Advanced field shimming technology to reduce the influence of a screening current in a REBCO coil for a high-resolution NMR magnet

被引:28
作者
Iguchi, S. [1 ,2 ]
Piao, R. [1 ,3 ]
Hamada, M. [4 ]
Matsumoto, S. [5 ]
Suematsu, H. [6 ]
Takao, T. [2 ]
Saito, A. T. [3 ]
Li, J. [3 ]
Nakagome, H. [3 ]
Jin, X. [1 ]
Takahashi, M. [1 ]
Maeda, H. [1 ]
Yanagisawa, Y. [1 ,3 ]
机构
[1] RIKEN, Ctr Life Sci Technol, Yokohama, Kanagawa 2300045, Japan
[2] Sophia Univ, Fac Sci & Technol, Yotsuya 1028554, Japan
[3] Chiba Univ, Grad Sch Engn, Chiba 2368522, Japan
[4] Japan Superconductor Technol Inc, Kobe, Hyogo 6512271, Japan
[5] Natl Inst Mat Sci, Sperconducting Wire Unit, Tsukuba, Ibaraki 3050003, Japan
[6] JEOL RESONANCE Inc, Akishima, Tokyo 1968558, Japan
基金
日本科学技术振兴机构;
关键词
compact super-high field NMR; high-resolution NMR; REBCO coil; field inhomogeneity; screening current; shimming; ferromagnetic shim; 400 MHZ NMR; 1; GHZ; OPERATION;
D O I
10.1088/0953-2048/29/4/045013
中图分类号
O59 [应用物理学];
学科分类号
摘要
This paper describes a field shimming technology to obtain a spatially homogeneous magnetic field required for a high-resolution nuclear magnetic resonance (NMR) magnet under the influence of a screening current in a (RE) Ba2Cu3O7-x (REBCO) coil. Use of REBCO inner coils is one solution to realize a super-high field (>23.5 T, 1 GHz) NMR magnet. However, a REBCO coil generates a large amount of field error harmonics due to the inhomogeneous coil winding introduced by a thin tape conductor. In addition, the performance of a field correction coil and outer superconducting (SC) shim coils are significantly reduced due to the shielding effect of the screening current in the REBCO coil. Therefore, conventional shimming technology using SC shim coils and room temperature shim coils cannot adequately compensate those field error harmonics and high-resolution NMR measurements are not possible. In the present paper, an advanced field shimming technology including an inner SC shim coil and a novel type of ferromagnetic shim were installed in a 400 MHz low-temperature SC/REBCO NMR magnet. The inner SC shim coil and the ferromagnetic shim compensated for the reduction in the performances of the field correction coil and the SC shim coils, respectively. The field error harmonics were greatly compensated with the technology and a high NMR resolution <0.01 ppm was obtained. The results from the present work suggest an optimal shimming procedure for a super-high field NMR magnet with high-temperature superconductors inner coils, i.e. the best mix of shims.
引用
收藏
页数:12
相关论文
共 34 条
  • [1] Magnetic field generated by shielding current in high Tc superconducting coils for NMR magnets
    Amemiya, Naoyuki
    Akachi, Ken
    [J]. SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2008, 21 (09)
  • [2] Progress in performance of DI-BSCCO family
    Ayai, N.
    Kobayashi, S.
    Kikuchi, M.
    Ishida, T.
    Fujikami, J.
    Yamazaki, K.
    Yamade, S.
    Tatamidani, K.
    Hayashi, K.
    Sato, K.
    Kitaguchi, H.
    Kumakura, H.
    Osamura, K.
    Shimoyama, J.
    Kamijyo, H.
    Fukumoto, Y.
    [J]. PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2008, 468 (15-20): : 1747 - 1752
  • [3] Bascuñán J, 2014, IEEE T APPL SUPERCON, V24, DOI [10.1109/tasc.2013.2285781, 10.1109/TASC.2013.2285781]
  • [4] THIN SUPERCONDUCTORS IN A PERPENDICULAR MAGNETIC AC FIELD - GENERAL FORMULATION AND STRIP GEOMETRY
    BRANDT, EH
    [J]. PHYSICAL REVIEW B, 1994, 49 (13): : 9024 - 9040
  • [5] Bruce Montgomery D., 1969, Solenoid Magnet Design, The Magnetic and Mechanical Aspects of Resistive and Superconducting System
  • [6] Claridge T.D.W., 1999, High-Resolution NMR Technology in Organic Chemistry
  • [7] Achievement of 1020 MHz NMR
    Hashi, Kenjiro
    Ohki, Shinobu
    Matsumoto, Shinji
    Nishijima, Gen
    Goto, Atsushi
    Deguchi, Kenzo
    Yamada, Kazuhiko
    Noguchi, Takashi
    Sakai, Shuji
    Takahashi, Masato
    Yanagisawa, Yoshinori
    Iguchi, Seiya
    Yamazaki, Toshio
    Maeda, Hideaki
    Tanaka, Ryoji
    Nemoto, Takahiro
    Suematsu, Hiroto
    Miki, Takashi
    Saito, Kazuyoshi
    Shimizu, Tadashi
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2015, 256 : 30 - 33
  • [8] Recent Developments in 2G HTS Coil Technology
    Hazelton, Drew W.
    Selvamanickam, Venkat
    Duval, Jason M.
    Larbalestier, David C.
    Markiewicz, William Denis
    Weijers, Hubertus W.
    Holtz, Ronald L.
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2009, 19 (03) : 2218 - 2222
  • [9] High-order MR shimming: a simulation study of the effectiveness of competing methods, using an established susceptibility model of the human head
    Hillenbrand, DF
    Lo, KM
    Punchard, WFB
    Reese, TG
    Starewicz, PM
    [J]. APPLIED MAGNETIC RESONANCE, 2005, 29 (01) : 39 - 64
  • [10] Shimming for the 1020 MHz LTS/Bi-2223 NMR Magnet
    Iguchi, Seiya
    Yanagisawa, Yoshinori
    Takahashi, Masato
    Takao, Tomoaki
    Hashi, Kenjiro
    Ohki, Shinobu
    Nishijima, Gen
    Matsumoto, Shinji
    Noguchi, Takashi
    Tanaka, Ryoji
    Suematsu, Hiroto
    Saito, Kazuyoshi
    Shimizu, Tadashi
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (04)