Cerebral foreign body reaction due to hydrophilic polymer embolization following aneurysm treatment by pipeline flow diversion device

被引:11
作者
Geisbush, Thomas Rand [1 ]
Marks, Michael P. [2 ]
Heit, Jeremy J. [2 ]
机构
[1] Rosalind Franklin Univ Med & Sci, Chicago Med Sch, Chicago, IL USA
[2] Stanford Hosp & Clin, 300 Pasteur Dr,Room S-047, Stanford, CA 94305 USA
关键词
Granulomatous inflammation; foreign body reaction; hydrophilic polymer embolization; flow diverting stent; cerebral aneurysm; ENDOVASCULAR TREATMENT; COMPLICATIONS;
D O I
10.1177/1591019919830767
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The use of flow diverting stents for wide based, intracranial aneurysms has become an invaluable treatment option. While intracranial hemorrhage and ischemic stroke from dislodged atherosclerotic emboli are common adverse events, the potential for delayed granulomatous inflammation from possible hydrophilic polymer emboli is rarely recognized. We present a unique case in which visible chipping of the pusher wire for stent placement was observed, followed by clinical and radiographic evidence suggestive of a delayed foreign body reaction to intracranial hydrophilic polymer emboli. A 55-year-old woman underwent placement of a Pipeline embolization device for a left-sided, broad-based aneurysm at the base of the internal carotid artery and posterior communicating artery. Two months later she developed right-sided focal neurological deficits. Imaging showed ipsilateral focal edema and enhancing lesions with contrast. Although not confirmed with biopsy and histopathology, clinical and radiographic evidence suggests that this patient probably experienced a delayed foreign body reaction to hydrophilic polymer emboli from compromised procedural equipment during flow diverting stent placement. Although previously described, this is the first instance to our knowledge in whichvisible chipping of the pusher wire was observed on a Pipeline embolization device.
引用
收藏
页码:447 / 453
页数:7
相关论文
共 20 条
[1]  
Ansari SA, 2018, J NEUROINTERV SURG, V19, P1
[2]  
Barnwell SL, 1997, AM J NEURORADIOL, V18, P1886
[3]   Pipeline for Uncoilable or Failed Aneurysms: Results from a Multicenter Clinical Trial [J].
Becske, Tibor ;
Kallmes, David F. ;
Saatci, Isil ;
McDougall, Cameron G. ;
Szikora, Istvn ;
Lanzino, Giuseppe ;
Moran, Christopher J. ;
Woo, Henry H. ;
Lopes, Demetrius K. ;
Berez, Aaron L. ;
Cher, Daniel J. ;
Siddiqui, Adnan H. ;
Levy, Elad I. ;
Albuquerque, Felipe C. ;
Fiorella, David J. ;
Berentei, Zsolt ;
Marosfoi, Miklos ;
Cekirge, Saruhan H. ;
Nelson, Peter K. .
RADIOLOGY, 2013, 267 (03) :858-868
[4]   Endovascular treatment of cerebral aneurysms using flow-diverter devices: A systematic review [J].
Briganti, Francesco ;
Leone, Giuseppe ;
Marseglia, Mariano ;
Mariniello, Giuseppe ;
Caranci, Ferdinando ;
Brunetti, Arturo ;
Maiuri, Francesco .
NEURORADIOLOGY JOURNAL, 2015, 28 (04) :365-375
[5]   Hydrophilic Polymer-associated Ischemic Enterocolitis [J].
Chavez, Jesus A. ;
Chen, Wei ;
Frankel, Wendy L. ;
Arnold, Christina A. .
AMERICAN JOURNAL OF SURGICAL PATHOLOGY, 2017, 41 (02) :271-276
[6]   Enhancing Brain Lesions after Endovascular Treatment of Aneurysms [J].
Cruz, J. P. ;
Marotta, T. ;
O'Kelly, C. ;
Holtmannspoetter, M. ;
Saliou, G. ;
Willinsky, R. ;
Krings, T. ;
Agid, R. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (10) :1954-1958
[7]   Complications of endovascular polymers associated with vascular introducer sheaths and metallic coils in 3 patients, with literature review [J].
Fealey, Michael E. ;
Edwards, William D. ;
Giannini, Caterina ;
Piepgras, David G. ;
Cloft, Harry ;
Rihal, Charanjit S. .
AMERICAN JOURNAL OF SURGICAL PATHOLOGY, 2008, 32 (09) :1310-1316
[8]  
Food and Drug Administration, 2015, INTR MED DEV FDA SAF
[9]   Distal Embolization of Hydrophilic-Coating Material From Coronary Guidewires After Percutaneous Coronary Interventions [J].
Grundeken, Maik J. ;
Li, Xiaofei ;
Kurpershoek, C. Eline ;
Kramer, Miranda C. ;
Vink, Aryan ;
Piek, Jan J. ;
Tijssen, Jan G. P. ;
Koch, Karel T. ;
Wykrzykowska, Joanna J. ;
de Winter, Robbert J. ;
van der Wal, Allard C. .
CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2015, 8 (02)
[10]   Histopathological assessment of fatal ipsilateral intraparenchymal hemorrhages after the treatment of supraclinoid aneurysms with the Pipeline Embolization Device [J].
Hu, Yin C. ;
Deshmukh, Vivek R. ;
Albuquerque, Felipe C. ;
Fiorella, David ;
Nixon, Randal R. ;
Heck, Donald V. ;
Barnwell, Stanley L. ;
McDougall, Cameron G. .
JOURNAL OF NEUROSURGERY, 2014, 120 (02) :365-374