Characterization of penumbra sharpening and scattering by adaptive aperture for a compact pencil beam scanning proton therapy system

被引:14
|
作者
Grewal, Hardev S. [1 ,2 ]
Ahmad, Salahuddin [1 ]
Jin, Hosang [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Dept Radiat Oncol, 800 NE 10th St SCC L100, Oklahoma City, OK 73104 USA
[2] Oklahoma Proton Ctr, 5901 W Mem Rd, Oklahoma City, OK 73142 USA
关键词
adaptive aperture; compact proton therapy system; lateral penumbra; pencil beam scanning;
D O I
10.1002/mp.14771
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To quantitatively access penumbra sharpening and scattering by adaptive aperture (AA) under various beam conditions and clinical cases for a Mevion S250i compact pencil beam scanning proton therapy system. Methods First, in-air measurements were performed using a scintillation detector for single spot profile and lateral penumbra for five square field sizes (3 x 3 to 18 x 18 cm(2)), three energies (33.04, 147.36, and 227.16 MeV), and three snout positions (5, 15, and 33.6 cm) with Open and AA field. Second, treatment plans were generated in RayStation treatment planning system (TPS) for various combination of target size (3- and 10-cm cube), target depth (5, 10, and 15 cm) and air gap (5-20 cm) for both Open and AA field. These plans were delivered to EDR2 films in the solid water and penumbra reduction by AA was quantified. Third, the effect of the AA scattered protons on the surface dose was studied at 5 mm depth by EDR2 film and the RayStation TPS computation. Finally, dosimetric advantage of AA over Open field was studied for five brain and five prostate cases using the TPS simulation. Results The spot size changed dramatically from 3.8 mm at proton beam energy of 227.15 MeV to 29.4 mm at energy 33.04 MeV. In-air measurements showed that AA substantially reduced the lateral penumbra by 30% to 60%. The EDR2 film measurements in solid water presented the maximum penumbra reduction of 10 to 14 mm depending on the target size. The maximum increase of 25% in field edge dose at 5 mm depth as compared to central axis was observed. The substantial penumbra reduction by AA produced less dose to critical structures for all the prostate and brain cases. Conclusions Adaptive aperture sharpens the penumbra by factor of two to three depending upon the beam condition. The absolute penumbra reduction with AA was more noticeable for shallower target, smaller target, and larger air gap. The AA-scattered protons contributed to increase in surface dose. Clinically, AA reduced the doses to critical structures.
引用
收藏
页码:1508 / 1519
页数:12
相关论文
共 50 条
  • [21] Gantry-based pencil beam scanning proton therapy for uveal melanoma: IMPT versus proton arc therapy
    Qi, Hang
    Hu, Lei
    Huang, Sheng
    Lee, Yen-Po
    Yu, Francis
    Chen, Qing
    Yang, Yunjie
    Kang, Minglei
    Zhai, Huifang
    Vermeulen, Milo
    Shim, Andy
    Park, Peter
    Ding, Xuanfeng
    Zhou, Jun
    Abramson, David H.
    Francis, Jasmine H.
    Simone, Charles B.
    Barker, Christopher A.
    Lin, Haibo
    RADIATION ONCOLOGY, 2025, 20 (01)
  • [22] Evaluation of the ray-casting analytical algorithm for pencil beam scanning proton therapy
    Winterhalter, Carla
    Zepter, Stefan
    Shim, Sojin
    Meier, Gabriel
    Bolsi, Alessandra
    Fredh, Anna
    Hrbacek, Jan
    Oxley, David
    Zhang, Ye
    Weber, Damien C.
    Lomax, Antony
    Safai, Sairos
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (06)
  • [23] A pre-absorber optimization technique for pencil beam scanning proton therapy treatments
    Fracchiolla, Francesco
    Fellin, Francesco
    Innocenzi, Marta
    Lipparini, Mirko
    Lorentini, Stefano
    Widesott, Lamberto
    Farace, Paolo
    Schwarz, Marco
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 57 : 145 - 152
  • [24] Comment on 'Collimated proton pencil-beam scanning for superficial targets: impact of the order of range shifter and aperture'
    Winterhalter, Carla
    Lomax, Antony J.
    Oxley, David
    Weber, Damien C.
    Safai, Sairos
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (20)
  • [25] Intrafraction tumor motion monitoring and dose reconstruction for liver pencil beam scanning proton therapy
    Nankali, Saber
    Worm, Esben Schjodt
    Thomsen, Jakob Borup
    Stick, Line Bjerregaard
    Bertholet, Jenny
    Hoyer, Morten
    Weber, Britta
    Mortensen, Hanna Rahbek
    Poulsen, Per Rugaard
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [26] Collection efficiency of a monitor parallel plate ionization chamber for pencil beam scanning proton therapy
    Rong-Cheng Han
    Yong-Jiang Li
    Yue-Hu Pu
    Nuclear Science and Techniques, 2020, 31 (02) : 3 - 12
  • [27] Collection efficiency of a monitor parallel plate ionization chamber for pencil beam scanning proton therapy
    Rong-Cheng Han
    Yong-Jiang Li
    Yue-Hu Pu
    Nuclear Science and Techniques, 2020, 31
  • [28] Robust treatment planning in whole pelvis pencil beam scanning proton therapy for prostate cancer
    Butala, Anish A.
    Ingram, W. Scott
    O'Reilly, Shannon E.
    Hartl, Brett
    Kassaee, Ali
    Deville, Curtiland
    Vapiwala, Neha
    MEDICAL DOSIMETRY, 2020, 45 (04) : 334 - 338
  • [29] Validation and clinical implementation of an accurate Monte Carlo code for pencil beam scanning proton therapy
    Huang, Sheng
    Kang, Minglei
    Souris, Kevin
    Ainsley, Christopher
    Solberg, Timothy D.
    McDonough, James E.
    Simone, Charles B.
    Lin, Liyong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 558 - 572
  • [30] Increased Efficiency of Range Verification in Routine QA for Pencil-Beam Scanning Proton Therapy
    Jeong, Seonghoon
    Yoon, Myonggeun
    Son, Jaeman
    Chung, Kwangzoo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 73 (07) : 983 - 989