Differentiation of heroin and cocaine using dual-energy CT-an experimental study

被引:12
|
作者
Grimm, Jochen [1 ,2 ,3 ,4 ]
Wudy, Ramona [1 ]
Ziegeler, Edvard [1 ]
Wirth, Stefan [1 ]
Uhl, Michael [5 ]
Reiser, Maximilian F. [1 ]
Scherr, Michael [1 ]
机构
[1] Ludwig Maximilians Univ Hosp Munich, Inst Clin Radiol, D-80336 Munich, Germany
[2] Univ Lausanne Hosp, Dept Med Radiol, CH-1011 Lausanne, Switzerland
[3] Univ Lausanne, CH-1011 Lausanne, Switzerland
[4] Ctr Legal Med Lausanne & Geneva, CH-1011 Lausanne, Switzerland
[5] Bavarian State Criminal Police Off, D-80636 Munich, Germany
关键词
Body packing; Dual-energy CT; Dual-energy index; Material differentiation; Illicit drugs; DOSE COMPUTED-TOMOGRAPHY; BODY-PACKING; PACKERS; RADIOGRAPHY; STUFFERS; PUSHERS; SURGERY; DRUGS;
D O I
10.1007/s00414-014-0980-6
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
To evaluate if heroin and cocaine can be distinguished using dual-energy CT. Twenty samples of heroin and cocaine at different concentrations and standardized compression (SC) were scanned in dual-energy mode on a newest generation Dual Energy 64-row MDCT scanner. CT number, spectral graphs, and dual-energy index (DEI) were evaluated. Results were prospectively tested on six original samples from a body packer. Wilcoxon's test was used for statistical evaluation. Values are given as median and range. Under SC, the CT number of cocaine samples (-29.87 Hounsfield unit (HU) [-125.85; 16.16 HU]) was higher than the CT number of heroin samples (-184.37 HU [-199.81; -159.25 HU]; p < 0.01). Slope of spectral curves for cocaine was -2.36 HU/keV [-7.15; -0.67 HU/keV], and for heroin, 1.75 HU/keV [1.28; 2.5 HU/keV] (p < 0.01). DEI was 0.0352 [0.0081; 0.0528] for cocaine and significantly higher than for heroin samples (-0.0127 [-0.0097; -0.0159]; p < 0.001). While CT number was inconclusive, all six original packs were correctly classified after evaluation of the spectral curve and DEI. In contrast to the CT number, slope of the spectral curve and DEI were independent of concentration and compression. The slope of the spectral curve and the DEI from dual-energy CT data can be used to distinguish heroin and cocaine in vitro; these results are independent of compression and concentration in the measured range.
引用
收藏
页码:475 / 482
页数:8
相关论文
共 50 条
  • [41] Quantitative assessment of the degree of differentiation in colon cancer with dual-energy spectral CT
    Yang Chuang-bo
    He Tai-ping
    Duan Hai-feng
    Jia Yong-jun
    Zhang Xi-rong
    Ma Guang-ming
    Ren Chenglong
    Wang Jun
    Yu Yong
    Abdominal Radiology, 2017, 42 : 2591 - 2596
  • [42] Quantitative assessment of the degree of differentiation in colon cancer with dual-energy spectral CT
    Yang Chuang-bo
    He Tai-ping
    Duan Hai-feng
    Jia Yong-jun
    Zhang Xi-rong
    Ma Guang-ming
    Ren Chenglong
    Wang Jun
    Yu Yong
    ABDOMINAL RADIOLOGY, 2017, 42 (11) : 2591 - 2596
  • [43] Differentiation of neoplastic from bland macroscopic portal vein thrombi using dual-energy spectral CT imaging: a pilot study
    Qian, Li Jun
    Zhu, Jiong
    Zhuang, Zhi Guo
    Xia, Qiang
    Cheng, Yu Fan
    Li, Jian Ying
    Xu, Jian Rong
    EUROPEAN RADIOLOGY, 2012, 22 (10) : 2178 - 2185
  • [44] Differentiation of neoplastic from bland macroscopic portal vein thrombi using dual-energy spectral CT imaging: a pilot study
    Li Jun Qian
    Jiong Zhu
    Zhi Guo Zhuang
    Qiang Xia
    Yu Fan Cheng
    Jian Ying Li
    Jian Rong Xu
    European Radiology, 2012, 22 : 2178 - 2185
  • [45] Dual-Energy CT and Spot Sign
    Vargas, M. I.
    Lovblad, K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2016, 37 (10) : E63 - E63
  • [46] Dual-Energy CT of Intrahepatic Splenosis
    Zhang, Hui-Hui
    Tan, Xian-Zheng
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 223 (06)
  • [47] Dual-Energy CT and Radiation Dose
    Reher, Thomas
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2020, 17 (01) : 95 - 96
  • [48] Renal applications of dual-energy CT
    Kaza, Ravi K.
    Platt, Joel F.
    ABDOMINAL RADIOLOGY, 2016, 41 (06) : 1122 - 1132
  • [49] Stratification of Patients With Liver Fibrosis Using Dual-Energy CT
    Lamb, Peter
    Sahani, Dushyant V.
    Fuentes-Orrego, Jorge M.
    Patino, Manuel
    Ghosh, Asish
    Mendonca, Paulo R. S.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2015, 34 (03) : 807 - 815
  • [50] Material decomposition using dual-energy CT with unsupervised learning
    Chang, Hui-Yu
    Liu, Chi-Kuang
    Huang, Hsuan-Ming
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (04) : 1607 - 1617