Comparison of the Icare Tonometer and the Hand-held Goldmann Applanation Tonometer in Pediatric Aphakia

被引:11
|
作者
Li, Yunping [1 ]
Tang, Luosheng [1 ]
Xiao, Manyi [1 ]
Jia, Songbai [1 ]
Liu, Ping [1 ]
Zhou, Yi [1 ]
Chen, Huihui [1 ]
Zeng, Jiexi [1 ]
机构
[1] Cent S Univ, Xiangya Hosp 2, Dept Ophthalmol, Changsha 410011, Hunan, Peoples R China
关键词
tonometer; chlorate hydrate; pediatric aphakia; Perkins; MEASURING INTRAOCULAR-PRESSURE; OCULAR-RESPONSE-ANALYZER; REBOUND TONOMETER; CLINICAL-EVALUATION; GLAUCOMA; CHILDREN;
D O I
10.1097/IJG.0b013e31825afc87
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: Purpose: To compare the intraocular pressure (IOP) measurements obtained by the Icare and the hand-held Goldmann applanation tonometer (also called Perkins) in aphakic children after congenital cataract surgery. Methods: Methods: We investigated 125 children with aphakia after congenital cataract surgery in this study. A younger group (3 to 30 mo) and an elder group (31 to 72 mo) were divided in those patients by their age. In the younger group, all measurements were performed under sedation using chloral hydrate. Axial length of the eye and central corneal thickness (CCT) were also measured from all patients. Results: Results: Significant correlation was found on IOP readings obtained by the Icare and the Perkins tonometers (r=0.943, P < 0.001). After establishing a Bland-Altman plot, we found that 95% limit of the agreement between the 2 methods distributed between -1.6 to 5.6 mm Hg. The IOPs recorded from the Icare increased faster than that from the Perkins tonometer with the increase of the CCT thickness; a significant association relationship was found on the IOP difference between the 2 measurements (r=0.408, P < 0.001). However, no statistical correlation was identified between the axial length and the IOPs recorded by either tonometer. Conclusions: Conclusions: Most of the young patients accepted the Icare tonometer under unsedated conditions. This significant advantage indicated that the Icare tonometer will be overall better tolerated in pediatric aphakia population, although it could overestimate the IOPs compared with the measurements obtained from the Perkins tonometer. Differences in readings between the 2 tonometers become bigger as the CCT increase.
引用
收藏
页码:550 / 554
页数:5
相关论文
共 50 条
  • [1] Icare-Pro Rebound Tonometer Versus Hand-held Applanation Tonometer for Pediatric Screening
    Arribas-Pardo, Paula
    Mendez-Hernandez, Carmen
    Valls-Ferran, Isabel
    Puertas-Bordallo, Diego
    JOURNAL OF PEDIATRIC OPHTHALMOLOGY & STRABISMUS, 2018, 55 (06) : 382 - 386
  • [2] Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer
    Gao, Feng
    Liu, Hu
    Zhao, Qing
    Pan, Yingzhe
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2017, 13 (05) : 1912 - 1916
  • [3] The Icare-Pro Rebound Tonometer Versus the Hand-held Applanation Tonometer in Congenital Glaucoma
    Borrego Sanz, Lara
    Morales-Fernandez, Laura
    Martinez de-la-Casa, Jose M.
    Saenz-Frances, Federico
    Fuentes, Manuel
    Garcia-Feijoo, Julian
    JOURNAL OF GLAUCOMA, 2016, 25 (02) : 149 - 154
  • [4] Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients
    Brusini, P
    Salvetat, ML
    Zeppieri, M
    Tosoni, C
    Parisi, L
    JOURNAL OF GLAUCOMA, 2006, 15 (03) : 213 - 217
  • [5] Comparison of ICare, dynamic contour tonometer, and ocular response analyzer with Goldmann applanation tonometer in patients with glaucoma
    Vandewalle, Evelien
    Vandenbroeck, Sofie
    Stalmans, Ingeborg
    Zeyen, Thierry
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2009, 19 (05) : 783 - 789
  • [6] Evaluation of a New Rebound Self-tonometer, Icare HOME: Comparison With Goldmann Applanation Tonometer
    Takagi, Daisuke
    Sawada, Akira
    Yamamoto, Tetsuya
    JOURNAL OF GLAUCOMA, 2017, 26 (07) : 613 - 618
  • [7] Clinical comparison of the Icare tonometer and Goldmann applanation tonometry
    Pakrou, Nima
    Gray, Tim
    Mills, Richard
    Landers, John
    Craig, Jamie
    JOURNAL OF GLAUCOMA, 2008, 17 (01) : 43 - 47
  • [8] Comparison of the iCare tonometer with the Goldmann tonometer in Malawi
    Hohmann, J.
    Schulze-Schwering, M.
    Nyaka, T. Chirambo
    Moyo, V.
    Kayange, P. C.
    Doycheva, D.
    Batumba, N. H.
    Spitzer, M. S.
    OPHTHALMOLOGE, 2012, 109 (11): : 1098 - 1102
  • [9] Comparison of the ICare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists
    L M Abraham
    N C R Epasinghe
    D Selva
    R Casson
    Eye, 2008, 22 : 503 - 506
  • [10] Comparison of the ICare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists
    Abraham, L. M.
    Epasinghe, N. C. R.
    Selva, D.
    Casson, R.
    EYE, 2008, 22 (04) : 503 - 506