摘要
Purpose. High-energy photons are most commonly used in radiotherapy to treat cancer. Wedge filters are required to obtain homogeneous dose distribution in the patient. Different wedge filter types create different surface doses. In this study, the effect of the virtual and physical wedge filters on the surface and build-up region doses was examined for 6- and 18-MV high-energy photon beams. Materials and methods. The measurements were made in a water equivalent phantom in the build-up region at a 100-cm source-to-surface distance for various field sizes using virtual and physical wedge filters having different angles. A parallel-plate ion chamber was used to measure the percent depth doses. Results. The percentage dose at the surface increased as the field size increased for open, virtual, and physical wedged beams. For open, physical, and virtual wedged beams, the surface doses were found to be 15.4%, 9.9%, and 15.9% with 6-MV photons and 10.6%, 8.8%, 11.9% with 18-MV photons, respectively, at 10 x 10 cm(2) field size. Conclusion. Build-up doses of virtual wedged beams were similar to those of open beams. Surface and buildup doses of physical wedged beams were lower than those of open and virtual wedged beams.
引用
收藏
页码:110 / 116
页数:7
相关论文
共 22 条
[1]   Extrapolated surface dose measurements with radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
MEDICAL PHYSICS, 1999, 26 (03) :485-488
[2]   Accurate skin dose measurements using radiochromic film in clinical applications [J].
Devic, S ;
Seuntjens, J ;
Abdel-Rahman, W ;
Evans, M ;
Olivares, M ;
Podgorsak, EB ;
Vuong, T ;
Soares, CG .
MEDICAL PHYSICS, 2006, 33 (04) :1116-1124
[3]   EFFECTS OF BEAM MODIFIERS AND IMMOBILIZATION DEVICES ON THE DOSE IN THE BUILDUP REGION [J].
FONTENLA, DP ;
NAPOLI, JJ ;
HUNT, M ;
FASS, D ;
MCCORMICK, B ;
KUTCHER, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (01) :211-219
[4]   MEASUREMENT OF DOSE IN THE BUILDUP REGION USING FIXED-SEPARATION PLANE-PARALLEL IONIZATION CHAMBERS [J].
GERBI, BJ ;
KHAN, FM .
MEDICAL PHYSICS, 1990, 17 (01) :17-26
[5]   THE POLARITY EFFECT FOR COMMERCIALLY AVAILABLE PLANE-PARALLEL IONIZATION CHAMBERS [J].
GERBI, BJ ;
KHAN, FM .
MEDICAL PHYSICS, 1987, 14 (02) :210-215
[6]  
*IAEA, 1997, IAEA TECHN REP SER, P381
[7]   Photon beam skin dose analyses for different clinical setups [J].
Kim, S ;
Liu, CR ;
Zhu, TC ;
Palta, JR .
MEDICAL PHYSICS, 1998, 25 (06) :860-866
[8]   ENTRANCE AND EXIT DOSE REGIONS FOR A CLINAC-2100C [J].
KLEIN, EE ;
PURDY, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (02) :429-435
[9]  
Klein Eric E, 2003, J Appl Clin Med Phys, V4, P1, DOI 10.1120/1.1520113
[10]  
LEAVITT DD, 1993, MED PHYS, V20, P877