Quality assurance of geometric accuracy based on an electronic portal imaging device and log data analysis for Dynamic WaveArc irradiation

被引:6
作者
Hirashima, Hideaki [1 ]
Miyabe, Yuki [1 ]
Nakamura, Mitsuhiro [1 ]
Mukumoto, Nobutaka [1 ]
Mizowaki, Takashi [1 ]
Hiraoka, Masahiro [1 ]
机构
[1] Kyoto Univ, Grad Sch Med, Dept Radiat Oncol & Image Appl Therapy, Kyoto, Japan
关键词
Dynamic WaveArc; electronic portal imaging device; log data; quality assurance; MODULATED ARC THERAPY; RADIATION-THERAPY; DELIVERY; RADIOTHERAPY; SYSTEM; VMAT; VERIFICATION; IMRT; QUANTIFICATION; RAPIDARC;
D O I
10.1002/acm2.12324
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to develop a simple verification method for the routine quality assurance (QA) of Dynamic WaveArc (DWA) irradiation using electronic portal imaging device (EPID) images and log data analysis. First, an automatic calibration method utilizing the outermost multileaf collimator (MLC) slits was developed to correct the misalignment between the center of the EPID and the beam axis. Moreover, to verify the detection accuracy of the MLC position according to the EPID images, various positions of the MLC with intentional errors in the range 0.1-1 mm were assessed. Second, to validate the geometric accuracy during DWA irradiation, tests were designed in consideration of three indices. Test 1 evaluated the accuracy of the MLC position. Test 2 assessed dose output consistency with variable dose rate (160-400 MU/min), gantry speed (2.2-6 degrees/s), and ring speed (0.5-2.7 degrees/s). Test 3 validated dose output consistency with variable values of the above parameters plus MLC speed (1.6-4.2 cm/s). All tests were delivered to the EPID and compared with those obtained using a stationary radiation beam with a 0 degrees gantry angle. Irradiation log data were recorded simultaneously. The 0.1-mm intentional error on the MLC position could be detected by the EPID, which is smaller than the EPID pixel size. In Test 1, the MLC slit widths agreed within 0.20 mm of their exposed values. The averaged root-mean-square error (RMSE) of the dose outputs was less than 0.8% in Test 2 and Test 3. Using log data analysis in Test 3, the RMSE between the planned and recorded data was 0.1 mm, 0.12 degrees, and 0.07 degrees for the MLC position, gantry angle, and ring angle, respectively. The proposed method is useful for routine QA of the accuracy of DWA.
引用
收藏
页码:234 / 242
页数:9
相关论文
共 35 条
[1]   Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries [J].
Agnew, A. ;
Agnew, C. E. ;
Grattan, M. W. D. ;
Hounsell, A. R. ;
McGarry, C. K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09) :N49-N63
[2]   Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT [J].
Agnew, Christina E. ;
Irvine, Denise M. ;
McGarry, Conor K. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (06) :204-216
[3]  
Alber M, 2008, RADIOTHER ONCOL, V76, P1
[4]   A quantification of the effectiveness of EPID dosimetry and software-based plan verification systems in detecting incidents in radiotherapy [J].
Bojechko, Casey ;
Phillps, Mark ;
Kalet, Alan ;
Ford, Eric C. .
MEDICAL PHYSICS, 2015, 42 (09) :5363-5369
[5]   Treating patients with Dynamic Wave Arc: First clinical experience [J].
Burghelea, Manuela ;
Verellen, Dirk ;
Dhont, Jennifer ;
Hung, Cecilia ;
Gevaert, Thierry ;
Van den Begin, Robbe ;
Collen, Christine ;
Poels, Kenneth ;
Tournel, Koen ;
Boussaer, Marlies ;
Jaudet, Cyril ;
Reynders, Truus ;
Simon, Viorica ;
de Ridder, Mark .
RADIOTHERAPY AND ONCOLOGY, 2017, 122 (03) :347-351
[6]   Initial characterization, dosimetric benchmark and performance validation of Dynamic Wave Arc [J].
Burghelea, Manuela ;
Verellen, Dirk ;
Poels, Kenneth ;
Hung, Cecilia ;
Nakamura, Mitsuhiro ;
Dhont, Jennifer ;
Gevaert, Thierry ;
Van den Begin, Robbe ;
Collen, Christine ;
Matsuo, Yukinori ;
Kishi, Takahiro ;
Simon, Viorica ;
Hiraoka, Masahiro ;
de Ridder, Mark .
RADIATION ONCOLOGY, 2016, 11
[7]   Geometric Verification of Dynamic Wave Arc Delivery With the Vero System Using Orthogonal X-ray Fluoroscopic Imaging [J].
Burghelea, Manuela ;
Verellen, Dirk ;
Poels, Kenneth ;
Gevaert, Thierry ;
Depuydt, Tom ;
Tournel, Koen ;
Hung, Cecilia ;
Simon, Viorica ;
Hiraoka, Masahiro ;
de Ridder, Mark .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 92 (04) :754-761
[8]   Use of EPID for leaf position accuracy QA of dynamic multi-leaf collimator (DMLC) treatment [J].
Chang, J ;
Obcemea, CH ;
Sillanpaa, J ;
Mechalakos, J ;
Burman, C .
MEDICAL PHYSICS, 2004, 31 (07) :2091-2096
[9]   FEASIBILITY OF SINGLE-ISOCENTER VOLUMETRIC MODULATED ARC RADIOSURGERY FOR TREATMENT OF MULTIPLE BRAIN METASTASES [J].
Clark, Grant M. ;
Popple, Richard A. ;
Young, P. Edward ;
Fiveash, John B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (01) :296-302
[10]   A treatment planning study comparing volumetric arc modulation with RapidArc and fixed field IMRT for cervix uteri radiotherapy [J].
Cozzi, Luca ;
Dinshaw, Ketayun Ardeshir ;
Shrivastava, Shyam Kishore ;
Mahantshetty, Umesh ;
Engineer, Reena ;
Deshpande, Deepak Dattatray ;
Jamema, S. V. ;
Vanetti, Eugenio ;
Clivio, Alessandro ;
Nicolini, Giorgia ;
Fogliata, Antonella .
RADIOTHERAPY AND ONCOLOGY, 2008, 89 (02) :180-191