Estimation of ELISA results using a parallel curve analysis

被引:10
作者
Bursa, Francis [1 ]
Yellowlees, Ann [1 ]
Bishop, Alka [2 ]
Beckett, Angela [2 ,3 ]
Hallis, Bassam [4 ]
Matheson, Mary [4 ]
机构
[1] Quant Biostat, Edinburgh, Midlothian, Scotland
[2] Porton Biopharma Ltd, Porton Down, England
[3] Univ Portsmouth, Ctr Enzyme Innovat, Portsmouth, Hants, England
[4] Publ Hlth England, Porton Down, England
关键词
ELISA; ELISA statistics; Concentration estimation; Immunoassay; ERROR;
D O I
10.1016/j.jim.2020.112836
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We introduce a new method for the analysis of enzyme-linked immunosorbent assay (ELISA) data. The new method can use data near the asymptotes and does not give undue weight to responses on the flatter parts of the dose-response curve. We apply it to simulated data and to two real-world assays and show it is more accurate and more precise than the traditional interpolation method. In particular, the new method works much better for very low-concentration samples for which the traditional method is often unable to give a result.
引用
收藏
页数:9
相关论文
共 11 条
[1]  
[Anonymous], 2018, R LANG ENV STAT COMP
[2]   Statistical estimation of antibody concentration using multiple dilutions [J].
Cheung, Yin Bun ;
Xu, Ying ;
Remarque, Edmond J. ;
Milligan, Paul .
JOURNAL OF IMMUNOLOGICAL METHODS, 2015, 417 :115-123
[3]   Evaluating concentration estimation errors in ELISA microarray experiments [J].
Daly, DS ;
White, AM ;
Varnum, SM ;
Anderson, KK ;
Zangar, RC .
BMC BIOINFORMATICS, 2005, 6 (1)
[4]  
Fleetwood Kelly, 2015, PDA J Pharm Sci Technol, V69, P248, DOI 10.5731/pdajpst.2015.01016
[5]  
Hauck WW, 2005, PDA J PHARM SCI TECH, V59, P127
[6]   The effect of serial dilution error on calibration inference in immunoassay [J].
Higgins, KM ;
Davidian, M ;
Chew, G ;
Burge, H .
BIOMETRICS, 1998, 54 (01) :19-32
[7]   EQUIVALENCE TESTING FOR PARALLELISM IN THE FOUR-PARAMETER LOGISTIC MODEL [J].
Jonkman, Jeffrey N. ;
Sidik, Kurex .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2009, 19 (05) :818-837
[8]  
Lansky D., 2012, DESIGN ANAL POTENCY, P107
[9]  
Quantics, 2019, QUBAS VERS 2 01
[10]   A 2-COMPONENT MODEL FOR MEASUREMENT ERROR IN ANALYTICAL-CHEMISTRY [J].
ROCKE, DM ;
LORENZATO, S .
TECHNOMETRICS, 1995, 37 (02) :176-184