Characterization of penumbra sharpening and scattering by adaptive aperture for a compact pencil beam scanning proton therapy system

被引:14
作者
Grewal, Hardev S. [1 ,2 ]
Ahmad, Salahuddin [1 ]
Jin, Hosang [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Dept Radiat Oncol, 800 NE 10th St SCC L100, Oklahoma City, OK 73104 USA
[2] Oklahoma Proton Ctr, 5901 W Mem Rd, Oklahoma City, OK 73142 USA
关键词
adaptive aperture; compact proton therapy system; lateral penumbra; pencil beam scanning;
D O I
10.1002/mp.14771
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To quantitatively access penumbra sharpening and scattering by adaptive aperture (AA) under various beam conditions and clinical cases for a Mevion S250i compact pencil beam scanning proton therapy system. Methods First, in-air measurements were performed using a scintillation detector for single spot profile and lateral penumbra for five square field sizes (3 x 3 to 18 x 18 cm(2)), three energies (33.04, 147.36, and 227.16 MeV), and three snout positions (5, 15, and 33.6 cm) with Open and AA field. Second, treatment plans were generated in RayStation treatment planning system (TPS) for various combination of target size (3- and 10-cm cube), target depth (5, 10, and 15 cm) and air gap (5-20 cm) for both Open and AA field. These plans were delivered to EDR2 films in the solid water and penumbra reduction by AA was quantified. Third, the effect of the AA scattered protons on the surface dose was studied at 5 mm depth by EDR2 film and the RayStation TPS computation. Finally, dosimetric advantage of AA over Open field was studied for five brain and five prostate cases using the TPS simulation. Results The spot size changed dramatically from 3.8 mm at proton beam energy of 227.15 MeV to 29.4 mm at energy 33.04 MeV. In-air measurements showed that AA substantially reduced the lateral penumbra by 30% to 60%. The EDR2 film measurements in solid water presented the maximum penumbra reduction of 10 to 14 mm depending on the target size. The maximum increase of 25% in field edge dose at 5 mm depth as compared to central axis was observed. The substantial penumbra reduction by AA produced less dose to critical structures for all the prostate and brain cases. Conclusions Adaptive aperture sharpens the penumbra by factor of two to three depending upon the beam condition. The absolute penumbra reduction with AA was more noticeable for shallower target, smaller target, and larger air gap. The AA-scattered protons contributed to increase in surface dose. Clinically, AA reduced the doses to critical structures.
引用
收藏
页码:1508 / 1519
页数:12
相关论文
共 50 条
  • [31] Successful Implementation of Image-Guided Pencil-Beam Scanning Proton Therapy in Medulloblastomas
    Das, Anindita
    Gaikwad, Utpal
    Krishnan, Ganapathy
    Rajendran, Adhithyan
    Patil, Sushama
    Subramaniam, Preethi
    Krishna, Uday
    Wakde, Manoj
    Chilukuri, Srinivas
    Jalali, Rakesh
    DIAGNOSTICS, 2023, 13 (21)
  • [32] Increased Efficiency of Range Verification in Routine QA for Pencil-Beam Scanning Proton Therapy
    Seonghoon Jeong
    Myonggeun Yoon
    Jaeman Son
    Kwangzoo Chung
    Journal of the Korean Physical Society, 2018, 73 : 983 - 989
  • [33] Voxel-wise dose rate calculation in clinical pencil beam scanning proton therapy
    Daartz, Juliane
    Madden, Thomas M.
    Lalonde, Arthur
    Cascio, Ethan
    Verburg, Joost
    Shih, Helen
    MacDonald, Shannon
    Hachadorian, Rachael
    Schuemann, Jan
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (06)
  • [34] Collection efficiency of a monitor parallel plate ionization chamber for pencil beam scanning proton therapy
    Han, Rong-Cheng
    Li, Yong-Jiang
    Pu, Yue-Hu
    NUCLEAR SCIENCE AND TECHNIQUES, 2020, 31 (02)
  • [35] The dosimetric enhancement of GRID profiles using an external collimator in pencil beam scanning proton therapy
    Smith, Blake R.
    Nelson, Nicholas P.
    Geoghegan, Theodore J.
    Patwardhan, Kaustubh A.
    Hill, Patrick M.
    Yu, Jen
    Gutierrez, Alonso N.
    Allen, Bryan G.
    Hyer, Daniel E.
    MEDICAL PHYSICS, 2022, 49 (04) : 2684 - 2698
  • [36] Stochastic model for predicting the temporal structure of the plan delivery in a synchrotron-based pencil beam scanning proton therapy system
    Burguete, J.
    Garcia-Cardosa, M.
    Antolin, E.
    Aguilar, B.
    Azcona, J. D.
    RADIATION PHYSICS AND CHEMISTRY, 2025, 226
  • [37] Prognostic impact of the "Sekhar grading system for cranial Chordomas" in patients treated with pencil beam scanning proton therapy: an institutional analysis
    Hottinger, Anna-Lena
    Bojaxhiu, Beat
    Ahlhelm, Frank
    Walser, Marc
    Bachtiary, Barbara
    Zepter, Stefan
    Lomax, Tony
    Pica, Alessia
    Weber, Damien C.
    RADIATION ONCOLOGY, 2020, 15 (01)
  • [38] Prognostic impact of the “Sekhar grading system for cranial Chordomas” in patients treated with pencil beam scanning proton therapy: an institutional analysis
    Anna-Lena Hottinger
    Beat Bojaxhiu
    Frank Ahlhelm
    Marc Walser
    Barbara Bachtiary
    Stefan Zepter
    Tony Lomax
    Alessia Pica
    Damien C. Weber
    Radiation Oncology, 15
  • [39] Simulation and experimental benchmarking of a proton pencil beam scanning nozzle model for development of MR-integrated proton therapy
    Oborn, Bradley M.
    Semioshkina, Ekaterina
    van der Kraaij, Erik
    Hoffmann, Aswin L.
    MEDICAL PHYSICS, 2024, 51 (09) : 6196 - 6205
  • [40] Assessment of pencil beam scanning proton therapy beam delivery accuracy through machine learning and log file analysis
    Ranjith, C. P.
    Krishnan, Mayakannan
    Raveendran, Vysakh
    Chaudhari, Lalit
    Laskar, Siddhartha
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 127