   Sexual function in patients treated for benign prostatic hyperplasia [J].
Kassabian, VS .
LANCET, 2003, 361 (9351) :60-62
[4]   Performance and safety of holmium: YAG laser optical fibers [J].
Knudsen, BE ;
Glickman, RD ;
Stallman, KJ ;
Maswadi, S ;
Chew, BH ;
Beiko, DT ;
Denstedt, JD ;
Teichman, JMH .
JOURNAL OF ENDOUROLOGY, 2005, 19 (09) :1092-1097
[5]   Transurethral holmium laser enucleation of the prostate versus transurethral electrocautery resection of the prostate: A randomized prospective trial in 200 patients [J].
Kuntz, RM ;
Ahyai, S ;
Lehrich, K ;
Fayad, A .
JOURNAL OF UROLOGY, 2004, 172 (03) :1012-1016
[6]  
MALEK RS, 2004, CONT UROL WYD POL, V10, P20
[7]  
MIEKOS M, 1996, LEK WOJ S, V3, P199
[8]  
MIEKOS M, 1996, LEK WOJ S, V3, P74
[9]   Review of laser fibers: A practical guide for urologists [J].
Nazif, OA ;
Teichman, JMH ;
Glickman, RD ;
Welch, AJ .
JOURNAL OF ENDOUROLOGY, 2004, 18 (09) :818-829
[10]   High power (80 W) potassium-titanyl-phosphate laser vaporization of the prostate in 66 high risk patients [J].
Reich, O ;
Bachmann, A ;
Siebels, M ;
Hofstetter, A ;
Stief, CG ;
Sulser, T .
JOURNAL OF UROLOGY, 2005, 173 (01) :158-160