摘要
Chirp coded excitation is an effective method to improve the signal-to-noise ratio (SNR) and penetration depth of high-frequency endoscopic ultrasound (EUS) imaging. In coded excitation, pulse compression is applied to compress the elongated coded signals into a short pulse, which determines the final imaging performance, including spatial resolution and SNR. However, with the current pulse compression methods, it is hard to get high performance in the peak sidelobe level (PSL), image contrast, and axial resolution at the same time. To solve this problem, in this article, a new method named compressed pulse weighting method (CPWM) was proposed based on the combination of two kinds of pulse compression signals. A brief theoretical derivation proved the feasibility of method. The proposed method was evaluated by the simulation and phantom experiments. Compared with traditional method, the results showed that the proposed adaptive weighting method can provide increases of 32.42% in the penetration depth, 9.48 dB in the SNR, 5.60 dB in the contrast ratio (CR), 5.46 in the contrast-to-noise ratio (CNR), and 0.13 mm in the axial imaging resolution for 12-MHz EUS. Therefore, this method can effectively improve the ultrasound penetration depth and imaging quality, which made it have good potential for high-frequency ultrasound imaging.
引用
收藏
页码:446 / 452
页数:7
相关论文
共 22 条
[1]   Parameter optimization of pulse compression in ultrasound imaging systems with coded excitation [J].
Behar, V ;
Adam, D .
ULTRASONICS, 2004, 42 (10) :1101-1109
[2]   Development of a 35-MHz piezo-composite ultrasound array for medical imaging [J].
Cannata, JM ;
Williams, JA ;
Zhou, QF ;
Ritter, TA ;
Shung, KK .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (01) :224-236
[3]   Dual-frequency chirp imaging for contrast detection [J].
Cheng, Chih-Hao ;
Shen, Che-Chou ;
Yeh, Chih-Kuang .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (09) :2767-2778
[4]   Coded excitation for diagnostic ultrasound: A system developer's perspective [J].
Chiao, RY ;
Hao, XH .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (02) :160-170
[5]   Utilization of miniprobes in modern endoscopic ultrasonography [J].
Cysewska-Sobusiak, Anna ;
Skrzywanek, Pawel ;
Sowier, Aleksander .
IEEE SENSORS JOURNAL, 2006, 6 (05) :1323-1330
[6]  
GOSS S A, 1979, Ultrasound in Medicine and Biology, V5, P181, DOI 10.1016/0301-5629(79)90086-3
[7]  
Jensen J. A., 1996, Medical & Biological Engineering & Computing, V34, P351
[8]   The Study of Cable Effect on High-Frequency Ultrasound Transducer Performance [J].
Jian, Xiaohua ;
Li, Zhangjian ;
Han, Zhile ;
Xu, Jie ;
Liu, Pengbo ;
Liu, Yufei ;
Cui, Yaoyao ;
Huang, Wenbin .
IEEE SENSORS JOURNAL, 2018, 18 (13) :5265-5271
[9]   A High Frequency Geometric Focusing Transducer Based on 1-3 Piezocomposite for Intravascular Ultrasound Imaging [J].
Jian, Xiaohua ;
Han, Zhile ;
Liu, Pengbo ;
Xu, Jie ;
Li, Zhangjian ;
Li, Peiyang ;
Shao, Weiwei ;
Cui, Yaoyao .
BIOMED RESEARCH INTERNATIONAL, 2017, 2017
[10]   Chirp-coded excitation imaging with a high-frequency ultrasound annular array [J].
Mamou, Jonathan ;
Ketterling, Jeffrey A. ;
Silverman, Ronald H. .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2008, 55 (02) :508-513