The "Sandwich Phenomenon": A Rare Complication in Adjacent, Double-Level X-Stop Surgery Report of Three Cases and Review of the Literature

被引:16
作者
Barbagallo, Giuseppe M. V. [1 ]
Corbino, Leonardo A. [1 ]
Olindo, Giuseppe [1 ]
Foti, Pietro [2 ]
Albanese, Vincenzo [1 ]
Signorelli, Francesco [3 ]
机构
[1] Azienda Osped Univ Policlin G Rodolico, Dept Neurosurg, Catania, Italy
[2] Azienda Osped Univ Policlin G Rodolico, Dept Radiol, Catania, Italy
[3] Magna Graecia Univ Catanzaro, Dept Neurosurg, Catanzaro, Italy
关键词
complication; distractor; inerspinous device; lumbar stenosis; sandwich phenomenon; X-Stop; LUMBAR SPINAL STENOSIS; NEUROGENIC INTERMITTENT CLAUDICATION; INTERSPINAL PROCESS DECOMPRESSION; DEGENERATIVE SPONDYLOLISTHESIS; DISTRACTION DEVICE; PROCESS IMPLANT; SYSTEM; MULTICENTER; KINEMATICS; INSERTION;
D O I
10.1097/BRS.0b013e3181c83820
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Case reports. Objective. To report the atraumatic fracture of the intervening (L4) spinous process (SP) in patients treated with X-Stop at 2 adjacent levels and discuss the potential underlying causes. Summary of Background Data. Limited evidence on complications in double-level X-Stop surgery is available. Methods. Three men, 47, 63, and 75 years old, respectively, underwent surgery with insertion of X-Stop at L3-L4 and L4-L5 because of low back pain and neurogenic claudication due to degenerative lumbar spine conditions. Two 10 mm devices were implanted in the first patient and two 16 mm distractors in the second man. One 10 mm X-Stop at L3-L4 and one 12 mm at L4-L5, respectively, were implanted in the third patient. No intraoperative complications occurred and the postoperative course was uneventful, with symptoms relief. Results. The first 2 patients presented because of recurrent symptoms 4 and 6 months after surgery, respectively. Imaging revealed "spontaneous" fracture of the L4 SP in both of them. Revision surgery was required, with removal of the interspinous devices, decompression and fixation. In the third patient the L4 SP fracture was detected when the patient presented because of recurrent back pain 18 months after the index surgery, but revision surgery was not consented. Conclusion. To our knowledge this is the first report describing the "sandwich phenomenon," i.e., the atraumatic fracture of the intervening SP in patients with adjacent, double-level X-Stop. Possible underlying theories and anatomic peculiarities which may predispose to this rare event are discussed.
引用
收藏
页码:E96 / E100
页数:5
相关论文
共 22 条
[1]   Treatment of neurogenic claudication by interspinous decompression: application of the X STOP device in patients with lumbar degenerative spondylolisthesis [J].
Anderson, Paul A. ;
Tribus, Cliff B. ;
Kitchel, Scott H. .
JOURNAL OF NEUROSURGERY-SPINE, 2006, 4 (06) :463-471
[2]   ANALYSIS OF COMPLICATIONS IN PATIENTS TREATED WITH THE X-STOP INTERSPINOUS PROCESS DECOMPRESSION SYSTEM: PROPOSAL FOR A NOVEL ANATOMIC SCORING SYSTEM FOR PATIENT SELECTION AND REVIEW OF THE LITERATURE [J].
Barbagallo, Giuseppe M. V. ;
Olindo, Giuseppe ;
Corbino, Leonardo ;
Albanese, Vincenzo .
NEUROSURGERY, 2009, 65 (01) :111-120
[3]   Self-rated evaluation of outcome of the implantation of interspinous process distraction (X-Stop) for neurogenic claudication [J].
Brussee, Paul ;
Hauth, Jakob ;
Donk, Roland D. ;
Verbeek, Andre L. M. ;
Bartels, Ronald H. M. A. .
EUROPEAN SPINE JOURNAL, 2008, 17 (02) :200-203
[4]  
Epstein Nancy E, 2009, Spine J, V9, pe6, DOI 10.1016/j.spinee.2008.08.004
[5]  
HEIJNEN SAF, 2004, NED TIJDSCHR ORTHOP, V11, P199
[6]   Quality of life of lumbar stenosis-treated patients in whom the X STOP interspinous device was implanted [J].
Hsu, Ken Y. ;
Zucherman, James F. ;
Hartjen, Charles A. ;
Mehalic, Thomas F. ;
Implicito, Dante A. ;
Martin, Michael J. ;
Johnson, Donald R., II ;
Skidmore, Grant A. ;
Vessa, Paul P. ;
Dwyer, James W. ;
Cauthen, Joseph C. ;
Ozuna, Richard M. .
JOURNAL OF NEUROSURGERY-SPINE, 2006, 5 (06) :500-507
[7]   A novel technique of intra-spinous process injection of PMMA to augment the strength of an inter-spinous process device such as the X STOP [J].
Idler, Cary ;
Zucherman, James F. ;
Yerby, Scott ;
Hsu, Ken Y. ;
Hannibal, Matthew ;
Kondrashov, Dimitriy .
SPINE, 2008, 33 (04) :452-456
[8]  
Kondrashov DG, 2007, INTERNET J MINIM INV, P1
[9]   Interspinous process decompression with the X-STOP device for lumbar spinal stenosis - A 4-year follow-up study [J].
Kondrashov, Dimitriy G. ;
Hannibal, Matthew ;
Hsu, Ken Y. ;
Zucherman, James F. .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (05) :323-327
[10]   An interspinous process distractor (X STOP) for lumbar spinal stenosis in elderly patients - Preliminary experiences in 10 consecutive coses [J].
Lee, JB ;
Hida, K ;
Seki, T ;
Iwasaki, Y ;
Minoru, A .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2004, 17 (01) :72-77