Compliant positioning of total hip components for optimal range of motion

被引:433
作者
Widmer, KH [1 ]
Zurfluh, B [1 ]
机构
[1] Univ Basel, Dept Orthopaed, Orthopaed Biomech Lab, Felix Platter Spital, CH-4052 Basel, Switzerland
关键词
total hip arthroplasty; component orientation; neck/cup impingement; dislocation; range of motion;
D O I
10.1016/j.orthres.2003.11.001
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Impingement between femoral neck and endoprosthetic cup is one of the causes for dislocation in total hip arthroplasty (THA). Choosing a correct combined orientation of both components, the acetabular cup and femoral stem, in manual or computer-assisted implantation will yield a maximized, stable range of motion (ROM) and will reduce the risk for dislocation. A mathematical model of a THA was developed to determine the optimal combination of cup inclination, cup anteversion, and stem antetorsion for maximizing ROM and minimizing the risk for cup-neck impingement. Single and combined hip joint motions were tested. A radiographic definition was used for component orientation. Additional parameters, such as stem-neck (CCD) angle, head-neck ratio, and the design of the acetabular opening, were also considered. The model showed that a maximized and safe ROM requires compliant, well-defined combinations of cup inclination, cup anteversion, and stem antetorsion depending on the intended ROM. Radiographic cup anteversion and stem antetorsion were linearly correlated. Additional internal rotation reduced flexion, and additional external rotation reduced extension, abduction and adduction. The articulating hemispheric surface of acetabular cups should be oriented between 40degrees and 45degrees of radiographic inclination, between 20degrees and 28degrees of radiographic cup anteversion, and should be combined with stem antetorsion so that the sum of cup anteversion plus 0.7 times the stem antetorsion equals 37degrees. Final component orientation must also consider cup containment, implant impingement with bone and soft tissue, and preoperative skeletal contractures or deformities to achieve the optimal compromise for each patient. (C) 2003 Orthopaedic Research Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:815 / 821
页数:7
相关论文
共 29 条
  • [1] Ceramic acetabular cups for total hip replacement - Part 6: Cup design, inclination angle and antetorsion angle have an influence on the range of motion and impingement
    Bader, R
    Willmann, G
    [J]. BIOMEDIZINISCHE TECHNIK, 1999, 44 (7-8): : 212 - 219
  • [2] The effect of femoral component head size on posterior dislocation of the artificial hip joint
    Bartz, RL
    Noble, PC
    Kadakia, NR
    Tullos, HS
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (09) : 1300 - 1307
  • [3] CHARNLEY J, 1970, CLIN ORTHOP RELAT R, P7
  • [4] CUNNINGHAM, 1922, PELVIS CUNNINGHAMS T, P255
  • [5] The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios
    D'Lima, DD
    Urquhart, AG
    Buehler, KO
    Walker, RH
    Colwell, CW
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (03) : 315 - 321
  • [6] Effects of acetabular abduction on cup wear rates in total hip arthroplasty
    Del Schutte, H
    Lipman, AJ
    Bannar, SM
    Livermore, JT
    Ilstrup, D
    Morrey, BF
    [J]. JOURNAL OF ARTHROPLASTY, 1998, 13 (06) : 621 - 626
  • [7] CONTRIBUTION TO TECHNIQUE OF ACETABULAR IMPLANTATION IN TOTAL HIP ARTHROPLASTY
    DICK, W
    MORSCHER, E
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1975, 83 (02) : 215 - 220
  • [8] A JOINT COORDINATE SYSTEM FOR THE CLINICAL DESCRIPTION OF 3-DIMENSIONAL MOTIONS - APPLICATION TO THE KNEE
    GROOD, ES
    SUNTAY, WJ
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (02): : 136 - 144
  • [9] HARRIS WH, 1980, CLIN ORTHOP RELAT R, P188
  • [10] Jaramaz B, 1998, CLIN ORTHOP RELAT R, V355, P8